Evaluating the Lower-Body Electromyogram Signal Acquired From the Feet As a Noise Reference for Standing Ballistocardiogram Measurements

被引:21
作者
Inan, Omer T. [1 ]
Kovacs, Gregory T. A. [2 ]
Giovangrandi, Laurent [1 ]
机构
[1] Stanford Univ, Dept Elect Engn, Stanford, CA 94305 USA
[2] DARPA, Microelect Technol Off, Arlington, VA 22203 USA
来源
IEEE TRANSACTIONS ON INFORMATION TECHNOLOGY IN BIOMEDICINE | 2010年 / 14卷 / 05期
关键词
Ballistocardiogram (BCG); electromyogram (EMG); motion artifacts; noninvasive cardiovascular monitoring; SNR estimation; WEIGHING SCALE; HEART-RATE;
D O I
10.1109/TITB.2010.2044185
中图分类号
TP [自动化技术、计算机技术];
学科分类号
0812 ;
摘要
The ballistocardiogram (BCG) is a measure of the reaction force of the body to cardiac ejection of blood. A variety of systems can be used for BCG detection, including beds, tables, chairs, and weighing scales. Weighing scales, in particular, have several practical advantages over the alternatives: low cost, small size, unobtrusiveness, and familiarity to the user; one disadvantage is that the subject must stand during the recording, rather than sit or lay supine, resulting in a higher susceptibility to motion artifacts in the measured signal. This paper evaluates the electromyogram (EMG) signal acquired from the feet of the subject during BCG recording as a noise reference for standing BCG measurements. As a subject moves while standing on the scale, muscle contractions in the feet are detected by the EMG signal, and used to flag segments of the BCG signal that are corrupted by elevated noise. For the purposes of evaluating this method, estimates of the BCG noise-to-signal ratio (NSR) were independently calculated with an ensemble average method, using the R-wave of a simultaneously-acquired chest ECG as a timing reference. The linear correlation between EMG power alone and BCG NSR from 14 subjects was found to be moderate (r = 0.58, F-statistic p-value < 0.05); combined with body-mass index (BMI), multiple linear regression yielded a stronger correlation (r = 0.73, F-statistic p-value = (1.01). Additionally, an example usage of the lower-leg EMG for improving BCG measurement robustness is provided.
引用
收藏
页码:1188 / 1196
页数:9
相关论文
共 20 条
  • [1] A NEW METHOD FOR LONG-TERM MONITORING OF THE BALLISTOCARDIOGRAM, HEART-RATE, AND RESPIRATION
    ALIHANKA, J
    VAAHTORANTA, K
    SAARIKIVI, I
    [J]. AMERICAN JOURNAL OF PHYSIOLOGY, 1981, 240 (05): : R384 - R392
  • [2] Brown H.R., 1952, CLIN BALLISTOCARDIOG
  • [3] CLINICAL EVALUATION OF THE BALLISTOCARDIOGRAM .2. HEART DISEASE - HYPERTENSION, ANGINA PECTORIS, AND MYOCARDIAL INFARCTION
    CHESKY, K
    MOSER, M
    TAYMOR, RC
    MASTER, AM
    PORDY, L
    [J]. AMERICAN HEART JOURNAL, 1951, 42 (03) : 328 - 333
  • [4] Heart rate detection from an electronic weighing scale
    Gonzalez-Landaeta, R.
    Casas, O.
    Pallas-Areny, R.
    [J]. PHYSIOLOGICAL MEASUREMENT, 2008, 29 (08) : 979 - 988
  • [5] BALLISTOCARDIOGRAPHY - AN APPRAISAL OF TECHNIC, PHYSIOLOGIC PRINCIPLES, AND CLINICAL VALUE
    GUBNER, RS
    RODSTEIN, M
    UNGERLEIDER, HE
    [J]. CIRCULATION, 1953, 7 (02) : 268 - 286
  • [6] Non-invasive cardiac output trending during exercise recovery on a bathroom-scale-based ballistocardiograph
    Inan, O. T.
    Etemadi, M.
    Paloma, A.
    Giovangrandi, L.
    Kovacs, G. T. A.
    [J]. PHYSIOLOGICAL MEASUREMENT, 2009, 30 (03) : 261 - 274
  • [7] Robust ballistocardiogram acquisition for home monitoring
    Inan, O. T.
    Etemadi, M.
    Wiard, R. M.
    Giovangrandi, L.
    Kovacs, G. T. A.
    [J]. PHYSIOLOGICAL MEASUREMENT, 2009, 30 (02) : 169 - 185
  • [8] INAN OT, P IEEE CBMS 2008, P70
  • [9] KOIVISTOINEN T, P IEEE EMBC 2004, P2026
  • [10] VALIDATION OF TETRAPOLAR BIOELECTRICAL IMPEDANCE METHOD TO ASSESS HUMAN-BODY COMPOSITION
    LUKASKI, HC
    BOLONCHUK, WW
    HALL, CB
    SIDERS, WA
    [J]. JOURNAL OF APPLIED PHYSIOLOGY, 1986, 60 (04) : 1327 - 1332