Automated counting of white blood cells in synovial fluid

被引:41
作者
de Jonge, R
Brouwer, R
Smit, M
de Frankrijker-Merkestijn, M
Dolhain, RJEM
Hazes, JMW
van Toorenenbergen, AW
Lindemans, J
机构
[1] Univ Med Ctr Rotterdam, Erasmus MC, Dept Clin Chem, NL-3000 CA Rotterdam, Netherlands
[2] Univ Med Ctr Rotterdam, Erasmus MC, Dept Rheumatol, NL-3000 CA Rotterdam, Netherlands
关键词
synovial fluid; Sysmex XE-2100; automated; leucocytes;
D O I
10.1093/rheumatology/keh021
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Objectives. To evaluate the performance of automated leucocyte (white blood cell; WBC) counting by comparison with manual counting. Methods. The number of WBC was determined in heparinized synovial fluid samples by the use of (i) a standard urine cytometer (Kova) and a microscope (reference method) and (ii) a haematology analyser (Sysmex XE-2100; WBC/BASO and DIFF channels). Imprecision within and between days was determined by replicate analysis of a low (level A; WBC similar to0.560 x 10(9)/l) and a high (level B; WBC similar to1.081 x 10(9)/l) dedicated synovial fluid control (Quantimetrix). Results. The WBC count of the DIFF channel was highly correlated with the WBC count of the microscopic reference method (r = 0.99; WBC analyser = 0.870 x WBC reference method + 0.413). In contrast, no agreement existed between WBC counts generated by the WBC/BASO channel of the analyser and the reference method (r = 0.52; WBC analyser = 0.008 x WBC reference method + 0.079). Within-day imprecision (4 - 7%) and between-day imprecision (10%) of the haematology analyser were smaller than the within-day imprecision (12%) and the between-day imprecision (20 - 22%) of the manual reference method. For manual counting, inter-observer coefficients of variation were 35.9% (control level A) and 21.0% (control level B). Conclusions. The WBC count in synovial fluid can be reliably determined using the DIFF channel of the Sysmex XE-2100. Automated counting of WBC in synovial fluid offers more precise and faster results than manual counting.
引用
收藏
页码:170 / 173
页数:4
相关论文
共 15 条
[1]   The utilization of synovial fluid analysis in the UK [J].
Amer, H ;
Swan, A ;
Dieppe, P .
RHEUMATOLOGY, 2001, 40 (09) :1060-1063
[2]  
ATILOLA MAO, 1986, CAN J VET RES, V50, P282
[3]  
DIEPPE PA, 1979, Q J MED, V48, P533
[4]   DIAGNOSTIC-VALUE OF SYNOVIAL-FLUID MICROSCOPY - A REASSESSMENT AND RATIONALIZATION [J].
FREEMONT, AJ ;
DENTON, J ;
CHUCK, A ;
HOLT, PJL ;
DAVIES, M .
ANNALS OF THE RHEUMATIC DISEASES, 1991, 50 (02) :101-107
[5]   RAPID, SIMPLE DETECTION OF TRACE AMOUNTS OF SYNOVIAL-FLUID [J].
GOLDENBERG, DL ;
BRANDT, KD ;
COHEN, AS .
ARTHRITIS AND RHEUMATISM, 1973, 16 (04) :487-490
[6]   METHODS OF ASSESSING THE SYNOVIAL-FLUID CELL COUNT [J].
HASKARD, DO ;
REVELL, PA .
CLINICAL RHEUMATOLOGY, 1984, 3 (03) :319-322
[7]   VARIATION IN SYNOVIAL-FLUID ANALYSIS BY HOSPITAL LABORATORIES [J].
HASSELBACHER, P .
ARTHRITIS AND RHEUMATISM, 1987, 30 (06) :637-642
[8]  
Kwoh CK, 2002, ARTHRITIS RHEUM-US, V46, P328
[9]   Comparison of manual and automated cell counts in EDTA preserved synovial fluids. Storage has little influence on the results [J].
Salinas, M ;
Rosas, J ;
Iborra, J ;
Manero, H ;
Pascual, E .
ANNALS OF THE RHEUMATIC DISEASES, 1997, 56 (10) :622-626
[10]   REPRODUCIBILITY OF SYNOVIAL-FLUID ANALYSES - A STUDY AMONG 4 LABORATORIES [J].
SCHUMACHER, HR ;
SIECK, MS ;
ROTHFUSS, S ;
CLAYBURNE, GM ;
BAUMGARTEN, DF ;
MOCHAN, BS ;
KANT, JA .
ARTHRITIS AND RHEUMATISM, 1986, 29 (06) :770-774