Transvenous Extraction Performance of Expanded Polytetrafluoroethylene Covered ICD Leads in Comparison to Traditional ICD Leads in Humans

被引:34
作者
Di Cori, Andrea [1 ]
Bongiorni, Maria Grazia [1 ]
Zucchelli, Giulio [1 ]
Segreti, Luca [1 ]
Viani, Stefano [1 ]
Paperini, Luca [1 ]
Soldati, Ezio [1 ]
机构
[1] Univ Pisa, Div Cardiovasc Dis 2, Santa Chiara Hosp, Pisa, Italy
来源
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY | 2010年 / 33卷 / 11期
关键词
ePTFE lead; ICD lead extraction; transvenous removal; mechanical dilatation; DEFIBRILLATOR LEADS; CORONARY-SINUS; PACEMAKERS;
D O I
10.1111/j.1540-8159.2010.02879.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Methods: Seventeen consecutive ICD recipients (ePTFE Group A, 16 men, mean age 66 +/- 12 years) with 17 Endotak Reliance G dual-coil ICD leads (mean implantation time 23 +/- 26 months) underwent a transvenous removal procedure. They were compared with two control groups, including 20 Sprint Quattro 6944 (non-ePTFE Group B; Medtronic Inc., Minneapolis, MN, USA) and 36 Riata 1570 ICD leads (non-ePTFE Group C; St. Jude Medical, St. Paul, USA). The indication for lead extraction was local infection in 35 patients (48%), sepsis in 24 patients (33%), and lead malfunction in 14 patients (19%). Results: In all groups, all leads were successfully and completely removed without major complications. Overall manual traction was effective in six patients (8%) and more effective in the ePTFE Group (29%) compared to Group B (0%) and Group C (3%) (P = 0.001). Sixty-seven leads (92%) required mechanical dilatation by the venous entry site approach, with a shorter extraction time in the ePTFE Group (5 +/- 11 min) compared to Group B (21 +/- 22 min) and Group C (16 +/- 22 min) (P = 0.003). ePTFE leads showed a lower rate of fibrotic adherences at the superior vena cava level (P = 0.01) without statistically significant differences in the other sites. Conclusions: ePTFE-covered leads may be removed more easily and quickly than non-ePTFE leads, requiring less frequently mechanical dilatation. (PACE 2010; 1376-1381).
引用
收藏
页码:1376 / 1381
页数:6
相关论文
共 14 条
[1]   Usefulness of mechanical transvenous dilation and location of areas of adherence in patients undergoing coronary sinus lead extraction [J].
Bongiorni, Maria Grazia ;
Zucchelli, Giulio ;
Soldati, Ezio ;
Arena, Giuseppe ;
Giannola, Gabriele ;
Di Cori, Andrea ;
Lapira, Federica ;
Bartoli, Chiara ;
Segreti, Luca ;
De Lucia, Raffaele ;
Barsotti, Antonio .
EUROPACE, 2007, 9 (01) :69-73
[2]   Transvenous removal of pacing and implantable cardiac defibrillating leads using single sheath mechanical dilatation and multiple venous approaches: high success rate and safety in more than 2000 leads [J].
Bongiorni, Maria Grazia ;
Soldati, Ezio ;
Zucchelli, Giulio ;
Di Cori, Andrea ;
Segreti, Luca ;
De Lucia, Raffaele ;
Solarino, Gianluca ;
Balbarini, Alberto ;
Marzilli, Mario ;
Mariani, Mario .
EUROPEAN HEART JOURNAL, 2008, 29 (23) :2886-2893
[3]   ICD leads:: Design and chronic dysfunctions [J].
Gradaus, R ;
Breithardt, G ;
Böcker, D .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2003, 26 (02) :649-657
[4]   The Implantable Cardioverter-Defibrillator Lead: Principles, Progress, and Promises [J].
Haqqani, Haris M. ;
Mond, Harry G. .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2009, 32 (10) :1336-1353
[5]   CARDIAC PATHOLOGY OF TRANSVENOUS PACEMAKERS [J].
HUANG, TY ;
BABA, N .
AMERICAN HEART JOURNAL, 1972, 83 (04) :469-&
[6]  
Koplan BA, 2008, PACE, V31, P47, DOI [10.1111/j.1540-8159.2007.00924.x, 10.1111/j.1540-8159.2008.01211.x]
[7]  
KOUHT AR, 2009, HEART RHYTHM, V6, pS132
[8]   Implantable transvenous cardioverter defibrillator leads: The dark side [J].
Lawton, JS ;
Wood, MA ;
Gilligan, DM ;
Stambler, BS ;
Damiano, RJ ;
Ellenbogen, KA .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1996, 19 (09) :1273-1278
[9]   AUTOPSY FINDINGS WITH PERMANENT PERVENOUS PACEMAKERS [J].
ROBBOY, SJ ;
HARTHORNE, JW ;
LEINBACH, RC ;
SANDERS, CA ;
AUSTEN, WG .
CIRCULATION, 1969, 39 (04) :495-+
[10]   5-YEARS EXPERIENCE WITH INTRAVASCULAR LEAD EXTRACTION [J].
SMITH, HJ ;
FEARNOT, NE ;
BYRD, CL ;
WILKOFF, BL ;
LOVE, CJ ;
SELLERS, TD .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1994, 17 (11) :2016-2020