Implementation of biological variation-based analytical performance specifications in the laboratory: Stringent evaluation of Improvacutor blood collection tubes

被引:2
作者
Chung, Hee-Jung [1 ]
Song, Yoon Kyung [1 ]
Hong, Sung Kuk [1 ]
Hwang, Sang-Hyun [2 ]
Seo, Hee Seung [1 ]
Whang, Dong Hee [3 ]
Nam, Myung-Hyun [4 ]
Lee, Do Hoon [1 ]
机构
[1] Natl Canc Ctr, Dept Lab Med, Goyang, South Korea
[2] Univ Ulsan, Asan Med Ctr, Coll Med, Dept Lab Med, Seoul, South Korea
[3] Inje Univ, Seoul Paik Hosp, Coll Med, Dept Lab Med, Seoul, South Korea
[4] Korea Univ, Ansan Hosp, Coll Med, Dept Lab Med, Ansan, South Korea
来源
PLOS ONE | 2017年 / 12卷 / 12期
关键词
CLINICAL-CHEMISTRY; MEASUREMENT UNCERTAINTY; CRITERIA; QUALITY; GLUCOSE; ASSAYS;
D O I
10.1371/journal.pone.0189882
中图分类号
O [数理科学和化学]; P [天文学、地球科学]; Q [生物科学]; N [自然科学总论];
学科分类号
07 ; 0710 ; 09 ;
摘要
Recently, because the quality of laboratory analyses has increased along with the need for quality improvement, several external quality control bodies have adapted performance specifications using the Desirable Biological Variation Database, termed "Ricos goals"; these criteria are more stringent than those presented in CLIA 88. In this study, we aimed to validate newly introduced serum separator tubes, Improvacutor, for routine clinical chemistry testing in accordance with Ricos goals and CLIA 88. Blood samples were collected from 100 volunteers into three types of serum vacuum tubes: Greiner Vacuette, Becton Dickinson (BD) Vacutainer, and Improve Improvacutor. The samples were subjected to 16 routine chemistry tests using a TBA-200fr NEO chemistry autoanalyzer. In the comparison analysis, all 16 test results were acceptable according to CLIA 88. However, in the comparison of Improve and BD tubes, creatinine showed 4.31% (+0.08 mu mol/L) bias. This slightly exceeded the Desirable Specification for Inaccuracy Ricos limit of +/- 3.96%, but still satisfied the CLIS88 limit of +/- 26.52 mu mol/L. The remaining 15 analytes performed acceptably according to the Desirable Specifications of Ricos. The correlation coefficient of 12 analytes was greater than 0.95 in Passing-Bablok regression analysis among the three tubes, but was lower for four analytes: calcium, sodium, potassium, and chloride. In the stability assay, only potassium tested in the Greiner tube revealed a larger positive bias (2.18%) than the Ricos Desirable Specification for Inaccuracy based on biologic variation (1.8%). The BD tube also showed a positive bias of 1.74%, whereas the new Improve tube showed the smallest positive bias of 1.17% in potassium level after 72 h storage. Thus, the results of this study demonstrate that recently introduced analytical performance specifications based on components of biological variation (Rico's goal) could be extended to criterion for performance evaluation and applied.
引用
收藏
页数:10
相关论文
共 30 条
  • [1] [Anonymous], 2014, EP15A3 CLSI, V3rd
  • [2] [Anonymous], 1992, FED REG
  • [3] [Anonymous], 2010, GP44A4 CLSI
  • [4] [Anonymous], EP09 A3 Measurement Procedures Comparison and Bias Estimation Using Patient Sample
  • [5] Anraku H SY, 1989, U.S. Patent, Patent No. [8,565,334, 8565334]
  • [6] STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT
    BLAND, JM
    ALTMAN, DG
    [J]. LANCET, 1986, 1 (8476) : 307 - 310
  • [7] Interferences from blood collection tube components on clinical chemistry assays
    Bowen, Raffick A. R.
    Remaley, Alan T.
    [J]. BIOCHEMIA MEDICA, 2014, 24 (01) : 31 - 44
  • [8] Immunoassay interference by a commonly used blood collection tube additive, the organosilicone surfactant Silwet L-720
    Bowen, RAR
    Chan, Y
    Ruddel, ME
    Hortin, GL
    Csako, G
    Demosky, SJ
    Remaley, AT
    [J]. CLINICAL CHEMISTRY, 2005, 51 (10) : 1874 - 1882
  • [9] Effect of blood collection tubes on total triiodothyronine and other laboratory assays
    Bowen, RAR
    Chan, Y
    Cohen, J
    Rehak, NN
    Hortin, GL
    Csako, G
    Remaley, AT
    [J]. CLINICAL CHEMISTRY, 2005, 51 (02) : 424 - 433
  • [10] Advancements in blood collection devices
    Bush, VJ
    Leonard, L
    Szamosi, DI
    [J]. LABORATORY MEDICINE, 1998, 29 (10) : 616 - 622