Occlusion of the pilot tube in a laser-shield II endotracheal tube caused by methylene blue crystals: A case report

被引:2
作者
Hashimoto, T [1 ]
Armstead, VE [1 ]
机构
[1] Thomas Jefferson Univ, Jefferson Med Coll, Dept Anesthesiol, Philadelphia, PA 19107 USA
关键词
equipment and supplies : endotracheal tube; pilot balloon valve; extubation; intratracheal; complications; laser surgery;
D O I
10.1016/S0952-8180(98)00070-1
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
The Laser-Shield II (Xomed-Treace, Jacksonville, FL) endotracheal tube is equipped with methylene blue crystals to aid in the prompt detection of tube cuff rupture. As the tube cuff is being inflated with saline, the crystals are supposed to dissolve fully in the saline, forming a solution that is readily visible in the event of cuff rupture. We describe a case of occlusion of the pilot balloon tube caused by undissolved methylene blue crystals, a situation that made it impossible to deflate the Laser-Shield II tube cuff: Our experience may aid in future modification of endotracheal tube design. (C) 1998 by Elsevier Science Inc.
引用
收藏
页码:522 / 523
页数:2
相关论文
共 8 条
[1]  
BLANC VF, 1974, ANESTH ANALG, V53, P202
[2]   DIFFICULTY IN EXTUBATION - A CAUSE FOR CONCERN [J].
BROCKUTNE, JG ;
JAFFE, RA ;
ROBINS, B ;
RATNER, E .
ANAESTHESIA, 1992, 47 (03) :229-230
[3]  
DORSH JA, 1994, UNDERSTANDING ANESTH, P441
[4]   PROBLEMS ASSOCIATED WITH TRACHEAL EXTUBATION [J].
HARTLEY, M ;
VAUGHAN, RS .
BRITISH JOURNAL OF ANAESTHESIA, 1993, 71 (04) :561-568
[5]   HEAT SINK PROTECTION AGAINST LASERING ENDOTRACHEAL CUFFS [J].
LEJEUNE, FE ;
LETARD, F ;
GUICE, C ;
MARICE, H .
ANNALS OF OTOLOGY RHINOLOGY AND LARYNGOLOGY, 1982, 91 (06) :606-607
[6]  
RAMPIL IJ, 1992, ANESTH ANALG, V74, P424
[7]  
SOSIS MB, 1993, ANESTHESIOL CLIN N A, V11, P573
[8]   DIFFICULT EXTUBATION DUE TO A KINKED PILOT TUBE [J].
TANSKI, J ;
JAMES, RH .
ANAESTHESIA, 1986, 41 (10) :1060-1060