Cone beam CT based validation of neural network generated synthetic CTs for radiotherapy in the head region

被引:13
作者
Irmak, Sinan [1 ]
Zimmermann, Lukas [1 ,2 ,3 ]
Georg, Dietmar [1 ]
Kuess, Peter [1 ]
Lechner, Wolfgang [1 ]
机构
[1] Med Univ Vienna, Dept Radiat Oncol, Vienna, Austria
[2] Univ Appl Sci, Fac Engn, Wiener Neustadt, Austria
[3] Univ Appl Sci, Competence Ctr Preclin Imaging & Biomed Engn, Wiener, Wiener Neustadt, Austria
基金
奥地利科学基金会;
关键词
CBCT; MRI only; synthetic CT; DEFORMABLE IMAGE REGISTRATION; DOSE CALCULATION; COMPUTED-TOMOGRAPHY; ONLY RADIOTHERAPY; MRI; THERAPY; FEASIBILITY; CHALLENGES;
D O I
10.1002/mp.14987
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose In the past years, many different neural network-based conversion techniques for synthesizing computed tomographys (sCTs) from MR images have been published. While the model's performance can be checked during the training against the test set, test datasets can never represent the whole population. Conversion errors can still occur for special cases, for example, for unusual anatomical situations. Therefore, the performance of sCT conversion needs to be verified on a patient specific level, especially in the absence of a planning CT (pCT). In this study, the capability of cone-beam CTs (CBCTs) for the validation of sCTs generated by a neural network was investigated. Methods 41 patients with tumors in the head region were selected. 20 of them were used for model training and 10 for validation. Different implementations of CycleGAN (with/without identity and feature loss) were used to generate sCTs. The pixel (MAE, RMSE, PSNR) and geometric error (DICE, Sensitivity, Specificity) values were reported to identify the best model. VMAT plans were created for the remaining 11 patients on the pCTs. These plans were re-calculated on sCTs and CBCTs. An automatic density overriding method (CBCTRS) and a population-based dose calculation method (CBCTPop) were employed for CBCT-based dose calculation. The dose distributions were analysed using 3D global gamma analysis, applying a threshold of 10% with respect to the prescribed dose. Differences in DVH metrics for the PTV and the organs-at-risk were compared among the dose distributions based on pCTs, sCTs, and CBCTs. Results The best model was the CycleGAN without identity and feature matching loss. Including the identity loss led to a metric decrease of 10% for DICE and a metric increase of 20-60 HU for MAE. Using the 2%/2 mm gamma criterion and pCT as reference, the mean gamma pass rates were 99.0 +/- 0.4% for sCTs. Mean gamma pass rate values comparing pCT and CBCT were 99.0 +/- 0.8% and 99.1 +/- 0.8% for the CBCTRS and CBCTPop, respectively. The mean gamma pass rates comparing sCT and CBCT resulted in 98.4 +/- 1.6% and 99.2 +/- 0.6% for CBCTRS and CBCTPop, respectively. The differences between the gamma-pass-rates of the sCT and two CBCT-based methods were not significant. The majority of deviations of the investigated DVH metrices between sCTs and CBCTs were within 2%. Conclusion The dosimetric results demonstrate good agreement between sCT, CBCT, and pCT based calculations. A properly applied CBCT conversion method can serve as a tool for quality assurance procedures in an MR only radiotherapy workflow for head patients. Dosimetric deviations of DVH metrics between sCT and CBCTs of larger than 2% should be followed up. A systematic shift of approximately 1% should be taken into account when using the CBCTRS approach in an MR only workflow.
引用
收藏
页码:4560 / 4571
页数:12
相关论文
共 51 条
  • [1] Patch-based generation of a pseudo CT from conventional MRI sequences for MRI-only radiotherapy of the brain
    Andreasen, Daniel
    Van Leemput, Koen
    Hansen, Rasmus H.
    Andersen, Jon A. L.
    Edmund, Jens M.
    [J]. MEDICAL PHYSICS, 2015, 42 (04) : 1596 - 1605
  • [2] Feasibility of CBCT-based dose with a patient-specific stepwise HU-to-density curve to determine time of replanning
    Chen, Shifeng
    Le, Quynh
    Mutaf, Yildirim
    Lu, Wei
    Nichols, Elizabeth M.
    Yi, Byong Yong
    Leven, Tish
    Prado, Karl L.
    D'Souza, Warren D.
    [J]. JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2017, 18 (05): : 64 - 69
  • [3] Technical Note: U-net-generated synthetic CT images for magnetic resonance imaging-only prostate intensity-modulated radiation therapy treatment planning
    Chen, Shupeng
    Qin, An
    Zhou, Dingyi
    Yan, Di
    [J]. MEDICAL PHYSICS, 2018, 45 (12) : 5659 - 5665
  • [4] Magnetic resonance-guided radiation therapy: A review
    Chin, Stephen
    Eccles, Cynthia L.
    McWilliam, Alan
    Chuter, Robert
    Walker, Emma
    Whitehurst, Philip
    Berresford, Joseph
    Van Herk, Marcel
    Hoskin, Peter J.
    Choudhury, Ananya
    [J]. JOURNAL OF MEDICAL IMAGING AND RADIATION ONCOLOGY, 2020, 64 (01) : 163 - 177
  • [5] MR-guidance in clinical reality: current treatment challenges and future perspectives
    Corradini, S.
    Alongi, F.
    Andratschke, N.
    Belka, C.
    Boldrini, L.
    Cellini, F.
    Debus, J.
    Guckenberger, M.
    Hoerner-Rieber, J.
    Lagerwaard, F. J.
    Mazzola, R.
    Palacios, M. A.
    Philippens, M. E. P.
    Raaijmakers, C. P. J.
    Terhaard, C. H. J.
    Valentini, V.
    Niyazi, M.
    [J]. RADIATION ONCOLOGY, 2019, 14 (1)
  • [6] State of the art MRI in head and neck cancer
    Dai, Y. L.
    King, A. D.
    [J]. CLINICAL RADIOLOGY, 2018, 73 (01) : 45 - 59
  • [7] MR-Only Brain Radiation Therapy: Dosimetric Evaluation of Synthetic CTs Generated by a Dilated Convolutional Neural Network
    Dinkla, Anna M.
    Wolterink, Jelmer M.
    Maspero, Matteo
    Savenije, Mark H. F.
    Verhoeff, Joost J. C.
    Seravalli, Enrica
    Isgum, Ivana
    Seevinck, Peter R.
    van den Berg, Cornelis A. T.
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2018, 102 (04): : 801 - 812
  • [8] The Value of Magnetic Resonance Imaging for Radiotherapy Planning
    Dirix, Piet
    Haustermans, Karin
    Vandecaveye, Vincent
    [J]. SEMINARS IN RADIATION ONCOLOGY, 2014, 24 (03) : 151 - 159
  • [9] Comparison of CT number calibration techniques for CBCT-based dose calculation
    Dunlop, Alex
    McQuaid, Dualta
    Nill, Simeon
    Murray, Julia
    Poludniowski, Gavin
    Hansen, Vibeke N.
    Bhide, Shreerang
    Nutting, Christopher
    Harrington, Kevin
    Newbold, Kate
    Oelfke, Uwe
    [J]. STRAHLENTHERAPIE UND ONKOLOGIE, 2015, 191 (12) : 970 - 978
  • [10] A review of substitute CT generation for MRI-only radiation therapy
    Edmund, Jens M.
    Nyholm, Tufve
    [J]. RADIATION ONCOLOGY, 2017, 12