Heterophilic antibody interference in commercial immunoassays; a screening study using paired native and pre-blocked sera

被引:40
作者
Bolstad, Nils [1 ]
Warren, David J. [1 ]
Bjerner, Johan [2 ]
Kravdal, Gunnhild [3 ]
Schwettmann, Lutz [4 ]
Olsen, Kari H. [1 ]
Rustad, Pal [2 ]
Nustad, Kjell [1 ]
机构
[1] Oslo Univ Hosp, Radiumhosp, Dept Med Biochem, N-0310 Oslo, Norway
[2] Furst Med Lab, Oslo, Norway
[3] Akershus Univ Hosp, Div Diagnost & Technol, Unit Med Biochem, Lorenskog, Norway
[4] Alesund Hosp, Dept Med Biochem, Alesund, Norway
关键词
HAMAs; heterophile antibody; heterophilic; interference; immunoassay; IMMUNOMETRIC ASSAY; STIMULATING-HORMONE; PROSTATE-CANCER; IMMUNOGLOBULIN; SPECIFICITY; ANTIGEN; TESTS; IGG;
D O I
10.1515/CCLM.2011.702
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Background: Heterophilic antibodies are still an important source of interference in immunoassays. We have conducted a screening study for interference in a panel of commercially available assays using two sera known to contain high titer Fe-reactive heterophilic antibodies. Methods: The sera were distributed to laboratories participating in the Nordic External Quality Assessment cooperation (EQANord). Duplicate samples pre-blocked with aggregated murine monoclonal MAK33 were also supplied. Discrepancies (>50%) between the results for native and blocked samples were used to classify the tested assays as susceptible to interference. A total of 170 different assay kits covering 91 analytes were tested. Results: We found that 21 assays, covering 19 different analytes, were susceptible to interference from the heterophilic antibodies in the two sera. Many of these are clinically and commercially important assays. Some of the false results were grossly elevated and could have been detrimental to patient care in a clinical setting. Conclusions: Heterophilic antibodies with Fc-reactivity remain a threat. A more widespread use of antibody fragments and aggregated immunoglobulin could potentially improve the heterophilic antibody resistance of assays intended for clinical use.
引用
收藏
页码:2001 / 2006
页数:6
相关论文
共 30 条
[1]  
Bertholf RL, 2002, ANN CLIN LAB SCI, V32, P414
[2]   Human heterophilic antibodies display specificity for murine IgG subclasses [J].
Bjerner, J ;
Olsen, KH ;
Bormer, OP ;
Nustad, K .
CLINICAL BIOCHEMISTRY, 2005, 38 (05) :465-472
[3]   Testing and validating a homogeneous immunometric assay for interference [J].
Bjerner, J ;
Olsen, KH ;
Bormer, OP ;
Nustad, K .
CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2004, 42 (02) :208-214
[4]  
Bjerner J, 2002, CLIN CHEM, V48, P613
[6]  
BOSCATO LM, 1986, CLIN CHEM, V32, P1491
[7]  
BOSCATO LM, 1988, CLIN CHEM, V34, P27
[8]  
CSAKO G, 1988, CLIN CHEM, V34, P1481
[9]   Evaluation of heterophilic antibody blocking agents in reducing false positive interference in immunoassays for IL-17AA, IL-17FF, and IL-17AF [J].
DeForge, Laura E. ;
Loyet, Kelly M. ;
Delarosa, Donnie ;
Chinn, Jason ;
Zamanian, Fojan ;
Chuntharapai, Anan ;
Lee, James ;
Hass, Phil ;
Wei, Nathan ;
Townsend, Michael J. ;
Wang, Jianyong ;
Wong, Wai Lee T. .
JOURNAL OF IMMUNOLOGICAL METHODS, 2010, 362 (1-2) :70-81
[10]  
Després N, 1998, CLIN CHEM, V44, P440