Resolution of Untoward Nonauditory Stimulation and Cochlear Implant Performance Gain After Extraction of a Stainless Steel Dental Crown in a Patient With Cochlear Otosclerosis

被引:2
作者
Carlson, Matthew L. [1 ]
Gifford, Rene H. [1 ]
Shallop, Jon K. [1 ]
Driscoll, Colin L. W. [1 ]
机构
[1] Mayo Clin, Dept Otolaryngol Head & Neck Surg, Sch Med, Rochester, MN USA
关键词
Cochlear implant(s); Device malfunction; Otosclerosis; FACIAL-NERVE STIMULATION;
D O I
10.1097/MAO.0b013e3182382a94
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
Objective: To report a unique case of untoward nonauditory stimulation and cochlear implant performance deterioration after the placement of a stainless steel dental crown. Study Design: Clinical capsule report. Setting: Tertiary academic referral center. Patient: A 64-year-old female with a history of advanced otosclerosis and sequential bilateral cochlear implantation was evaluated at our institution for concerns of poor device performance. Approximately 3 years earlier, the patient had undergone stainless steel mandibular molar crown placement. After her dental procedure, she began experiencing worsening facial nerve stimulation, pain, and poor performance with the implant in her right ear. In-house device integrity testing demonstrated a series of phase reversals in the electrode voltages involving the mid-portion of the array that were consistent with otosclerosis and current shunting. Results: The patient underwent removal of the mandibular crown 5 years after initial placement and experienced diminished facial nerve stimulation, resolution of pain, and a statistically significant improvement in monosyllabic word scores and speech in noise testing (Consonant-Nucleus-Consonant, 36%-76%; Bamford-Kowal-Bench Speech-in-Noise, 15-10.5 dB SNR). These findings were reproducible with serial audiometric testing over the subsequent 1.5 years, and she subjectively reported a substantial and lasting improvement in sound quality. Conclusion: We present the first published report documenting resolution of nonauditory stimulation and objective device performance improvement after stainless steel crown extraction. Although we cannot make any definitive conclusions based on a single report, it remains plausible that dental work incorporating metal substrates may interfere with device function through participation in aberrant current shunting in patients with otosclerosis.
引用
收藏
页码:1455 / 1458
页数:4
相关论文
共 10 条
[1]  
[Anonymous], BKB SIN US MAN
[2]  
Broomfield S, 2000, Cochlear Implants Int, V1, P55, DOI 10.1179/cim.2000.1.1.55
[3]   Peripheral Facial Nerve Communications and Their Clinical Implications [J].
Diamond, Mark ;
Wartmann, Christopher T. ;
Tubbs, R. Shane ;
Shoja, Mohammadali M. ;
Cohen-Gadol, Aaron A. ;
Loukas, Marios .
CLINICAL ANATOMY, 2011, 24 (01) :10-18
[4]  
Kelsall DC, 1997, AM J OTOL, V18, P336
[5]   Cochlear implantation in cochlear otosclerosis [J].
Marshall, AH ;
Fanning, N ;
Symons, S ;
Shipp, D ;
Chen, JM ;
Nedzelski, JM .
LARYNGOSCOPE, 2005, 115 (10) :1728-1733
[6]  
Micco Alan Gerard, 2006, Curr Opin Otolaryngol Head Neck Surg, V14, P352, DOI 10.1097/01.moo.0000244195.04926.a0
[7]   Revision cochlear implantation for facial nerve stimulation in otosclerosis [J].
Polak, M ;
Ulubil, SA ;
Hodges, AV ;
Balkany, TJ .
ARCHIVES OF OTOLARYNGOLOGY-HEAD & NECK SURGERY, 2006, 132 (04) :398-404
[8]   Speech Perception after Cochlear Implantation in 53 Patients with Otosclerosis: Multicentre Results [J].
Rotteveel, Liselotte J. C. ;
Snik, Ad F. M. ;
Cooper, Huw ;
Mawman, Deborah J. ;
van Olphen, Adriaan F. ;
Mylanus, Emmanuel A. M. .
AUDIOLOGY AND NEURO-OTOLOGY, 2010, 15 (02) :128-136
[9]   CONVERGENCE OF CUTANEOUS, TOOTH-PULP, VISCERAL, NECK AND MUSCLE AFFERENTS ONTO NOCICEPTIVE AND NONNOCICEPTIVE NEURONS IN TRIGEMINAL SUBNUCLEUS CAUDALIS (MEDULLARY DORSAL HORN) AND ITS IMPLICATIONS FOR REFERRED PAIN [J].
SESSLE, BJ ;
HU, JW ;
AMANO, N ;
ZHONG, G .
PAIN, 1986, 27 (02) :219-235
[10]   SPEECH-DISCRIMINATION SCORES MODELED AS A BINOMIAL VARIABLE [J].
THORNTON, AR ;
RAFFIN, MJM .
JOURNAL OF SPEECH AND HEARING RESEARCH, 1978, 21 (03) :507-518