Adhesive-based liquid metal radio-frequency microcoil for magnetic resonance relaxometry measurement

被引:40
作者
Kong, Tian Fook [1 ,2 ]
Peng, Weng Kung [2 ]
Trung Dung Luong [1 ]
Nam-Trung Nguyen [1 ,2 ]
Han, Jongyoon [2 ,3 ,4 ]
机构
[1] Nanyang Technol Univ, Sch Mech & Aerosp Engn, Singapore 639798, Singapore
[2] Singapore MIT Alliance Res & Technol SMART Ctr, BioSyst & Micromech BioSyM IRG, Singapore 117543, Singapore
[3] MIT, Dept Elect Engn & Comp Engn, Cambridge, MA USA
[4] MIT, Dept Biol Engn, Cambridge, MA USA
关键词
NMR; BLOOD; CHIP;
D O I
10.1039/c1lc20853e
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
This paper reports the fabrication and characterization of an adhesive-based liquid-metal microcoil for magnetic resonance relaxometry (MRR). Conventionally, microcoils are fabricated by various techniques such as electroplating, microcontact printing and focused ion beam milling. These techniques require considerable fabrication efforts and incur high cost. In this paper, we demonstrate a novel technique to fabricate three-dimensional multilayer liquid-metal microcoils together with the microfluidic network by lamination of dry adhesive sheets. One of the unique features of the adhesive-based technique is that the detachable sample chamber can be disposed after each experiment and the microcoil can be reused without cross-contamination multiple times. The integrated microcoil has a low direct-current (DC) resistance of 0.3 Omega and a relatively high inductance of 67.5 nH leading to a high quality factor of approximately 30 at 21.65 MHz. The microcoil was characterized for similar to 0.5 T proton MRR measurements. The optimal pulse duration, amplitude, and frequency for the 90 degrees pulse were 131 mu s, -30 dB (1.56 W) and 21.6553 MHz, respectively. In addition, we used the liquid-metal microcoil to perform a parametric study on the transverse relaxation rate of human red blood cells at different hematocrit levels. The transverse relaxation rate increases quadratically with the hematocrit level. The results from the liquid-metal microcoil were verified by measurements with a conventional solenoid coil.
引用
收藏
页码:287 / 294
页数:8
相关论文
共 37 条
[1]   On-chip three dimensional microcoils for MRI at the microscale [J].
Badilita, Vlad ;
Kratt, Kai ;
Baxan, Nicoleta ;
Mohmmadzadeh, Mohammad ;
Burger, Tobias ;
Weber, Hans ;
v. Elverfeldt, Dominik ;
Hennig, Juergen ;
Korvink, Jan G. ;
Wallrabe, Ulrike .
LAB ON A CHIP, 2010, 10 (11) :1387-1390
[2]   Modeling and optimization of planar microcoils [J].
Beyzavi, Ali ;
Nguyen, Nam-Trung .
JOURNAL OF MICROMECHANICS AND MICROENGINEERING, 2008, 18 (09)
[3]  
Brown M.A., 2003, MRI BASIC PRINCIPLES, P21
[4]  
CASANOVA F, 2011, SINGLE SIDED NMR, P59
[5]   Microfluidic stretchable RF electronics [J].
Cheng, Shi ;
Wu, Zhigang .
LAB ON A CHIP, 2010, 10 (23) :3227-3234
[6]   A new magnetic bead-based, filterless bio-separator with planar electromagnet surfaces for integrated bio-detection systems [J].
Choi, JW ;
Ahn, CH ;
Bhansali, S ;
Henderson, HT .
SENSORS AND ACTUATORS B-CHEMICAL, 2000, 68 (1-3) :34-39
[7]   Magnetic bead handling on-chip: new opportunities for analytical applications [J].
Gijs, MAM .
MICROFLUIDICS AND NANOFLUIDICS, 2004, 1 (01) :22-40
[8]   Two-Layer Electroplated Microcoils With a PECVD Silicon Dioxide Interlayer Dielectric [J].
Golda, Dariusz ;
Lang, Jeffrey H. ;
Culpepper, Martin L. .
JOURNAL OF MICROELECTROMECHANICAL SYSTEMS, 2008, 17 (06) :1537-1545
[9]   The Optimal Hematocrit [J].
Harder, Louise ;
Boshkov, Lynn .
CRITICAL CARE CLINICS, 2010, 26 (02) :335-+
[10]   Micro-NMR for Rapid Molecular Analysis of Human Tumor Samples [J].
Haun, Jered B. ;
Castro, Cesar M. ;
Wang, Rui ;
Peterson, Vanessa M. ;
Marinelli, Brett S. ;
Lee, Hakho ;
Weissleder, Ralph .
SCIENCE TRANSLATIONAL MEDICINE, 2011, 3 (71)