Evaluation of a 2D detector array for patient-specific VMAT QA with different setups

被引:49
作者
Boggula, Ramesh [1 ]
Birkner, Mattias [2 ]
Lohr, Frank [1 ]
Steil, Volker [1 ]
Wenz, Frederik [1 ]
Wertz, Hansjoerg [1 ]
机构
[1] Univ Med Ctr Mannheim, Dept Radiat Oncol, Mannheim, Germany
[2] IBA Dosimetry, Schwarzenbruck, Germany
关键词
VOLUMETRIC MODULATED ARC; PIXEL IONIZATION-CHAMBER; PHOTON-BEAM ATTENUATION; IMRT QUALITY-ASSURANCE; CARBON-FIBER COUCH; TREATMENT DELIVERY; THERAPY; VERIFICATION; RADIOTHERAPY; MATRIXX;
D O I
10.1088/0031-9155/56/22/011
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
For pre-treatment plan verification of advanced treatment techniques such as intensity-modulated arc therapy, a fast and reliable dosimetric device is required. In this study, we investigated the suitability of MatriXX in different setups for verification of volumetric modulated arc therapy (VMAT) plans. If MatriXX is used in a stationary phantom (MULTICube), the measured dose is dependent on the beam angle. For the first setup (MatriXX/MULTICube), we developed correction factors (CFs) for each detector element (1020 CFs). We investigated the accuracy of these CFs by verifying 12 VMAT plans. In the second setup, we also assessed the suitability of MatriXX in a dedicated holder. Using this setup (MatriXX/Holder), 30 additional VMAT plans were verified. Deviations of up to similar to 17% and similar to 11% were noted for one of the ion chambers at 90 degrees and 180 degrees gantry positions. The influence of the beam angle dependence (MULTICube) can explicitly be seen when a gamma criterion of 2%/2 mm was chosen. An overall improvement of 4.3% of passing pixels (pp) was noted after applying beam angular-dependent CFs. When the gamma criterion was 3%/3 mm, the %pp was >= 95% without and similar to 100% with correction. With the second setup, MatriXX/holder, we showed excellent agreement between measurements and calculations. The % pp averaged over all plans (30 VMAT treatment plans) was nearly similar to 100%. The combination of MatriXX with MULTICube or with holder proved to be a fast and reliable method for pretreatment verification of arc therapy with sufficient accuracy.
引用
收藏
页码:7163 / 7177
页数:15
相关论文
共 50 条
[31]   A critical evaluation of the PTW 2D-ARRAY seven29 and OCTAVIUS II phantom for IMRT and VMAT verification [J].
Hussein, Mohammad ;
Adams, Elizabeth J. ;
Jordan, Thomas J. ;
Clark, Catharine H. ;
Nisbet, Andrew .
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2013, 14 (06) :274-292
[32]   Patient-specific biokinetics and hybrid 2D/3D approach integration in OEDIPE software: Application to radioiodine therapy [J].
Bensiali, M. ;
Anizan, N. ;
Leboulleux, S. ;
Lamart, S. ;
Davesne, E. ;
Broggio, D. ;
Desbree, A. ;
France, D. .
PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2023, 113
[33]   The root cause analysis on failed patient-specific measurements of pencil beam scanning protons using a 2D detection array with finite size ionization chambers [J].
Ricci, Jacob C. ;
Hsi, Wen C. ;
Su, Zhong ;
Mund, Karl ;
Dawson, Robert ;
Indelicato, Daniel J. .
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2021, 22 (08) :175-190
[34]   Evaluation of the sensitivity of two 3D diode array dosimetry systems to setup error for quality assurance (QA) of volumetric-modulated arc therapy (VMAT) [J].
Li, Guangjun ;
Bai, Sen ;
Chen, Nianyong ;
Henderson, Lansdale ;
Wu, Kui ;
Xiao, Jianghong ;
Zhang, Yingjie ;
Jiang, Qingfeng ;
Jiang, Xiaoqin .
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2013, 14 (05) :13-24
[35]   Technical Note: Evaluation of plastic scintillator detector for small field stereotactic patient-specific quality assurance [J].
Qin, Yujiao ;
Gardner, Stephen J. ;
Kim, Joshua ;
Huang, Yimei ;
Wen, Ning ;
Doemer, Anthony ;
Chetty, Indrin J. .
MEDICAL PHYSICS, 2017, 44 (10) :5509-5516
[36]   Correlation between patient-specific quality assurance in volumetric modulated arc therapy and 2D dose image features [J].
Guo, Yixiao ;
Hu, Jinyan ;
Li, Yang ;
Ran, Juntao ;
Cai, Hongyi .
SCIENTIFIC REPORTS, 2023, 13 (01)
[37]   Transit and non-transit 3D EPID dosimetry versus detector arrays for patient specific QA [J].
Olaciregui-Ruiz, Igor ;
Vivas-Maiques, Begona ;
Kaas, Jochem ;
Perik, Thijs ;
Wittkamper, Frits ;
Mijnheer, Ben ;
Mans, Anton .
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2019, 20 (06) :79-90
[38]   Dosimetric characteristics of the novel 2D ionization chamber array OCTAVIUS Detector 1500 [J].
Stelljes, T. S. ;
Harmeyer, A. ;
Reuter, J. ;
Looe, H. K. ;
Chofor, N. ;
Harder, D. ;
Poppe, B. .
MEDICAL PHYSICS, 2015, 42 (04) :1528-1537
[39]   Evaluation of patient DVH-based QA metrics for prostate VMAT: correlation between accuracy of estimated 3D patient dose and magnitude of MLC misalignment [J].
Kadoya, Noriyuki ;
Saito, Masahide ;
Ogasawara, Makoto ;
Fujita, Yukio ;
Ito, Kengo ;
Sato, Kiyokazu ;
Kishi, Kazuma ;
Dobashi, Suguru ;
Takeda, Ken ;
Jingu, Keiichi .
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2015, 16 (03) :179-189
[40]   Characterization and clinical evaluation of a novel 2D detector array for conventional and flattening filter free (FFF) IMRT pre-treatment verification [J].
Sekar, Yuvaraj ;
Thoelking, Johannes ;
Eckl, Miriam ;
Kalichava, Irakli ;
Sihono, Dwi Seno Kuncoro ;
Lohr, Frank ;
Wenz, Frederik ;
Wertz, Hansjoerg .
ZEITSCHRIFT FUR MEDIZINISCHE PHYSIK, 2018, 28 (02) :134-141