Reference interval for immature platelet fraction on Sysmex XN hematology analyzer: a comparison study with Sysmex XE-2100

被引:32
|
作者
Ko, Young Jin [2 ]
Hur, Mina [1 ]
Kim, Hanah [3 ]
Choi, Sang Gyeu [3 ]
Moon, Hee-Won [3 ]
Yun, Yeo-Min [3 ]
机构
[1] Konkuk Univ, Konkuk Univ Hosp, Sch Med, Dept Lab Med, Seoul 143729, South Korea
[2] Univ Ulsan, Coll Med, Asan Med Ctr, Dept Lab Med, Seoul, South Korea
[3] Konkuk Univ, Sch Med, Dept Lab Med, Seoul 143729, South Korea
关键词
immature platelet fraction; reference interval; Sysmex XN; UMBILICAL-CORD BLOOD; RETICULATED PLATELETS; PERFORMANCE EVALUATION; THROMBOELASTOMETRY; ESTABLISHMENT; PARAMETERS; RECOVERY; TERM;
D O I
10.1515/cclm-2014-0839
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Background: Recently introduced hematology analyzer, the Sysmex XN modular system (Sysmex, Kobe, Japan), has newly adopted a florescent channel to detect platelets and immature platelet fraction (IPF). This study aimed to establish new reference intervals for %-IPF and absolute number of IPF (A-IPF) on Sysmex XN. Platelet counts, %-IPF, and A-IPF were also compared between Sysmex XN and XE-2100 systems (Sysmex). Methods: Except outliers, blood samples from 2104 healthy individuals and 140 umbilical cord blood were analyzed using both Sysmex XN and XE-2100. The results of two systems were compared using Bland-Altman plot. The reference intervals for %-IPF and A-IPF were defined using non-parametric percentile methods according to the Clinical and Laboratory Standard Institute guideline (C28-A3). Results: The platelet counts, %-IPF, and A-IPF showed non-parametric distributions. The mean difference between Sysmex XN and XE-2100 in healthy individuals revealed a positive bias in platelets (+8.0x10(9)/L), %-IPF (+1.2%), and A-IPF (+3.0x10(9)/L). The reference intervals for %-IPF and A-IPF on Sysmex XN were: 1.0%-7.3% and 2.49-15.64x10(9)/L in healthy individuals; and 1.0%-4.4% and 2.94-12.82x10(9)/L in umbilical cord blood. Conclusions: This large-scale study demonstrates a clear difference of platelet counts and IPF between Sysmex XN and XE-2100. The new reference intervals for IPF on Sysmex XN would provide fundamental data for clinical practice and future research.
引用
收藏
页码:1091 / 1097
页数:7
相关论文
共 50 条
  • [1] Performance evaluation of Sysmex XN hematology analyzer in umbilical cord blood: a comparison study with Sysmex XE-2100
    Kim, Hanah
    Hur, Mina
    Choi, Sang-Gyeu
    Moon, Hee-Won
    Yun, Yeo-Min
    Hwang, Han Sung
    Kwon, Han Sung
    Sohn, In Sook
    CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2014, 52 (12) : 1771 - 1779
  • [2] Evaluation of the Sysmex XE-2100® hematology analyzer in hospital use
    Nakul-Aquaronne, D
    Sudaka-Sammarcelli, I
    Ferrero-Vacher, C
    Starck, B
    Bayle, J
    JOURNAL OF CLINICAL LABORATORY ANALYSIS, 2003, 17 (04) : 113 - 123
  • [3] Evaluation of the performance of sysmex XN-3100 automated hematology analyzer regarding the sysmex XE-2100 and microscopic examination
    Incir, Said
    Palaoglu, Kerim Erhan
    TURKISH JOURNAL OF BIOCHEMISTRY-TURK BIYOKIMYA DERGISI, 2021, 46 (01): : 27 - 35
  • [4] Design of Malaria Diagnostic Criteria for the Sysmex XE-2100 Hematology Analyzer
    Campuzano-Zuluaga, German
    Alvarez-Sanchez, Gonzalo
    Elcy Escobar-Gallo, Gloria
    Marina Valencia-Zuluaga, Luz
    Marcela Rios-Orrego, Alexandra
    Pabon-Vidal, Adriana
    Felipe Miranda-Arboleda, Andres
    Blair-Trujillo, Silvia
    Campuzano-Maya, German
    AMERICAN JOURNAL OF TROPICAL MEDICINE AND HYGIENE, 2010, 82 (03): : 402 - 411
  • [5] REFERENCE INTERVAL FOR IMMATURE PLATELET FRACTION IN SYSMEX XN MODULAR SYSTEM
    Kim, Hanah
    Ko, Young Jin
    Hur, Mina
    Choi, Sang Gyeu
    Moon, Hee-Won
    Yun, Yeo-Min
    INTERNATIONAL JOURNAL OF LABORATORY HEMATOLOGY, 2014, 36 : 84 - 84
  • [6] Optimization of Criteria for Verification of Automated Platelet Counts Generated by the Sysmex XE-2100 Hematology Analyzer
    Gulati, Gene
    Cote, Sean
    Behling, Eric
    Kocher, William
    LABMEDICINE, 2009, 40 (09): : 535 - 539
  • [7] Two cases of false platelet clumps flagged by the automated hematology analyzer Sysmex XE-2100
    Dai, Qingkai
    Zhang, Ge
    Lai, Chunqi
    Du, Zeli
    Chen, Lan
    Chen, Qi
    Peng, Leiwen
    Wang, Yuefang
    Yang, Hui
    Ye, Lei
    Yang, Xianjun
    Jiang, Yongmei
    CLINICA CHIMICA ACTA, 2014, 429 : 152 - 156
  • [8] Establishment of reference interval for immature platelet fraction using Sysmex XN9000 hematology analyzer in Anhui province in China
    Yang, Baixia
    Zhou, Yindi
    Ding, Shoufu
    Zha, Helin
    Zhang, Fan
    He, Wentao
    Xu, Yuanhong
    INTERNATIONAL JOURNAL OF CLINICAL AND EXPERIMENTAL MEDICINE, 2017, 10 (02): : 4214 - 4220
  • [9] Pseudoeosinophilia associated with malaria infection determined in the Sysmex XE-2100 hematology analyzer
    Huh, J
    Jung, JS
    Yoon, H
    Chung, WS
    ANNALS OF HEMATOLOGY, 2005, 84 (06) : 400 - 402
  • [10] Pseudoeosinophilia associated with malaria infection determined in the Sysmex XE-2100 hematology analyzer
    Jungwon Huh
    Junseop Jung
    Hyungdu Yoon
    Whasoon Chung
    Annals of Hematology, 2005, 84 : 400 - 402