Intramedullary stabilization of extraarticular proximal tibial fractures: A biomechanical comparison of intramedullary and extramedullary implants including a new Proximal Tibia Nail (PTN)

被引:38
作者
Hansen, Matthias [1 ]
Mehler, Dorothea [2 ]
Hessmann, Martin Henri [3 ]
Blum, Jochen [1 ]
Rommens, Pol Maria [3 ]
机构
[1] City Hosp, Clin Trauma Hand & Reconstruct Surg, Worms, Germany
[2] Johannes Gutenberg Univ Mainz, Univ Hosp, Dept Trauma Surg, Biomech Res Lab, D-6500 Mainz, Germany
[3] Johannes Gutenberg Univ Mainz, Univ Hosp, Dept Trauma Surg, D-6500 Mainz, Germany
关键词
proximal fracture; tibia; intramedullary nail; double plate osteosynthesis; LISS; UTN; external fixator;
D O I
10.1097/BOT.0b013e31815aba5e
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Objectives: To determine in the laboratory whether there are or are not differences between individual geometrical designs of intramedullary and extramedullary devices used for the fixation of extraarticular proximal tibial fractures. Methods: Five devices were tested: a newly developed Proximal Tibia Nail (PTN), conventional double-plate osteosynthesis (DPO), the Less Invasive Stabilization System (LISS), an augmented Unreamed Tibial Nail with a T-stabilization-plate (UTN + TSP), and an external fixator (ExFix). A 10-mm defect osteotomy was performed on paired human tibiae, and the proximal and distal ends were potted in polymethylmethacrylate cement (PMMA). Each pair of bones was randomly stabilized with the new PTN in 1 tibia (Groups PTN I through PTN4) and in 1 of the 4 comparative implants in the corresponding contralateral bone. A biomechanical test of the bone implant construct was then performed with a vertical axial force of 350, 600, and 900 N, a bending moment of 6 Nm and a bidirectional rotational strain of 8 Nrn. Displacement of bone fragments was measured and depicted as a force-displacement diagram. Results: For axial loading, significant differences were seen between the PTN 2 group compared to the LISS group (P=0.016) and the PTN 4 group compared to the ExFix group (P=0.016). No statistically significant differences were seen for the PTN I group compared to the DPO group (P=0.125) and the PTN 3 group compared to the UTN + TSP group (P=0.453). The bending stiffness of the PTN 1 - 4 groups was not significantly different from any of the 4 alternative implants. There was comparable torsional stiffness in all implant groups except for the UTN + TSP group, which was less stable and significantly different from the PTN 3 group (P=0.016). Conclusions: Given the parameters of this investigation, the new PTN would theoretically provide the same mechanical stability as the DPO in axial loading. Higher stability in axial loading may be present when compared to the LISS or the ExFix. Further clinical investigation of this implant will determine its usefulness among proximal tibial fixation devices.
引用
收藏
页码:701 / 709
页数:9
相关论文
共 37 条
[1]  
BOLHOFNER BR, 1995, CLIN ORTHOP RELAT R, P75
[2]   A technique for intramedullary nailing of proximal third tibia fractures [J].
Buehler, KC ;
Green, J ;
Woll, TS ;
Duwelius, PJ .
JOURNAL OF ORTHOPAEDIC TRAUMA, 1997, 11 (03) :218-223
[3]  
BURSTEIN AH, 1997, BIOMECHANIK ORTHOPAD, P103
[4]  
Cole JD, 1998, TECH ORTHOP, V13, P27, DOI DOI 10.1097/00013611-199801000-00004
[5]   Force transfer between the plate and the bone: relative importance of the bending stiffness of the screws and the friction between plate and bone [J].
Cordey, J ;
Borgeaud, M ;
Perren, SM .
INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 2000, 31 :21-28
[6]   Treatment of high-energy tibial plateau fractures by the Ilizarov circular fixator [J].
Dendrinos, GK ;
Kontos, S ;
Katsenis, D ;
Dalas, A .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1996, 78B (05) :710-717
[7]   Mechanical conditions in the internal stabilization of proximal tibial defects [J].
Duda, GN ;
Mandruzzato, F ;
Heller, M ;
Kassi, JP ;
Khodadadyan, C ;
Haas, NP .
CLINICAL BIOMECHANICS, 2002, 17 (01) :64-72
[8]   Mechanical boundary conditions of fracture healing: borderline indications in the treatment of unreamed tibial nailing [J].
Duda, GN ;
Mandruzzato, F ;
Heller, M ;
Goldhahn, J ;
Moser, R ;
Hehli, M ;
Claes, L ;
Haas, NP .
JOURNAL OF BIOMECHANICS, 2001, 34 (05) :639-650
[9]  
FREEDMAN EL, 1995, CLIN ORTHOP RELAT R, P25
[10]   The development of the distal femur Less Invasive Stabilization System (LISS) [J].
Frigg, R ;
Appenzeller, A ;
Christensen, R ;
Frenk, A ;
Gilbert, S ;
Schavan, R .
INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 2001, 32 :24-31