Evaluation of a 2D detector array for patient-specific VMAT QA with different setups

被引:49
作者
Boggula, Ramesh [1 ]
Birkner, Mattias [2 ]
Lohr, Frank [1 ]
Steil, Volker [1 ]
Wenz, Frederik [1 ]
Wertz, Hansjoerg [1 ]
机构
[1] Univ Med Ctr Mannheim, Dept Radiat Oncol, Mannheim, Germany
[2] IBA Dosimetry, Schwarzenbruck, Germany
关键词
VOLUMETRIC MODULATED ARC; PIXEL IONIZATION-CHAMBER; PHOTON-BEAM ATTENUATION; IMRT QUALITY-ASSURANCE; CARBON-FIBER COUCH; TREATMENT DELIVERY; THERAPY; VERIFICATION; RADIOTHERAPY; MATRIXX;
D O I
10.1088/0031-9155/56/22/011
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
For pre-treatment plan verification of advanced treatment techniques such as intensity-modulated arc therapy, a fast and reliable dosimetric device is required. In this study, we investigated the suitability of MatriXX in different setups for verification of volumetric modulated arc therapy (VMAT) plans. If MatriXX is used in a stationary phantom (MULTICube), the measured dose is dependent on the beam angle. For the first setup (MatriXX/MULTICube), we developed correction factors (CFs) for each detector element (1020 CFs). We investigated the accuracy of these CFs by verifying 12 VMAT plans. In the second setup, we also assessed the suitability of MatriXX in a dedicated holder. Using this setup (MatriXX/Holder), 30 additional VMAT plans were verified. Deviations of up to similar to 17% and similar to 11% were noted for one of the ion chambers at 90 degrees and 180 degrees gantry positions. The influence of the beam angle dependence (MULTICube) can explicitly be seen when a gamma criterion of 2%/2 mm was chosen. An overall improvement of 4.3% of passing pixels (pp) was noted after applying beam angular-dependent CFs. When the gamma criterion was 3%/3 mm, the %pp was >= 95% without and similar to 100% with correction. With the second setup, MatriXX/holder, we showed excellent agreement between measurements and calculations. The % pp averaged over all plans (30 VMAT treatment plans) was nearly similar to 100%. The combination of MatriXX with MULTICube or with holder proved to be a fast and reliable method for pretreatment verification of arc therapy with sufficient accuracy.
引用
收藏
页码:7163 / 7177
页数:15
相关论文
共 50 条
[41]   Evaluation of Delta4DVH Anatomy in 3D Patient-Specific IMRT Quality Assurance [J].
Tang, Du ;
Yang, Zhen ;
Dai, Xunzhang ;
Cao, Ying .
TECHNOLOGY IN CANCER RESEARCH & TREATMENT, 2020, 19
[42]   Evaluation of tumor microvasculature with 3D ultrasound localization microscopy based on 2D matrix array [J].
Zhang, Changlu ;
Lei, Shuang ;
Ma, Aiqing ;
Wang, Bing ;
Wang, Shuo ;
Liu, Jiamei ;
Shang, Dongqing ;
Zhang, Qi ;
Li, Yongchuan ;
Zheng, Hairong ;
Ma, Teng .
EUROPEAN RADIOLOGY, 2024, 34 (08) :5250-5259
[43]   Patient-Specific Hyperparameter Optimization of a Deep Learning-Based Tumor Autocontouring Algorithm on 2D Liver, Prostate, and Lung Cine MR Images: A Pilot Study [J].
Han, Gawon ;
Wachowicz, Keith ;
Usmani, Nawaid ;
Yee, Don ;
Wong, Jordan ;
Elangovan, Arun ;
Yun, Jihyun ;
Fallone, B. Gino .
ALGORITHMS, 2025, 18 (04)
[44]   Commissioning and clinical implementation of the first commercial independent Monte Carlo 3D dose calculation to replace CyberKnife M6™ patient-specific QA measurements [J].
Milder, Maaike T. W. ;
Alber, Markus ;
Soehn, Matthias ;
Hoogeman, Mischa S. .
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2020, 21 (11) :304-311
[45]   A technique for quantifying the sensitivity of dosimetric tool gamma with 2D detector array in pretreatment IMRT plans by segment deletion method [J].
Krishnan, M. P. Arun ;
Bhagyalakshmi, A. T. ;
Holla, Raghavendra ;
Puzakkal, Niyas ;
Ranjith, C. P. ;
Vysakh, R. ;
Irfad, M. P. ;
Ramasubramanian, V ;
Hu, Jianping ;
Momeen, M. Ummal .
RADIOLOGIA MEDICA, 2021, 126 (03) :453-459
[46]   Clinical introduction of a linac head-mounted 2D detector array based quality assurance system in head and neck IMRT [J].
Korevaar, Erik W. ;
Wauben, David J. L. ;
van der Hulst, Peter C. ;
Langendijk, Johannes A. ;
Van't Veld, Aart A. .
RADIOTHERAPY AND ONCOLOGY, 2011, 100 (03) :446-452
[47]   The evaluation of a 2D diode array in "magic phantom" for use in high dose rate brachytherapy pretreatment quality assurance [J].
Espinoza, A. ;
Petasecca, M. ;
Fuduli, I. ;
Howie, A. ;
Bucci, J. ;
Corde, S. ;
Jackson, M. ;
Lerch, M. L. F. ;
Rosenfeld, A. B. .
MEDICAL PHYSICS, 2015, 42 (02) :663-673
[48]   Evaluation of a 3D-printed heterogeneous anthropomorphic head and neck phantom for patient-specific quality assurance in intensity-modulated radiation therapy [J].
Kadoya, Noriyuki ;
Abe, Kota ;
Nemoto, Hikaru ;
Sato, Kiyokazu ;
Ieko, Yoshiro ;
Ito, Kengo ;
Dobashi, Suguru ;
Takeda, Ken ;
Jingu, Keiichi .
RADIOLOGICAL PHYSICS AND TECHNOLOGY, 2019, 12 (03) :351-356
[49]   Evaluation of a 3D-printed heterogeneous anthropomorphic head and neck phantom for patient-specific quality assurance in intensity-modulated radiation therapy [J].
Noriyuki Kadoya ;
Kota Abe ;
Hikaru Nemoto ;
Kiyokazu Sato ;
Yoshiro Ieko ;
Kengo Ito ;
Suguru Dobashi ;
Ken Takeda ;
Keiichi Jingu .
Radiological Physics and Technology, 2019, 12 :351-356
[50]   Comprehensive Patient-Specific Intensity-Modulated Radiation Therapy Quality Assurance Comparing Mobius3D/FX to Conventional Methods of Evaluation [J].
Basavatia, Amar K. ;
Fiedler, Derek A. ;
Ritter, Johannes ;
Brodin, Patrik ;
Goddard, Lee C. ;
Jeong, Kyoungkeun ;
Hsu, Shu-Hui ;
Mynampati, Dinesh K. ;
Yaparpalvi, Ravindra ;
Tome, Wolfgang A. .
CUREUS JOURNAL OF MEDICAL SCIENCE, 2021, 13 (05)