A compact superconducting magnet for magnetic resonance microscopy

被引:2
作者
Crozier, S [1 ]
Doddrell, DM [1 ]
机构
[1] Univ Queensland, Ctr Magnet Resonance, Brisbane, Qld 4072, Australia
关键词
D O I
10.1063/1.1148680
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
Magnetic resonance microscopy (MRM) depends on the use of high field, superconducting magnet systems for its operation. The magnets that are conventionally used are those that were initially designed for chemical structural analysis work. A novel, compact magnet designed specifically for MRM is presented here, and while preserving high field, high homogeneity conditions, has a length less than one-third that of conventional systems. This enables much better access to samples, an important consideration in many MRM experiments. As the homogeneity of a magnet is strongly dependent on its length, novel geometries and optimization techniques are required to meet the requirements of MRM in a compact system. An important outcome of the stochastic optimization performed in this work, is that the use used of a thin superconducting solenoid surrounded by counterwound disk windings provides a mechanism for drastic length reductions over conventional magnet designs. (C) 1998 American Institute of Physics.
引用
收藏
页码:448 / 451
页数:4
相关论文
共 21 条
[1]  
Brechna H., 1973, Superconducting Magnet Systems
[2]   Optimization of transverse gradient coils with coaxial return paths by simulated annealing [J].
Buszko, ML ;
Kempka, MF ;
Szczesniak, E ;
Wang, DC ;
Andrew, ER .
JOURNAL OF MAGNETIC RESONANCE SERIES B, 1996, 112 (03) :207-213
[3]  
Callaghan P. T., 1991, PRINCIPLES NMR MICRO
[4]  
Corona A, 1987, ACM T MATH SOFTWARE, V13, P262
[5]  
Craik D., 1995, MAGNETISM PRINCIPLES
[6]   DESIGN OF SHIELDED QUADRUPOLAR GRADIENT COILS FOR MAGNETIC-RESONANCE MICROSCOPY BY SIMULATED ANNEALING [J].
CROZIER, S ;
DODD, S ;
DODDRELL, DM .
IEEE TRANSACTIONS ON MAGNETICS, 1994, 30 (03) :1242-1246
[7]   THE DESIGN OF TRANSVERSE GRADIENT COILS OF RESTRICTED LENGTH BY SIMULATED ANNEALING [J].
CROZIER, S ;
FORBES, LK ;
DODDRELL, DM .
JOURNAL OF MAGNETIC RESONANCE SERIES A, 1994, 107 (01) :126-128
[8]   GRADIENT-COIL DESIGN BY SIMULATED ANNEALING [J].
CROZIER, S ;
DODDRELL, DM .
JOURNAL OF MAGNETIC RESONANCE SERIES A, 1993, 103 (03) :354-357
[9]   A COMBINED STRATEGY FOR OPTIMIZATION IN NONLINEAR MAGNETIC PROBLEMS USING SIMULATED ANNEALING AND SEARCH TECHNIQUES [J].
DRAGO, G ;
MANELLA, A ;
NERVI, M ;
REPETTO, M ;
SECONDO, G .
IEEE TRANSACTIONS ON MAGNETICS, 1992, 28 (02) :1541-1544
[10]  
ERNST RR, 1987, ONE 2 DIMENSIONS