High resolution imaging of the intracranial vessel wall at 3 and 7 T using 3D fast spin echo MRI

被引:65
|
作者
Zhu, Chengcheng [1 ]
Haraldsson, Henrik [1 ]
Tian, Bing [1 ]
Meisel, Karl [2 ]
Ko, Nerissa [2 ]
Lawton, Michael [3 ]
Grinstead, John [4 ]
Ahn, Sinyeob [4 ]
Laub, Gerhard [4 ]
Hess, Christopher [1 ,2 ]
Saloner, David [1 ,5 ]
机构
[1] Univ Calif San Francisco, Dept Radiol & Biomed Imaging, San Francisco, CA 94121 USA
[2] Univ Calif San Francisco, Dept Neurol, San Francisco, CA 94121 USA
[3] Univ Calif San Francisco, Dept Neurol Surg, San Francisco, CA 94121 USA
[4] Siemens Healthcare, San Francisco, CA USA
[5] VA Med Ctr, Serv Radiol, San Francisco, CA USA
关键词
Intracranial vessel wall; Atherosclerosis; Aneurysm; High resolution MRI; Contrast enhancement; CONTRAST-ENHANCED MRI; BLACK BLOOD MRI; CAROTID-ARTERY; TESLA; SUPPRESSION; ANEURYSMS; DISEASE;
D O I
10.1007/s10334-016-0531-x
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
High resolution MRI of the intracranial vessel wall provides important insights in the assessment of intracranial vascular disease. This study aims to refine high resolution 3D MRI techniques for intracranial vessel wall imaging at both 3 and 7 T using customized flip angle train design, and to explore their comparative abilities. 11 patients with intracranial artery disease (four atherosclerotic plaques, six aneurysms and one reversible cerebral vasoconstriction syndrome) were imaged at 3 and 7 T with a 3D T (1)-weighted fast-spin-echo sequence (SPACE) both pre and post Gd contrast injection. Wall to lumen contrast ratio (CRwall-lumen), contrast enhancement ratio (ER) and the sharpness of the vessel wall were quantified. Two experienced radiologists evaluated the image quality on a 0-5 scale. Both 3 and 7 T achieved good image quality with high resolution (nominal 0.5 mm isotropic) and whole brain coverage. The CRwall-lumen and the ER measurements were comparable (p > 0.05). The 7 T images were significantly sharper (sharpness: 2.69 +/- 0.50 vs. 1.88 +/- 0.53 mm(-1), p < 0.001) with higher image quality (reader 1 score: 3.5 +/- 1.1 vs. 2.4 +/- 1.1, p = 0.002) compared to 3 T. 3D T (1)-weighted SPACE can be used for intracranial vessel wall evaluation at both 3 and 7 T. 7 T provides significantly better image quality and improves the confidence of diagnosis.
引用
收藏
页码:559 / 570
页数:12
相关论文
共 50 条
  • [41] Usefulness of 3D High-resolution Vessel Wall MRI in Diffuse Nonaneurysmal SAH Patients
    Hye Na Jung
    Sang-il Suh
    Inseon Ryoo
    InSeong Kim
    Clinical Neuroradiology, 2021, 31 : 1071 - 1081
  • [42] Usefulness of 3D High-resolution Vessel Wall MRI in Diffuse Nonaneurysmal SAH Patients
    Jung, Hye Na
    Suh, Sang-il
    Ryoo, Inseon
    Kim, InSeong
    CLINICAL NEURORADIOLOGY, 2021, 31 (04) : 1071 - 1081
  • [43] Comparison of 3D Isotropic Resolution Fast Spin-Echo MRI with Conventional 2D MRI of Knee Ligaments and Menisci at 3.0T
    Subhas, N.
    Kao, A.
    Freire, M.
    Polster, J.
    Obuchowski, N.
    Winalski, C.
    AMERICAN JOURNAL OF ROENTGENOLOGY, 2010, 194 (05)
  • [44] High-resolution 3D myelin imaging using synthetic MRI
    Warntjes, M.
    Johansson, P.
    Lundberg, P.
    Tisell, A.
    MULTIPLE SCLEROSIS JOURNAL, 2019, 25 : 627 - 628
  • [45] Usefulness of 3D T1-Turbo Spin Echo Imaging for the Evaluation of Intracranial Stent Placement
    Mizuno, Hiroyuki
    Aihara, Masanori
    Sato, Koji
    Negishi, Chikashi
    Sasaguchi, Nobuo
    Kurihara, Hideyuki
    Yoshimoto, Yuhei
    JOURNAL OF NEUROENDOVASCULAR THERAPY, 2022, 17 (01) : 1 - 7
  • [46] Comparison of Gated and Ungated Black-Blood Fast Spin-echo MRI of Carotid Vessel Wall at 3T
    Zhu, Chengcheng
    Graves, Martin J.
    Sadat, Umar
    Young, Victoria E.
    Gillard, Jonathan H.
    Patterson, Andrew J.
    MAGNETIC RESONANCE IN MEDICAL SCIENCES, 2016, 15 (03) : 266 - 272
  • [47] Imaging Features of Susac Syndrome on High-Resolution Intracranial Vessel Wall MRI
    Yahyavi-Firouz-Abadi, Noushin
    Kiczek, Matthew
    Zeiler, Steven R.
    Wasserman, Bruce A.
    NEUROLOGY-NEUROIMMUNOLOGY & NEUROINFLAMMATION, 2021, 8 (04):
  • [48] Comparison of 3T Intracranial Vessel Wall MRI Sequences
    Lindenholz, A.
    Harteveld, A. A.
    Zwanenburg, J. J. M.
    Siero, J. C. W.
    Hendrikse, J.
    AMERICAN JOURNAL OF NEURORADIOLOGY, 2018, 39 (06) : 1112 - 1120
  • [49] 3D T2-weighted fast spin-echo MRI sialography of the parotid gland
    S. Sartoretti-Schefer
    S. Kollias
    W. Wichmann
    A. Valavanis
    Neuroradiology, 1999, 41 : 46 - 51
  • [50] 3D T2-weighted fast spin-echo MRI sialography of the parotid gland
    Sartoretti-Schefer, S
    Kollias, S
    Wichmann, W
    Valavanis, A
    NEURORADIOLOGY, 1999, 41 (01) : 46 - 51