Liquid ionization chambers for absorbed dose measurements in water at low dose rates and intermediate photon energies

被引:39
作者
Wickman, G [1 ]
Johansson, B
Bahar-Gogani, J
Holmstrom, T
Grindborg, JE
机构
[1] Umea Univ, Dept Radiat Phys, S-90185 Umea, Sweden
[2] Swedish Radiat Protect Inst, S-17116 Stockholm, Sweden
关键词
liquid ionization chamber; dosimetry in water; brachytherapy sources; intermediate photon energies;
D O I
10.1118/1.598268
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Two new liquid ionization chamber (LIC) designs, consisting of cylindrical and plane-parallel configurations, are presented. They are designed to be suitable for high-precision measurements of absorbed dose-to-water at dose rates and photon energies typical for LDR intermediate photon energy brachytherapy sources. The chambers have a sensitive liquid layer thickness of 1 mm and sensitive volumes of 7 mm(3) (plane-parallel) and 20 mm(3) (cylindrical). The liquids used as sensitive media in the chambers are either isooctane (C8H18), tetramethylsilane (Si(CH3)(4)) or mixtures of these two liquids in the approximate proportions 2 to 1. A chamber filled with such a liquid mixture and with a polarizing voltage of 300 V, provides a volume sensitivity of about 10(-9)C Gy(-1) mm(-3) for absorbed dose measurements in water in an x-ray radiation field with an effective photon energy of 120 keV. In the interval 30 to 140 keV, the relative change in sensitivity is less than +/-2.5%. The leakage current of the chambers is low and stable, which implies that absorbed dose measurements can be done with good reproducibility at dose-rates as low as 50 mu Gy min(-1) (sigma<3%). The long-term calibration stability was tested for a set of five chambers over a period of more than 1 year. No systematic change in their sensitivity could be observed. The general recombination at a polarizing voltage of 300 V is less than 2% for dose-rates up to about 100 mGy min(-1). The temperature dependence at room temperature is 0.5% per degrees C. The response is almost independent of the direction of the radiation for the plane-parallel LIC. (C) 1998 American Association of Physicists in Medicine. [S0094-2405(98)02406-7].
引用
收藏
页码:900 / 907
页数:8
相关论文
共 31 条
[1]  
Andreo P., 1987, ABSORBED DOSE DETERM
[2]  
[Anonymous], NISTIR5632 US DEP CO
[3]  
Burgemeister E. A., 1984, Radiation Protection Dosimetry, V6, P145
[4]   A GENERAL THEORY OF CAVITY IONISATION [J].
BURLIN, TE .
BRITISH JOURNAL OF RADIOLOGY, 1966, 39 (466) :727-&
[5]   A LIQUID IONIZATION-CHAMBER FOR NEUTRON DOSIMETRY [J].
CHU, JCH ;
GRANT, WH ;
ALMOND, PR .
PHYSICS IN MEDICINE AND BIOLOGY, 1980, 25 (06) :1133-1148
[6]   Accuracy of Monte Carlo photon transport simulation in characterizing brachytherapy dosimeter energy-response artefacts [J].
Das, RK ;
Li, Z ;
Perera, H ;
Williamson, JF .
PHYSICS IN MEDICINE AND BIOLOGY, 1996, 41 (06) :995-1006
[7]   A SECONDARY AIR KERMA STRENGTH STANDARD FOR YB-169 INTERSTITIAL BRACHYTHERAPY SOURCES [J].
DAS, RK ;
MISHRA, V ;
PERERA, H ;
MEIGOONI, AS ;
WILLIAMSON, JF .
PHYSICS IN MEDICINE AND BIOLOGY, 1995, 40 (05) :741-756
[8]  
Fluhs D, 1996, MED PHYS, V23, P427, DOI 10.1118/1.597736
[9]   SATURATION CHARACTERISTICS OF PARALLEL-PLATE IONIZATION CHAMBERS [J].
GREENING, JR .
PHYSICS IN MEDICINE AND BIOLOGY, 1964, 9 (02) :143-&
[10]   PHOTON MASS ATTENUATION AND MASS ENERGY-ABSORPTION COEFFICIENTS FOR H, C, N, O, AR, AND 7 MIXTURES FROM 0.1 KEV TO 20 MEV [J].
HUBBELL, JH .
RADIATION RESEARCH, 1977, 70 (01) :58-81