Evaluation of a 2D detector array for patient-specific VMAT QA with different setups

被引:48
作者
Boggula, Ramesh [1 ]
Birkner, Mattias [2 ]
Lohr, Frank [1 ]
Steil, Volker [1 ]
Wenz, Frederik [1 ]
Wertz, Hansjoerg [1 ]
机构
[1] Univ Med Ctr Mannheim, Dept Radiat Oncol, Mannheim, Germany
[2] IBA Dosimetry, Schwarzenbruck, Germany
关键词
VOLUMETRIC MODULATED ARC; PIXEL IONIZATION-CHAMBER; PHOTON-BEAM ATTENUATION; IMRT QUALITY-ASSURANCE; CARBON-FIBER COUCH; TREATMENT DELIVERY; THERAPY; VERIFICATION; RADIOTHERAPY; MATRIXX;
D O I
10.1088/0031-9155/56/22/011
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
For pre-treatment plan verification of advanced treatment techniques such as intensity-modulated arc therapy, a fast and reliable dosimetric device is required. In this study, we investigated the suitability of MatriXX in different setups for verification of volumetric modulated arc therapy (VMAT) plans. If MatriXX is used in a stationary phantom (MULTICube), the measured dose is dependent on the beam angle. For the first setup (MatriXX/MULTICube), we developed correction factors (CFs) for each detector element (1020 CFs). We investigated the accuracy of these CFs by verifying 12 VMAT plans. In the second setup, we also assessed the suitability of MatriXX in a dedicated holder. Using this setup (MatriXX/Holder), 30 additional VMAT plans were verified. Deviations of up to similar to 17% and similar to 11% were noted for one of the ion chambers at 90 degrees and 180 degrees gantry positions. The influence of the beam angle dependence (MULTICube) can explicitly be seen when a gamma criterion of 2%/2 mm was chosen. An overall improvement of 4.3% of passing pixels (pp) was noted after applying beam angular-dependent CFs. When the gamma criterion was 3%/3 mm, the %pp was >= 95% without and similar to 100% with correction. With the second setup, MatriXX/holder, we showed excellent agreement between measurements and calculations. The % pp averaged over all plans (30 VMAT treatment plans) was nearly similar to 100%. The combination of MatriXX with MULTICube or with holder proved to be a fast and reliable method for pretreatment verification of arc therapy with sufficient accuracy.
引用
收藏
页码:7163 / 7177
页数:15
相关论文
共 50 条
  • [21] Evaluation and Validation of IBA I'MatriXX Array for Patient-Specific Quality Assurance of TomoTherapy®
    Sresty, N. V. N. Madhusudhana
    Raju, A. Krishnam
    Reddy, B. Nagarjuna
    Sahithya, V. C.
    Mohmd, Yakub
    Kumar, G. Deleep
    Kumar, T. Anil
    Rushdi, Tasneem
    Aparna, S.
    Prasad, G. Durga
    Bajwa, Harjot Kaur
    JOURNAL OF MEDICAL PHYSICS, 2019, 44 (03) : 222 - 227
  • [22] A method to enhance 2D ion chamber array patient specific quality assurance for IMRT
    Diaz Moreno, Rogelio Manuel
    Venencia, Daniel
    Garrigo, Edgardo
    Pipman, Yakov
    AUSTRALASIAN PHYSICAL & ENGINEERING SCIENCES IN MEDICINE, 2017, 40 (01) : 145 - 151
  • [23] Patient-specific QA using 4D Monte Carlo phase space predictions and EPID dosimetry
    Popescu, I. A.
    Atwal, P.
    Lobo, J.
    Lucido, J.
    McCurdy, B. M. C.
    8TH INTERNATIONAL CONFERENCE ON 3D RADIATION DOSIMETRY (IC3DDOSE), 2015, 573
  • [24] Robotic radiosurgery system patient-specific QA for extracranial treatments using the planar ion chamber array and the cylindrical diode array
    Lin, Mu-Han
    Veltchev, Iavor
    Koren, Sion
    Ma, Charlie
    Li, Jinsgeng
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2015, 16 (04): : 290 - 305
  • [25] Sensitivity evaluation of two commercial quality assurance systems to organ-dose variations of patient-specific VMAT plans
    Oderinde, Oluwaseyi M.
    Du Plessis, Freek
    JOURNAL OF RADIATION RESEARCH AND APPLIED SCIENCES, 2019, 12 (01) : 132 - 139
  • [26] Machine Log File and Calibration Errors-based Patient-specific Quality Assurance (QA) for Volumetric Modulated Arc Therapy (VMAT)
    Ma, Yangguang
    Mou, Xuanqin
    Beeraka, Narasimha M.
    Guo, Yuexin
    Liu, Junqi
    Dai, Jianrong
    Fan, Ruitai
    CURRENT PHARMACEUTICAL DESIGN, 2023, 29 (34) : 2738 - 2751
  • [27] Commissioning and clinical implementation of an Autoencoder based Classification-Regression model for VMAT patient-specific QA in a multi-institution scenario
    Yang, Ruijie
    Yang, Xueying
    Wang, Le
    Li, Dingjie
    Guo, Yuexin
    Li, Ying
    Guan, Yumin
    Wu, Xiangyang
    Xu, Shouping
    Zhang, Shuming
    Chan, Maria F.
    Geng, Lisheng
    Sui, Jing
    RADIOTHERAPY AND ONCOLOGY, 2021, 161 : 230 - 240
  • [28] A critical evaluation of the PTW 2D-ARRAY seven29 and OCTAVIUS II phantom for IMRT and VMAT verification
    Hussein, Mohammad
    Adams, Elizabeth J.
    Jordan, Thomas J.
    Clark, Catharine H.
    Nisbet, Andrew
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2013, 14 (06): : 274 - 292
  • [29] The root cause analysis on failed patient-specific measurements of pencil beam scanning protons using a 2D detection array with finite size ionization chambers
    Ricci, Jacob C.
    Hsi, Wen C.
    Su, Zhong
    Mund, Karl
    Dawson, Robert
    Indelicato, Daniel J.
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2021, 22 (08): : 175 - 190
  • [30] Patient-specific biokinetics and hybrid 2D/3D approach integration in OEDIPE software: Application to radioiodine therapy
    Bensiali, M.
    Anizan, N.
    Leboulleux, S.
    Lamart, S.
    Davesne, E.
    Broggio, D.
    Desbree, A.
    France, D.
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2023, 113