In vitro optical performance of nonrotational symmetric and refractive-diffractive aspheric multifocal intraocular lenses: Impact of tilt and decentration

被引:55
作者
Montes-Mico, Robert [1 ]
Lopez-Gil, Norberto [2 ]
Perez-Vives, Cari [1 ]
Bonaque, Sergio [2 ]
Ferrer-Blasco, Teresa [1 ]
机构
[1] Univ Valencia, Dept Opt, Fac Phys, Optometry Res Grp, E-46100 Valencia, Spain
[2] Univ Murcia, Inst Univ Invest Envejecimineto, Murcia, Spain
关键词
SPHERICAL-ABERRATION; CONTRAST SENSITIVITY; VISUAL SIMULATION; PUPIL SIZE; DISTANCE; OUTCOMES; STABILITY; 1-PIECE; DESIGN; PIECE;
D O I
10.1016/j.jcrs.2012.03.040
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
PURPOSE: To evaluate the optical quality of Acrysof Restor SN6AD1 refractive-diffractive +3.00 D and Lentis Mplus LS-312 nonrotational symmetric +3.00 D aspheric multifocal intraocular lenses (IOLs) and analyze the effect of tilt and decentration. SETTING: University of Valencia, Valencia, and University of Murcia, Murcia, Spain. DESIGN: Experimental study. METHODS: An artificial model eye was used to measure wavefront aberrations of both IOL models. The modulation transfer function (MTF) and point-spread function (PSF) were calculated for a 5.0 mm pupil and under 5 situations: centered, 0.2 mm and 0.4 mm decentered, and 2 degrees and 4 degrees tilted. RESULTS: The refractive-diffractive IOL had the highest MTF values at all spatial frequencies. Similarly, the PSF was worse for the refractive-diffractive IOL. The MTFs for the refractive-diffractive IOL decreased when the IOL was decentered or tilted, being more robust for tilting than decentration. The MTFs for the nonrotational symmetric IOL were similar under all testing conditions. Cutoff frequencies for the refractive-diffractive IOL were stable over the range of tilting studied and comparable with the cutoff when the IOL was centered (approximately 50 cycles per degree [cpd]). When the IOL was decentered, the cutoff frequency decreased by approximately 40 cpd. The nonrotational symmetric IOL cutoff frequencies for misalignment and tilt were low (approximately 30 cpd) and similar under all conditions. CONCLUSIONS: The refractive-diffractive IOL provided better optical quality than the nonrotational symmetric IOL. Tilt and decentration had a significant impact on optical quality with both IOLs, being more severe with the nonrotational symmetric IOL.
引用
收藏
页码:1657 / 1663
页数:7
相关论文
共 30 条
  • [1] Visual function comparison of 2 aspheric multifocal intraocular lenses
    Alfonso, Jose F.
    Fernandez-Vega, Luis
    Blazquez, Jose I.
    Montes-Mico, Robert
    [J]. JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2012, 38 (02) : 242 - 248
  • [2] Comparative analysis of the clinical outcomes with 2 multifocal intraocular lens models with rotational asymmetry
    Alio, Jorge L.
    Plaza-Puche, Ana B.
    Pinero, David P.
    Javaloy, Jaime
    Jose Ayala, Maria
    [J]. JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2011, 37 (09) : 1605 - 1614
  • [3] Visual outcomes and optical performance of a monofocal intraocular lens and a new-generation multifocal intraocular lens
    Alio, Jorge L.
    Pinero, David P.
    Plaza-Puche, Ana B.
    Joanna Rodriguez, Maria
    [J]. JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2011, 37 (02) : 241 - 250
  • [4] DESIGN OF ASPHERIC INTRAOCULAR LENSES
    ATCHISON, DA
    [J]. OPHTHALMIC AND PHYSIOLOGICAL OPTICS, 1991, 11 (02) : 137 - 146
  • [5] Tilt and decentration of three-piece foldable high-refractive silicone and hydrophobic acrylic Intraocular lenses with 6-mm optics in an intraindividual comparison
    Baumeister, M
    Neidhardt, B
    Strobel, J
    Kohnen, T
    [J]. AMERICAN JOURNAL OF OPHTHALMOLOGY, 2005, 140 (06) : 1051 - 1058
  • [6] Problems in the measurement of wavefront aberration for eyes implanted with diffractive bifocal and multifocal intralocular lenses
    Charman, W. Neil
    Montes-Mico, Robert
    Radhakrishnan, Hema
    [J]. JOURNAL OF REFRACTIVE SURGERY, 2008, 24 (03) : 280 - 286
  • [7] Optical performance measurement and night driving simulation of ReSTOR, ReZoom, and Tecnis multifocal intraocular lenses in a model eye
    Choi, Junoh
    Schwiegerling, Jim
    [J]. JOURNAL OF REFRACTIVE SURGERY, 2008, 24 (03) : 218 - 222
  • [8] Visual outcomes after cataract surgery with implantation of a+3.00 D or+4.00 D aspheric diffractive multifocal intraocular lens: Comparative study
    de Vries, Niels E.
    Webers, Carroll A. B.
    Montes-Mico, Robert
    Ferrer-Blasco, Teresa
    Nuijts, Rudy M. M. A.
    [J]. JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2010, 36 (08) : 1316 - 1322
  • [9] Dietze HH, 2005, J REFRACT SURG, V21, pS541
  • [10] Effect of decentration and tilt on the image quality of aspheric intraocular lens designs in a model eye
    Eppig, Timo
    Scholz, Katja
    Loeffler, Andre
    Messner, Arthur
    Langenbucher, Achim
    [J]. JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2009, 35 (06) : 1091 - 1100