A general framework and review of scatter correction methods in cone beam CT. Part 2: Scatter estimation approaches

被引:137
作者
Ruehrnschopf, Ernst-Peter [1 ]
Klingenbeck, Klaus [1 ]
机构
[1] Siemens AG, Healthcare Sect, Imaging & Therapy Div, D-91301 Forchheim, Germany
关键词
cone beam CT; image quality; scatter correction; scatter suppression; scatter kernels; scatter kernel superposition; scatter integral transform; Monte Carlo; X-RAY SCATTER; UNIFIED RECONSTRUCTION ALGORITHM; INVERSE MONTE-CARLO; GLARE CORRECTION; COHERENT SCATTER; SPREAD FUNCTION; RADIATION; MODEL; KERNEL; REDUCTION;
D O I
10.1118/1.3589140
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The main components of scatter correction procedures are scatter estimation and a scatter compensation algorithm. This paper completes a previous paper where a general framework for scatter compensation was presented under the prerequisite that a scatter estimation method is already available. In the current paper, the authors give a systematic review of the variety of scatter estimation approaches. Scatter estimation methods are based on measurements, mathematical-physical models, or combinations of both. For completeness they present an overview of measurement-based methods, but the main topic is the theoretically more demanding models, as analytical, Monte-Carlo, and hybrid models. Further classifications are 3D image-based and 2D projection-based approaches. The authors present a system-theoretic framework, which allows to proceed top-down from a general 3D formulation, by successive approximations, to efficient 2D approaches. A widely useful method is the beam-scatter-kernel superposition approach. Together with the review of standard methods, the authors discuss their limitations and how to take into account the issues of object dependency, spatial variance, deformation of scatter kernels, external and internal absorbers. Open questions for further investigations are indicated. Finally, the authors refer on some special issues and applications, such as bow-tie filter, offset detector, truncated data, and dual-source CT. (C) 2011 American Association of Physicists in Medicine. [DOI: 10.1118/1.3589140]
引用
收藏
页码:5186 / 5199
页数:14
相关论文
共 102 条
[1]   Measurement of scattered radiation in a volumetric 64-slice CT scanner using three experimental techniques [J].
Akbarzadeh, A. ;
Ay, M. R. ;
Ghadiri, H. ;
Sarkar, S. ;
Zaidi, H. .
PHYSICS IN MEDICINE AND BIOLOGY, 2010, 55 (08) :2269-2280
[2]  
[Anonymous], P SPIE
[3]  
[Anonymous], P SPIE MED IM C
[4]  
[Anonymous], 2004, Foundations of Image Science
[5]  
[Anonymous], P SPIE
[6]   A slice-by-slice blurring model and kernel evaluation using the Klein-Nishina formula for 3D scatter compensation in parallel and converging beam SPECT [J].
Bai, CY ;
Zeng, GSL ;
Gullberg, GT .
PHYSICS IN MEDICINE AND BIOLOGY, 2000, 45 (05) :1275-1307
[7]   Cone beam X-ray scatter removal via image frequency modulation and filtering [J].
Bani-Hashemi, A ;
Blanz, E ;
Maltz, J ;
Hristov, D ;
Svatos, M .
MEDICAL PHYSICS, 2005, 32 (06) :2093-2093
[8]  
Barrett H.H., 1981, RADIOLOGICAL IMAGING
[9]   Megalopinakophobia: Its symptoms and cures [J].
Barrett, HH ;
Myers, KJ ;
Gallas, B ;
Clarkson, E ;
Zhang, HB .
MEDICAL IMAGING 2001: PHYSICS OF MEDICAL IMAGING, 2001, 4320 :299-307
[10]   Scatter correction for flat detector cone-beam CT based on simulated sphere models [J].
Bertram, M. ;
Hohmann, S. ;
Wiegert, J. .