Evaluation of a 2D detector array for patient-specific VMAT QA with different setups

被引:48
作者
Boggula, Ramesh [1 ]
Birkner, Mattias [2 ]
Lohr, Frank [1 ]
Steil, Volker [1 ]
Wenz, Frederik [1 ]
Wertz, Hansjoerg [1 ]
机构
[1] Univ Med Ctr Mannheim, Dept Radiat Oncol, Mannheim, Germany
[2] IBA Dosimetry, Schwarzenbruck, Germany
关键词
VOLUMETRIC MODULATED ARC; PIXEL IONIZATION-CHAMBER; PHOTON-BEAM ATTENUATION; IMRT QUALITY-ASSURANCE; CARBON-FIBER COUCH; TREATMENT DELIVERY; THERAPY; VERIFICATION; RADIOTHERAPY; MATRIXX;
D O I
10.1088/0031-9155/56/22/011
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
For pre-treatment plan verification of advanced treatment techniques such as intensity-modulated arc therapy, a fast and reliable dosimetric device is required. In this study, we investigated the suitability of MatriXX in different setups for verification of volumetric modulated arc therapy (VMAT) plans. If MatriXX is used in a stationary phantom (MULTICube), the measured dose is dependent on the beam angle. For the first setup (MatriXX/MULTICube), we developed correction factors (CFs) for each detector element (1020 CFs). We investigated the accuracy of these CFs by verifying 12 VMAT plans. In the second setup, we also assessed the suitability of MatriXX in a dedicated holder. Using this setup (MatriXX/Holder), 30 additional VMAT plans were verified. Deviations of up to similar to 17% and similar to 11% were noted for one of the ion chambers at 90 degrees and 180 degrees gantry positions. The influence of the beam angle dependence (MULTICube) can explicitly be seen when a gamma criterion of 2%/2 mm was chosen. An overall improvement of 4.3% of passing pixels (pp) was noted after applying beam angular-dependent CFs. When the gamma criterion was 3%/3 mm, the %pp was >= 95% without and similar to 100% with correction. With the second setup, MatriXX/holder, we showed excellent agreement between measurements and calculations. The % pp averaged over all plans (30 VMAT treatment plans) was nearly similar to 100%. The combination of MatriXX with MULTICube or with holder proved to be a fast and reliable method for pretreatment verification of arc therapy with sufficient accuracy.
引用
收藏
页码:7163 / 7177
页数:15
相关论文
共 50 条
[1]   Evaluation and improvement of angular response for a commercial 2D detector array for patient-specific QA [J].
Wang, Sha ;
Zheng, Yuanshui .
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2023, 24 (09)
[2]   Validation of a digital method for patient-specific verification of VMAT treatment using a 2D ionisation detector array [J].
U'wais, Fatima Adel ;
Radzi, Yasmin ;
Rizan, Naslinda Noor ;
Zin, Hafiz M. .
RADIATION PHYSICS AND CHEMISTRY, 2023, 202
[3]   Comparative performance analysis of 2D and 3D gamma metrics for patient specific QA in VMAT using Octavius 4D with 2D-Array 1500 [J].
Pal, Bipasha ;
Pal, Angshuman ;
Bag, Santanu ;
Ali, Md Abbas ;
Das, Suresh ;
Palit, Soura ;
Sarkar, Papai ;
Mallik, Suman ;
Goswami, Jyotirup ;
Das, Sayan ;
Manir, Kazi Sazzad ;
Sen, Arijit ;
Mondol, Monidipa .
PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2021, 91 :18-27
[4]   A study on correlation between 2D and 3D gamma evaluation metrics in patient-specific quality assurance for VMAT [J].
Rajasekaran, Dhanabalan ;
Jeevanandam, Prakash ;
Sukumar, Prabakar ;
Ranganathan, Arulpandiyan ;
Johnjothi, Samdevakumar ;
Nagarajan, Vivekanandan .
MEDICAL DOSIMETRY, 2014, 39 (04) :300-308
[5]   A feasibility study of deep learning prediction model for VMAT patient-specific QA [J].
Miao, Junjie ;
Xu, Yuan ;
Men, Kuo ;
Dai, Jianrong .
FRONTIERS IN ONCOLOGY, 2025, 15
[6]   Dosimetric characteristics of a 2D silicon diode array for stereotactic radiotherapy end-to-end patient-specific QA [J].
Bai, Long ;
Xiao, Qing ;
Wang, Qiang ;
Zhao, Jianling ;
Li, Guangjun ;
Bai, Sen .
RADIATION PHYSICS AND CHEMISTRY, 2020, 173
[7]   Curtailing patient-specific IMRT QA procedures from 2D dose error distribution [J].
Kurosu, Keita ;
Sumida, Iori ;
Mizuno, Hirokazu ;
Otani, Yuki ;
Oda, Michio ;
Isohashi, Fumiaki ;
Seo, Yuji ;
Suzuki, Osamu ;
Ogawa, Kazuhiko .
JOURNAL OF RADIATION RESEARCH, 2016, 57 (03) :258-264
[8]   Development of a patient-specific 3D dose evaluation program for QA in radiation therapy [J].
Lee, Suk ;
Chang, Kyung Hwan ;
Cao, Yuan Jie ;
Shim, Jang Bo ;
Yang, Dae Sik ;
Park, Young Je ;
Yoon, Won Sup ;
Kim, Chul Yong .
JOURNAL OF THE KOREAN PHYSICAL SOCIETY, 2015, 66 (06) :859-866
[9]   Error detection sensitivity test using complex errors on three patient-specific VMAT QA systems [J].
Thongsawad, S. ;
Srisatit, S. ;
Fuangrod, T. .
INTERNATIONAL NUCLEAR SCIENCE AND TECHNOLOGY CONFERENCE, 2019, 1285
[10]   Assessment of radiobiological metrics applied to patient-specific QA process of VMAT prostate treatments [J].
Clemente-Gutierrez, Francisco ;
Perez-Vara, Consuelo ;
Clavo-Herranz, Maria H. ;
Lopez-Carrizosa, Concepcion ;
Perez-Regadera, Jose ;
Ibanez-Villoslada, Carmen .
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2016, 17 (02) :341-367