Regional Hounsfield unit measurement of screw trajectory for predicting pedicle screw fixation using cortical bone trajectory: a retrospective cohort study

被引:36
|
作者
Matsukawa, Keitaro [1 ]
Abe, Yuichiro [2 ]
Yanai, Yoshihide [1 ]
Yato, Yoshiyuki [1 ]
机构
[1] Natl Hosp Org, Murayama Med Ctr, Dept Orthopaed Surg, Musashimurayama, Tokyo, Japan
[2] Wajokai Eniwa Hosp, Dept Orthopaed Surg, Eniwa, Hokkaido, Japan
关键词
Hounsfield units; Pedicle screw fixation; Screw loosening; Insertional torque; Bone density; Simulation analysis; Cortical bone trajectory; PULL-OUT STRENGTH; MINERAL DENSITY; INSERTIONAL TORQUE; IN-VITRO; STABILITY; FUSION; SPINE;
D O I
10.1007/s00701-017-3424-5
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
Background The sufficiency of screw anchoring is a critical factor for achieving successful spinal fusion; however, no reliable method for predicting pedicle screw fixation has been established. Recently, Hounsfield units (HU) obtained from computed tomography (CT) was developed as a new reliable tool to determine the bone quality. The purpose of the present study was to demonstrate the utility of regional HU measurement of the screw trajectory to predict the primary and long-term fixation strength of pedicle screws. Method The insertional torque of pedicle screws using the cortical bone trajectory technique was measured intraoperatively in 92 consecutive patients who underwent single-level posterior lumbar interbody fusion. The cylindrical area of each screw was plotted on the preoperative CT image by precisely confirming the screw position, and the screw trajectory was measured in HU. First, three parameters: the bone mineral density (BMD) of the femoral neck and lumbar vertebrae, and regional HU values of the screw trajectory, were correlated with the insertional torque and compared among three groups. Next, pedicle screw loosening was evaluated by postoperative CT obtained 12 months after surgery, and clinical and imaging data were analyzed to assess whether regional HU values could be used as a predictor of screw loosening. Results Regional HU values of the screw trajectory (r = 0.75, p < 0.001) had stronger correlation with the insertional torque than the femoral BMD (r = 0.59, p < 0.001) and lumbar BMD (r = 0.55, p < 0.001). The incidence of screw loosening was 4.6% (16/351). Multivariate logistic regression analysis revealed that regional HU value (odds ratio = 0.70; 95% confidence interval = 0.56-0.84; p = 0.018) was an independent risk factor significantly affected screw loosening. Conclusions Regional HU values of the screw trajectory could be a strong predictor of both primary and long-term screw fixation in vivo.
引用
收藏
页码:405 / 411
页数:7
相关论文
共 50 条
  • [1] Regional Hounsfield unit measurement of screw trajectory for predicting pedicle screw fixation using cortical bone trajectory: a retrospective cohort study
    Keitaro Matsukawa
    Yuichiro Abe
    Yoshihide Yanai
    Yoshiyuki Yato
    Acta Neurochirurgica, 2018, 160 : 405 - 411
  • [2] Lumbar pedicle cortical bone trajectory screw
    Song Tengfei
    Hsu, Wellington K.
    Ye Tianwen
    CHINESE MEDICAL JOURNAL, 2014, 127 (21) : 3808 - 3813
  • [3] Minimally invasive lumbar pedicle screw fixation using cortical bone trajectory - Screw accuracy, complications, and learning curve in 100 screw placements
    Dayani, Fara
    Chen, Yi-Ren
    Johnson, Eli
    Deb, Sayantan
    Wu, Yunfen
    Lan Pham
    Singh, Harminder
    JOURNAL OF CLINICAL NEUROSCIENCE, 2019, 61 : 106 - 111
  • [4] A biomechanical comparison between cortical bone trajectory fixation and pedicle screw fixation
    Oshino, Hiroki
    Sakakibara, Toshihiko
    Inaba, Tadashi
    Yoshikawa, Takamasa
    Kato, Takaya
    Kasai, Yuichi
    JOURNAL OF ORTHOPAEDIC SURGERY AND RESEARCH, 2015, 10
  • [5] Morphometric Measurement of Cortical Bone Trajectory for Lumbar Pedicle Screw Insertion Using Computed Tomography
    Matsukawa, Keitaro
    Yato, Yoshiyuki
    Nemoto, Osamu
    Imabayashi, Hideaki
    Asazuma, Takashi
    Nemoto, Koichi
    JOURNAL OF SPINAL DISORDERS & TECHNIQUES, 2013, 26 (06): : E248 - E253
  • [6] Evaluation of the Fixation Strength of Pedicle Screws Using Cortical Bone Trajectory What Is the Ideal Trajectory for Optimal Fixation?
    Matsukawa, Keitaro
    Taguchi, Eiko
    Yato, Yoshiyuki
    Imabayashi, Hideaki
    Hosogane, Naobumi
    Asazuma, Takashi
    Nemoto, Koichi
    SPINE, 2015, 40 (15) : E873 - E878
  • [7] A biomechanical comparison between cortical bone trajectory fixation and pedicle screw fixation
    Hiroki Oshino
    Toshihiko Sakakibara
    Tadashi Inaba
    Takamasa Yoshikawa
    Takaya Kato
    Yuichi Kasai
    Journal of Orthopaedic Surgery and Research, 10
  • [8] Fatigue Performance of Cortical Bone Trajectory Screw Compared With Standard Trajectory Pedicle Screw
    Akpolat, Yusuf Tahiri
    Inceoglu, Serkan
    Kinne, Nolan
    Hunt, Devon
    Cheng, Wayne K.
    SPINE, 2016, 41 (06) : E335 - E341
  • [9] Radiographic feasibility study of cortical bone trajectory and traditional pedicle screw dual trajectories
    Mullin, Jeffrey P.
    Perlmutter, Breanna
    Schmidt, Eric
    Benzel, Edward
    Steinmetz, Michael P.
    JOURNAL OF NEUROSURGERY-SPINE, 2016, 25 (06) : 727 - 732
  • [10] Percutaneous cortical bone trajectory screw fixation versus traditional open pedicle screw fixation for type A thoracolumbar fractures without neurological deficit
    Zheng, Zhangan
    Zhang, Li
    Zhu, Yu
    Chen, Jun
    Zhang, Xiaohai
    Xia, Taibao
    Wu, Tianliang
    Quan, Liangzhong
    Zhao, Guangchao
    Ji, Xuelei
    Gui, Zhaoliu
    Xue, Shuangtao
    Yin, Zongsheng
    JOURNAL OF ROBOTIC SURGERY, 2023, 17 (01) : 233 - 241