Evaluation of a 2D detector array for patient-specific VMAT QA with different setups

被引:49
作者
Boggula, Ramesh [1 ]
Birkner, Mattias [2 ]
Lohr, Frank [1 ]
Steil, Volker [1 ]
Wenz, Frederik [1 ]
Wertz, Hansjoerg [1 ]
机构
[1] Univ Med Ctr Mannheim, Dept Radiat Oncol, Mannheim, Germany
[2] IBA Dosimetry, Schwarzenbruck, Germany
关键词
VOLUMETRIC MODULATED ARC; PIXEL IONIZATION-CHAMBER; PHOTON-BEAM ATTENUATION; IMRT QUALITY-ASSURANCE; CARBON-FIBER COUCH; TREATMENT DELIVERY; THERAPY; VERIFICATION; RADIOTHERAPY; MATRIXX;
D O I
10.1088/0031-9155/56/22/011
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
For pre-treatment plan verification of advanced treatment techniques such as intensity-modulated arc therapy, a fast and reliable dosimetric device is required. In this study, we investigated the suitability of MatriXX in different setups for verification of volumetric modulated arc therapy (VMAT) plans. If MatriXX is used in a stationary phantom (MULTICube), the measured dose is dependent on the beam angle. For the first setup (MatriXX/MULTICube), we developed correction factors (CFs) for each detector element (1020 CFs). We investigated the accuracy of these CFs by verifying 12 VMAT plans. In the second setup, we also assessed the suitability of MatriXX in a dedicated holder. Using this setup (MatriXX/Holder), 30 additional VMAT plans were verified. Deviations of up to similar to 17% and similar to 11% were noted for one of the ion chambers at 90 degrees and 180 degrees gantry positions. The influence of the beam angle dependence (MULTICube) can explicitly be seen when a gamma criterion of 2%/2 mm was chosen. An overall improvement of 4.3% of passing pixels (pp) was noted after applying beam angular-dependent CFs. When the gamma criterion was 3%/3 mm, the %pp was >= 95% without and similar to 100% with correction. With the second setup, MatriXX/holder, we showed excellent agreement between measurements and calculations. The % pp averaged over all plans (30 VMAT treatment plans) was nearly similar to 100%. The combination of MatriXX with MULTICube or with holder proved to be a fast and reliable method for pretreatment verification of arc therapy with sufficient accuracy.
引用
收藏
页码:7163 / 7177
页数:15
相关论文
共 50 条
[21]   Improvement of deep learning prediction model in patient-specific QA for VMAT with MLC leaf position map and patient's dose distribution [J].
Tozuka, Ryota ;
Kadoya, Noriyuki ;
Tomori, Seiji ;
Kimura, Yuto ;
Kajikawa, Tomohiro ;
Sugai, Yuto ;
Xiao, Yushan ;
Jingu, Keiichi .
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2023, 24 (10)
[22]   Evaluation of a planar diode matrix for SRS patient-specific QA in comparison with GAFchromic films [J].
Infusino, Erminia ;
Ianiro, Anna ;
Luppino, Stefano ;
Nocentini, Sandro ;
Pugliatti, Cristina ;
Soriani, Antonella .
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2023, 24 (08)
[23]   Evaluation and Validation of IBA I'MatriXX Array for Patient-Specific Quality Assurance of TomoTherapy® [J].
Sresty, N. V. N. Madhusudhana ;
Raju, A. Krishnam ;
Reddy, B. Nagarjuna ;
Sahithya, V. C. ;
Mohmd, Yakub ;
Kumar, G. Deleep ;
Kumar, T. Anil ;
Rushdi, Tasneem ;
Aparna, S. ;
Prasad, G. Durga ;
Bajwa, Harjot Kaur .
JOURNAL OF MEDICAL PHYSICS, 2019, 44 (03) :222-227
[24]   A method to enhance 2D ion chamber array patient specific quality assurance for IMRT [J].
Diaz Moreno, Rogelio Manuel ;
Venencia, Daniel ;
Garrigo, Edgardo ;
Pipman, Yakov .
AUSTRALASIAN PHYSICAL & ENGINEERING SCIENCES IN MEDICINE, 2017, 40 (01) :145-151
[25]   Patient-specific QA using 4D Monte Carlo phase space predictions and EPID dosimetry [J].
Popescu, I. A. ;
Atwal, P. ;
Lobo, J. ;
Lucido, J. ;
McCurdy, B. M. C. .
8TH INTERNATIONAL CONFERENCE ON 3D RADIATION DOSIMETRY (IC3DDOSE), 2015, 573
[26]   Robotic radiosurgery system patient-specific QA for extracranial treatments using the planar ion chamber array and the cylindrical diode array [J].
Lin, Mu-Han ;
Veltchev, Iavor ;
Koren, Sion ;
Ma, Charlie ;
Li, Jinsgeng .
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2015, 16 (04) :290-305
[27]   Sensitivity evaluation of two commercial quality assurance systems to organ-dose variations of patient-specific VMAT plans [J].
Oderinde, Oluwaseyi M. ;
Du Plessis, Freek .
JOURNAL OF RADIATION RESEARCH AND APPLIED SCIENCES, 2019, 12 (01) :132-139
[28]   Machine Log File and Calibration Errors-based Patient-specific Quality Assurance (QA) for Volumetric Modulated Arc Therapy (VMAT) [J].
Ma, Yangguang ;
Mou, Xuanqin ;
Beeraka, Narasimha M. ;
Guo, Yuexin ;
Liu, Junqi ;
Dai, Jianrong ;
Fan, Ruitai .
CURRENT PHARMACEUTICAL DESIGN, 2023, 29 (34) :2738-2751
[29]   Commissioning and clinical implementation of an Autoencoder based Classification-Regression model for VMAT patient-specific QA in a multi-institution scenario [J].
Yang, Ruijie ;
Yang, Xueying ;
Wang, Le ;
Li, Dingjie ;
Guo, Yuexin ;
Li, Ying ;
Guan, Yumin ;
Wu, Xiangyang ;
Xu, Shouping ;
Zhang, Shuming ;
Chan, Maria F. ;
Geng, Lisheng ;
Sui, Jing .
RADIOTHERAPY AND ONCOLOGY, 2021, 161 :230-240
[30]   Improved patient-specific quality assurance for VMAT on film measurement with 2.5D gamma analysis [J].
Miura, Hideharu ;
Miyazawa, Masanori ;
Ozawa, Shuichi ;
Shida, Koichi ;
Hayata, Masahiro ;
Kagemoto, Masayuki .
PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2025, 135