A Catalog of Magnetic Resonance Imaging Compatibility of Penile Prostheses

被引:6
作者
Lowe, Gregory [1 ,2 ]
Smith, Ryan P. [1 ]
Costabile, Raymond A. [1 ]
机构
[1] Univ Virginia Hlth Syst, Dept Urol, Charlottesville, VA 22908 USA
[2] Ohio State Univ, Med Ctr, Dept Urol, Columbus, OH 43210 USA
关键词
Erectile Dysfunction; Penile Prosthesis; Magnetic Resonance Imaging; Safety; Penile Implant; RESERVOIR; PAIN;
D O I
10.1111/j.1743-6109.2012.02689.x
中图分类号
R5 [内科学]; R69 [泌尿科学(泌尿生殖系疾病)];
学科分类号
1002 ; 100201 ;
摘要
Introduction. Erectile dysfunction has been successfully treated with penile prosthesis implantation for over 50 years. Ferromagnetic implants or devices may create a potentially hazardous or painful situation during magnetic resonance imaging (MRI). A modern catalog of the MRI compatibility of penile prostheses is not available. Aim. Evaluate the safety profile of implanted, penile prostheses during MRI. Main Outcome Measure. Review available in vitro safety data and reported patient complications experienced during MRI with a penile prosthesis in place. Methods. A search of PubMed (TM) for articles documenting a penile prosthesis present during MRI was performed. Radiology texts and product information from manufacturers' producing a penile prosthesis were reviewed. Direct discussion with product manufacturers was also performed to obtain additional safety and compatibility information. Results. Nine clinical articles noted the presence of a penile prosthesis at the time of magnetic resonance imaging. No articles documented a complication from MRI of a man with a penile prosthesis. A single patient with an unnamed malleable prosthesis was noted to have twisting of the device during MRI which did not result in discomfort or malfunction of the device. In vitro studies support the safety of most prostheses during MRI. Conclusion. The available data suggests there is little risk for most patients with a penile prosthesis who undergo MRI. Notable exceptions include Dacomed's Omniphase and Duraphase device and the MRI-conditional Spectra device manufactured by AMS. Current manufacturers of penile implants provide wallet cards and medical letters to support safety when undergoing an MRI. Prior prosthesis implantation should not preclude patients from having an MRI. Lowe G, Smith RP, and Costabile RA. A catalog of magnetic resonance imaging compatibility of penile prostheses. J Sex Med 2012;9:14821487.
引用
收藏
页码:1482 / 1487
页数:6
相关论文
共 17 条
[1]  
A. M. S, 2010, AMS 700 MS PUMP
[2]  
A. M. S, AMS MED SOL SPECTR C
[3]   Ejaculatory pain as a result of inflatable penile prosthesis reservoir compressing a seminal vesicle [J].
Agrawal, Vineet ;
Rickards, David ;
Ralph, David John .
UROLOGY, 2006, 68 (04) :888-888
[4]   Penile prosthesis reservoir - an imaging pitfall [J].
Brook, O. R. ;
Ghersin, E. ;
Guralnik, L. ;
Kaftori, J. K. ;
Amendola, M. A. ;
Engel, A. .
CLINICAL RADIOLOGY, 2008, 63 (03) :343-347
[5]   Successful conservative treatment of infected penile protheses [J].
Derouet, H ;
Uder, M ;
Freyfogle, EB ;
Stoeckle, M .
EUROPEAN UROLOGY, 2002, 41 (01) :66-70
[6]   Updates in inflatable penile prostheses [J].
Henry, Gerard D. ;
Wilson, Steven K. .
UROLOGIC CLINICS OF NORTH AMERICA, 2007, 34 (04) :535-+
[7]  
Henry GD, 2009, J SEX MED, V6, P675, DOI [10.1111/j.1743-6109.2008.01145.x, 10.1111/j.1743-6109.2008.01200.x]
[8]   NORMAL PENILE ANATOMY AND ABNORMAL PENILE CONDITIONS - EVALUATION WITH MR IMAGING [J].
HRICAK, H ;
MAROTTI, M ;
GILBERT, TJ ;
LUE, TF ;
WETZEL, LH ;
MCANINCH, JW ;
TANAGHO, EA .
RADIOLOGY, 1988, 169 (03) :683-690
[9]   SILICONE-SILVER PENILE PROSTHESIS - DESCRIPTION, OPERATIVE APPROACH AND RESULTS [J].
JONAS, U ;
JACOBI, GH .
JOURNAL OF UROLOGY, 1980, 123 (06) :865-867
[10]   MR imaging of nonmalignant penile lesions [J].
Kirkham, Alexander P. S. ;
Illing, Rowland O. ;
Minhas, Suks ;
Allen, Clare .
RADIOGRAPHICS, 2008, 28 (03) :837-853