On the use of the MLC dosimetric leaf gap as a quality control tool for accurate dynamic IMRT delivery

被引:30
作者
Mei, Xiangyang [1 ]
Nygren, Ian [1 ]
Villarreal-Barajas, J. Eduardo [1 ,2 ,3 ]
机构
[1] Tom Baker Canc Clin, Alberta Hlth Serv, Calgary, AB T2N 4N2, Canada
[2] Univ Calgary, Dept Radiat Oncol, Calgary, AB T2N 4N2, Canada
[3] Univ Calgary, Dept Phys & Astron, Calgary, AB T2N 1N4, Canada
关键词
dosimetric leaf gap; IMRT; EPID; dynamic MLC; INTENSITY-MODULATED RADIOTHERAPY; MULTILEAF COLLIMATION; POSITION ACCURACY; MODE;
D O I
10.1118/1.3567148
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: MLC leaf gap consistency is critical for the accurate delivery of dynamic IMRT plans. It is estimated that a systematic MLC leaf gap change of 0.6 mm will result in a 2% change to the equivalent uniform dose to a clinical target volume for a typical head and neck sliding window (SW) IMRT plan. The aim of this work is to use the measured dosimetric leaf gap (DLG) to verify the dosimetric reproducibility of dynamically delivered SW IMRT plans. This study focuses on Varian linacs equipped with the 120 Millennium MLC and the Eclipse T treatment planning system (TPS), but can be extended to other linac/MLC/TPS combination. Methods: An ionization chamber, a diode array, and an electronic portal imaging device (EPID) were used to assess the DLG in zero( central axis(, one, and two dimensions, respectively. The DLG for zero and two dimensions was derived from measurements of SW fields of decreasing width (2, 1.5, 1, and 0.5 cm). The DLG in one dimension was measured directly from a single SW sweeping across a linear diode array. This one- dimensional DLG measurement was based on the full width at half maximum( FWHM( of the dose rate versus time spectrum. Results: The DLG derived from ion chamber measurements at central axis agrees to within 0.1 mm, with the DLG measured directly from the FWHM of dose rate versus time spectrum. The measured DLG depends on the control points used for the MLC SW fields. When two control points were used, the DLG measured at central axis showed an increase of 0.6 mm with respect to the same measurements performed using three or more control points. The two- dimensional distribution of DLG obtained using the EPID identified leaf gap errors as small as +/- 0.2 mm in isolated areas away from central axis. Conclusions: Comprehensive measurements of the DLG in 0D, 1D, and 2D provide an accurate assessment of DLG value required during TPS commissioning. These DLG measurements can also be used as a quality control tool to quantify changes of the MLC calibration and leaf gap consistency, which is critical for the accurate delivery of dynamically delivered SW IMRT plans. (C) 2011 American Association of Physicists in Medicine. [DOI: 10.1118/1.3567148]
引用
收藏
页码:2246 / 2255
页数:10
相关论文
共 12 条
[1]   Requirements for leaf position accuracy for dynamic multileaf collimation [J].
Budgell, GJ ;
Mott, JHL ;
Williams, PC ;
Brown, KJ .
PHYSICS IN MEDICINE AND BIOLOGY, 2000, 45 (05) :1211-1227
[2]   Use of EPID for leaf position accuracy QA of dynamic multi-leaf collimator (DMLC) treatment [J].
Chang, J ;
Obcemea, CH ;
Sillanpaa, J ;
Mechalakos, J ;
Burman, C .
MEDICAL PHYSICS, 2004, 31 (07) :2091-2096
[3]   The sliding slit test for dynamic IMRT: a useful tool for adjustment of MLC related parameters [J].
Chauvet, I ;
Petitfils, A ;
Lehobey, C ;
Kristner, JY ;
Brunet, Y ;
Lembrez, R ;
Gaboriaud, G ;
Mazal, A ;
Zefkili, S ;
Rosenwald, JC .
PHYSICS IN MEDICINE AND BIOLOGY, 2005, 50 (04) :563-580
[4]   Testing of dynamic multileaf collimation [J].
Chui, CS ;
Spirou, S ;
LoSasso, T .
MEDICAL PHYSICS, 1996, 23 (05) :635-641
[5]   Guidance document on delivery, treatment planning, and clinical implementation of IMRT: Report of the IMRT subcommittee of the AAPM radiation therapy committee [J].
Ezzell, GA ;
Galvin, JM ;
Low, D ;
Palta, JR ;
Rosen, I ;
Sharpe, MB ;
Xia, P ;
Xiao, Y ;
Xing, L ;
Yu, CX .
MEDICAL PHYSICS, 2003, 30 (08) :2089-2115
[6]   IMRT commissioning: Multiple institution planning and dosimetry comparisons, a report from AAPM Task Group 119 [J].
Ezzell, Gary A. ;
Burmeister, Jay W. ;
Dogan, Nesrin ;
LoSasso, Thomas J. ;
Mechalakos, James G. ;
Mihailidis, Dimitris ;
Molineu, Andrea ;
Palta, Jatinder R. ;
Ramsey, Chester R. ;
Salter, Bill J. ;
Shi, Jie ;
Xia, Ping ;
Yue, Ning J. ;
Xiao, Ying .
MEDICAL PHYSICS, 2009, 36 (11) :5359-5373
[7]   Challenges in Credentialing Institutions and Participants in Advanced Technology Multi-institutional Clinical Trials [J].
Radiological Physics Center, The University of Texas M. D. Anderson Cancer Center, Houston, TX, United States .
Int. J. Radiat. Oncol. Biol. Phys., 2008, 1 SUPPL. (S71-S75) :S71-S75
[8]   Physical and dosimetric aspects of a multileaf collimation system used in the dynamic mode for implementing intensity modulated radiotherapy [J].
LoSasso, T ;
Chui, CS ;
Ling, CC .
MEDICAL PHYSICS, 1998, 25 (10) :1919-1927
[9]   Comprehensive quality assurance for the delivery of intensity modulated radiotherapy with a multileaf collimator used in the dynamic mode [J].
LoSasso, T ;
Chui, CS ;
Ling, CC .
MEDICAL PHYSICS, 2001, 28 (11) :2209-2219
[10]  
LOSASSO TJ, 2003, PRACTICAL GUIDE INTE, P123