Effect of inhalation flow rate on the dosing characteristics of dry powder inhaler (DPI) and metered dose inhaler (MDI) products

被引:63
作者
Ross, DL
Schultz, RK
机构
[1] Inhalation Technology, 3M Pharmaceuticals, 3M Center 270-4S-17, St. Paul
[2] Dura Pharmaceuticals, San Diego, CA 92121
[3] 3M Pharmaceuticals, Inhalation Technology, 3M Center 270-4S-17, St. Paul
来源
JOURNAL OF AEROSOL MEDICINE-DEPOSITION CLEARANCE AND EFFECTS IN THE LUNG | 1996年 / 9卷 / 02期
关键词
dry powder inhaler; DPI; metered dose inhaler; MDI; inhalation flow rate; dosing characteristics; particle size; medication delivery; respirable dose;
D O I
10.1089/jam.1996.9.215
中图分类号
R1 [预防医学、卫生学];
学科分类号
1004 ; 120402 ;
摘要
Both the dose delivered from the device and the particle size of the medication are important parameters for inhalation products because they influence the amount of drug that is delivered to the patient's lung. The inspiratory flow rate may vary from dose to dose in a given patient and between patients. The Marple-Miller Cascade Impactor, a new multistage inertial impactor that operates at two flow rates (30 and 60 liters/min) with comparable particle size cut-offs, provides a means to study the effect of inhalation flow rate on the particle size distributions of inhalation products. The medication delivery, mass median aerodynamic diameter (MMAD), and fine particle mass were determined, in a randomized fashion, for albuterol, beclomethasone, budesonide, and terbutaline in both metered dose inhaler (MDI) and dry powder inhaler (DPI) products as a function of flow rate. In all cases, independent of drug or device used, the MDI products had a more reproducible respirable dose than the breath-actuated DPI products tested as a function of inhalation flow rate.
引用
收藏
页码:215 / 226
页数:12
相关论文
共 20 条
  • [1] PULMONARY DEPOSITION OF INHALED TERBUTALINE - COMPARISON OF SCANNING GAMMA-CAMERA AND URINARY-EXCRETION METHODS
    BORGSTROM, L
    NEWMAN, S
    WEISZ, A
    MOREN, F
    [J]. JOURNAL OF PHARMACEUTICAL SCIENCES, 1992, 81 (08) : 753 - 755
  • [2] TOTAL AND REGIONAL LUNG DEPOSITION OF TERBUTALINE SULFATE INHALED VIA A PRESSURIZED MDI OR VIA TURBUHALER (R)
    BORGSTROM, L
    NEWMAN, S
    [J]. INTERNATIONAL JOURNAL OF PHARMACEUTICS, 1993, 97 (1-3) : 47 - 53
  • [3] LUNG DEPOSITION OF BUDESONIDE INHALED VIA TURBUHALER(R) - A COMPARISON WITH TERBUTALINE SULFATE IN NORMAL SUBJECTS
    BORGSTROM, L
    BONDESSON, E
    MOREN, F
    TROFAST, E
    NEWMAN, SP
    [J]. EUROPEAN RESPIRATORY JOURNAL, 1994, 7 (01) : 69 - 73
  • [4] COMPARISON OF INHALED ALBUTEROL POWDER AND AEROSOL IN ASTHMA
    BRONSKY, E
    BUCHOLTZ, GA
    BUSSE, WW
    CHERVINSKY, P
    CONDEMI, J
    GHAFOURI, MA
    HUDSON, L
    LAKSHMINARAYAN, S
    LOCKEY, R
    REESE, ME
    RENNARD, SI
    SEGAL, A
    SMOLLEY, L
    SPECTOR, S
    STABLEIN, JJ
    VANAS, A
    WILSON, A
    [J]. JOURNAL OF ALLERGY AND CLINICAL IMMUNOLOGY, 1987, 79 (05) : 741 - 747
  • [5] MODELING THE DEPOSITION OF INHALED POWDERED DRUG AEROSOLS
    CLARK, AR
    EGAN, M
    [J]. JOURNAL OF AEROSOL SCIENCE, 1994, 25 (01) : 175 - 186
  • [6] Dolovich MB, 1988, AM REV RESPIR DIS, V137, pA433
  • [7] CLINICAL COMPARISON OF INHALED BUDESONIDE DELIVERED EITHER VIA PRESSURIZED METERED DOSE INHALER OR TURBUHALER
    ENGEL, T
    HEINIG, JH
    MALLING, HJ
    SCHARLING, B
    NIKANDER, K
    MADSEN, F
    [J]. ALLERGY, 1989, 44 (03) : 220 - 225
  • [8] ENGEL T, 1990, EUR RESPIR J, V3, P1037
  • [9] EFFECTS, SIDE-EFFECTS AND PLASMA-CONCENTRATIONS OF TERBUTALINE IN ADULT ASTHMATICS AFTER INHALING FROM A DRY POWDER INHALER DEVICE AT DIFFERENT INHALATION FLOWS AND VOLUMES
    ENGEL, T
    SCHARLING, B
    SKOVSTED, B
    HEINIG, JH
    [J]. BRITISH JOURNAL OF CLINICAL PHARMACOLOGY, 1992, 33 (04) : 439 - 444
  • [10] COMPARISON OF SALBUTAMOL ROTAHALER WITH CONVENTIONAL PRESSURIZED AEROSOL
    HETZEL, MR
    CLARK, TJH
    [J]. CLINICAL ALLERGY, 1977, 7 (06): : 563 - 568