Measurement of red blood cell aggregation in disposable capillary tubes

被引:15
作者
Baskurt, Oguz K. [1 ]
Uyuklu, Mehmet [2 ]
Ozdem, Sebahat [3 ]
Meiselman, Herbert J. [4 ]
机构
[1] Koc Univ, Sch Med, TR-34450 Istanbul, Turkey
[2] Akdeniz Univ, Fac Med, Dept Physiol, TR-07058 Antalya, Turkey
[3] Akdeniz Univ, Fac Med, Dept Biochem, TR-07058 Antalya, Turkey
[4] Univ So Calif, Keck Sch Med, Dept Physiol & Biophys, Los Angeles, CA 90033 USA
关键词
Acute phase response; disease activity; erythrocyte sedimentation rate; erythrocyte aggregation; ERYTHROCYTE SEDIMENTATION-RATE; SYSTEM; HEMATOCRIT; ANALYZER; DISEASE;
D O I
10.3233/CH-2011-1411
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
A new method is described in this paper that allows measurement of red blood cell (RBC) aggregation indexes in disposable glass tubes within minutes. Light transmission through the RBC suspension filled into a microhematocrit capillary at stasis is recorded during RBC aggregation; a novel method assures an initial dispersion of aggregates in the capillary. The resulting light transmittance-time data are analyzed to calculate various parameters. Measurement of erythrocyte sedimentation rate (ESR) and RBC aggregation using well established methods and the newly developed capillary tube aggregometer in blood samples with a wide range of RBC aggregation indicated significant correlations between these parameters. Additionally, light transmittance during complete disaggregation allows estimating hematocrit, thereby enabling hematocrit correction of the measured and calculated parameters. The newly developed capillary tube RBC aggregometer is suitable for use as a method to rapidly monitor disease activity and the acute phase response, especially at the point-of-care (e. g., health care facilities, physician's office) and for field studies.
引用
收藏
页码:295 / 305
页数:11
相关论文
共 35 条
[1]  
[Anonymous], REV PORT HEMORREOL
[2]  
[Anonymous], BR MED J
[3]  
[Anonymous], LYNCHS MED LAB TECHN
[4]  
Baskurt OK, 2007, INDIAN J EXP BIOL, V45, P25
[5]   Comparison of three instruments for measuring red blood cell aggregation [J].
Baskurt, Oguz K. ;
Uyuklu, Mehmet ;
Ulker, Pinar ;
Cengiz, Melike ;
Nemeth, Norbert ;
Alexy, Tamas ;
Shin, Sehyun ;
Hardeman, Max R. ;
Meiselman, Herbert J. .
CLINICAL HEMORHEOLOGY AND MICROCIRCULATION, 2009, 43 (04) :283-298
[6]   New guidelines for hemorheological laboratory techniques [J].
Baskurt, Oguz K. ;
Boynard, Michel ;
Cokelet, Giles C. ;
Connes, Philippe ;
Cooke, Brian M. ;
Forconi, Sandro ;
Liao, Fulong ;
Hardeman, Max R. ;
Jung, Friedrich ;
Meiselman, Herbert J. ;
Nash, Gerard ;
Nemeth, Norbert ;
Neu, Bjoern ;
Sandhagen, Bo ;
Shin, Sehyun ;
Thurston, George ;
Wautier, Jean Luc .
CLINICAL HEMORHEOLOGY AND MICROCIRCULATION, 2009, 42 (02) :75-97
[7]   In vivo correlates of altered blood rheology [J].
Baskurt, Oguz K. .
BIORHEOLOGY, 2008, 45 (06) :629-638
[8]  
Baskurt OK, 1998, CLIN HEMORHEOL MICRO, V19, P307
[9]  
BAUERSACHS RM, 1989, CLIN HEMORHEOL, V9, P1
[10]   Parameters of red blood cell aggregation as correlates of the inflammatory state [J].
Ben Ami, R ;
Barshtein, G ;
Zeltser, D ;
Goldberg, Y ;
Shapira, I ;
Roth, A ;
Keren, G ;
Miller, H ;
Prochorov, V ;
Eldor, A ;
Berliner, S ;
Yedgar, S .
AMERICAN JOURNAL OF PHYSIOLOGY-HEART AND CIRCULATORY PHYSIOLOGY, 2001, 280 (05) :H1982-H1988