Accuracy of automated blood pressure recorders in pregnancy

被引:11
|
作者
Brown, MA [1 ]
Robinson, A [1 ]
Buddle, ML [1 ]
机构
[1] St George Hosp, Dept Renal Med, Kogarah, NSW 2217, Australia
关键词
D O I
10.1111/j.1479-828X.1998.tb03062.x
中图分类号
R71 [妇产科学];
学科分类号
100211 ;
摘要
Automated blood pressure recorders are used with increasing frequency by pregnant women, mostly without proper evaluation of their accuracy. We compared blood pressures (BP) recorded by 2 automated noninvasive devices, the Spacelabs 90207 ambulatory blood pressure monitor and the OMRON HEM 705 CP portable selfinitiated device, with blood pressures recorded by routine sphygmomanometry in 79 pregnant women either considered 'at risk' for preeclampsia or,with mild hypertension in pregnancy. The Spacelabs device tended to overestimate systolic BP by a mean 11 (SD=8) mmHg and diastolic BP by 5 (SD=7) mmHg for phase 5 pressure (p<0.001) but was similar to routine BPs for diastolic phase 4 pressures. The OMRON device tended to underestimate diastolic (phase 4) pressure by 4 (SD=6) mmHg (p<0.001) but gave similar systolic and diastolic (phase 5) pressures to routine sphygmomanometry. However, for both devices there was considerable individual patient variability in accuracy. When using these devices to record a limited number of blood pressure recordings, as in this study, we suggest that individual comparison with mercury sphygmomanometry be made in each pregnant woman before accepting the validity of these recordings.
引用
收藏
页码:262 / 265
页数:4
相关论文
共 50 条
  • [1] ASSESSING AUTOMATED BLOOD-PRESSURE RECORDERS
    FITZGERALD, DJ
    OMALLEY, K
    OBRIEN, ET
    LANCET, 1981, 2 (8253): : 993 - 993
  • [2] AUTOMATED BLOOD-PRESSURE MEASUREMENT - COMPARISON OF SEVERAL RECORDERS
    SPYCKERELLE, Y
    GUEGUEN, R
    ANDRE, JL
    DESCHAMPS, JP
    ARCHIVES DES MALADIES DU COEUR ET DES VAISSEAUX, 1988, 81 (08): : 997 - 1000
  • [3] A COMPARATIVE-STUDY OF AUTOMATED BLOOD-PRESSURE RECORDERS
    RUBIN, P
    MCLEAN, K
    REID, J
    POSTGRADUATE MEDICAL JOURNAL, 1980, 56 (662) : 815 - 817
  • [4] ACCURACY OF AN AUTOMATED ULTRASOUND BLOOD PRESSURE MONITOR
    HOCHBERG, HM
    SALOMON, H
    CURRENT THERAPEUTIC RESEARCH-CLINICAL AND EXPERIMENTAL, 1971, 13 (02): : 129 - &
  • [5] ASSESSMENT OF AMBULATORY BLOOD-PRESSURE RECORDERS - ACCURACY AND CLINICAL-PERFORMANCE
    WHITE, WB
    CLINICAL AND INVESTIGATIVE MEDICINE-MEDECINE CLINIQUE ET EXPERIMENTALE, 1991, 14 (03): : 202 - 211
  • [6] AUTOMATED BLOOD-PRESSURE RECORDING IN PREGNANCY
    BROWN, MA
    ADSETT, D
    CLINICAL AND EXPERIMENTAL HYPERTENSION PART B-HYPERTENSION IN PREGNANCY, 1991, 10 (1-2): : 7 - 19
  • [7] AUTOMATED BLOOD-PRESSURE MEASUREMENT IN PREGNANCY
    MARX, GF
    SCHWALBE, SS
    AMERICAN JOURNAL OF OBSTETRICS AND GYNECOLOGY, 1994, 170 (02) : 703 - 703
  • [8] Automated blood pressure monitoring in normal pregnancy
    Stella, A
    Grella, PV
    INTERNATIONAL JOURNAL OF GYNECOLOGY & OBSTETRICS, 1996, 55 (01) : 11 - 17
  • [10] VALIDATION OF AUTOMATED BLOOD-PRESSURE RECORDING IN PREGNANCY
    FRANX, A
    VANDERPOST, JAM
    ELFERING, IM
    VEERMAN, DP
    MERKUS, HMWM
    BOER, K
    VANMONTFRANS, GA
    BRITISH JOURNAL OF OBSTETRICS AND GYNAECOLOGY, 1994, 101 (01): : 66 - 69