Factors Influencing Lesion Detection in SPECT Lung Images

被引:0
|
作者
Gifford, H. C. [1 ]
Zheng, X. M. [2 ]
Licho, R. [1 ]
Pretorius, P. H. [1 ]
Schneider, P. B. [1 ]
Simkin, P. H. [1 ]
King, M. A. [1 ]
机构
[1] Univ Massachusetts, Sch Med, Dept Radiol, Worcester, MA 01605 USA
[2] Charles Sturt Univ, Sch Clin Sci, Wagga Wagga, NSW, Australia
来源
2006 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOL 1-6 | 2006年
关键词
D O I
10.1109/NSSMIC.2006.356429
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
An earlier localization ROC (LROC) study that found attenuation correction (AC) degraded the detection of solitary pulmonary nodules (SPN) in hybrid SPECT lung images had several potential shortcomings related to the simulation methods. We sought to address these issues with a revised LROC study. Clinical Tc-99m NeoTect scans acquired with a simultaneous transmission-emission protocol defined the normal cases in a single-slice LROC study. Abnormal cases contained a simulated 1-cm lung lesion. Four rescaled-block-iterative EM (RBI) reconstruction strategies applied: 1) AC, scatter correction (SC), and resolution compensation (RC); 2) AC only; 3) RC only; and 4) no corrections (NC). Images from these strategies underwent 3D Gaussian post-smoothing. Performances were defined by the average area under the LROC curve obtained from three human observers. The strategy ranking in order of decreasing performance was: 1) RBI with RC; 2) RBI with all corrections; 3) RBI with AC; and 4) RBI with no corrections. A multireader-multicase (MRMC) analysis only found significant patient and patient-strategy effects. The conflicting results concerning AC from this study and the previous one may revolve around lesion masking effects, which, by design, were not a factor in the current study.
引用
收藏
页码:2662 / 2666
页数:5
相关论文
共 50 条
  • [1] Investigation of iterative reconstruction strategies for lung lesion detection in SPECT
    Gifford, HC
    Zheng, XM
    Boening, G
    Bruyant, PP
    King, MA
    2004 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOLS 1-7, 2004, : 4241 - 4245
  • [2] 99mTc-tetrofosmin SPECT/CT in lung lesion detection
    Spanu, A.
    Chessa, F.
    Marongiu, P.
    Sanna, D.
    Piras, B.
    Pirina, P.
    Madeddu, G.
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2009, 36 : S362 - S362
  • [3] Influencing Factors on Abdomen Lesion Detection Using Electrical Impedance Tomography
    Yu, Hao
    Ren, Shangjie
    Dong, Feng
    Zhao, Shu
    2018 IEEE INTERNATIONAL INSTRUMENTATION AND MEASUREMENT TECHNOLOGY CONFERENCE (I2MTC): DISCOVERING NEW HORIZONS IN INSTRUMENTATION AND MEASUREMENT, 2018, : 330 - 335
  • [4] Automated Detection of Lesion Regions in Lung Computed Tomography Images: A Review
    Han G.-H.
    Liu X.-B.
    Zheng G.-Y.
    Zidonghua Xuebao/Acta Automatica Sinica, 2017, 43 (12): : 2071 - 2090
  • [5] Factors influencing uniformity in helical SPECT
    de Swart, J.
    Groen, H. C.
    Kooij, P. P. M.
    de Jong, M.
    Konijnenberg, M.
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2012, 39 : S368 - S369
  • [6] The usefulness of SPECT/CT in lung lesion diagnosis
    Spanu, Angela
    Chessa, Francesca
    Sanna, Daniela
    Nuvoli, Susanna
    Madeddu, Giuseppe
    JOURNAL OF NUCLEAR MEDICINE, 2010, 51
  • [7] A filter design for optimization of lesion detection in SPECT
    Kim, JH
    Kim, KI
    Kwark, CE
    1996 IEEE NUCLEAR SCIENCE SYMPOSIUM - CONFERENCE RECORD, VOLS 1-3, 1997, : 1683 - 1687
  • [8] An investigation of reconstruction strategies for mediastinal lesion detection using hybrid Ga-67SPECT images
    Pereira, N. F.
    Gifford, H. C.
    Farncombe, T. H.
    Smyczynski, M.
    King, M. A.
    2005 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOLS 1-5, 2005, : 1691 - 1694
  • [9] Using lesion masking to facilitate nonlinaear alignment of SPECT images
    Stamatakis, EA
    Wilson, JTL
    Wyper, DJ
    NEUROIMAGE, 2001, 13 (06) : S42 - S42
  • [10] Comparison of planar vs. SPECT images for lesion detectability
    Sorensen, ES
    Zeng, GL
    Holt, B
    Scott, K
    Karen, S
    Ovard, T
    2002 IEEE NUCLEAR SCIENCE SYMPOSIUM, CONFERENCE RECORD, VOLS 1-3, 2003, : 1731 - 1735