Evaluation and empirical analysis of an exact FBP algorithm for spiral cone-beam CT

被引:9
作者
Katsevich, A [1 ]
Lauritsch, G [1 ]
Bruder, H [1 ]
Flohr, T [1 ]
Stierstorfer, K [1 ]
机构
[1] Univ Cent Florida, Dept Math, Orlando, FL 32816 USA
来源
MEDICAL IMAGING 2003: IMAGE PROCESSING, PTS 1-3 | 2003年 / 5032卷
关键词
exact inversion; cone-beam data; spiral tomography; filtered back-projection algorithm;
D O I
10.1117/12.481348
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Recently one of the authors proposed a reconstruction algorithm, which is theoretically exact and has the truly shift-invariant filtering and backprojection structure. Each voxel is reconstructed using the theoretically minimum section of the spiral, which is located between the endpoints of the PI segment of the voxel. Filtering is one-dimensional, performed along lines with variable tilt on the detector, and consists of five terms. We will present evaluation of the performance of the algorithm. We will also discuss and illustrate empirically the contributions of the five filtering terms to the overall image. A thorough evaluation proved the validity of the algorithm. Excellent image results were achieved even for high pitch values. Overall image quality can be regarded as at least equivalent to the less efficient, exact, Radon-based methods. However, the new algorithm significantly increases efficiency. Thus, the method has the potential to be applied in clinical scanners of the future. The empirical analysis leads to a simple, intuitive understanding of the otherwise obscure terms of the algorithm. Identification and skipping of the practically irrelevant fifth term allows significant speed-up of the algorithm due to uniform distance weighting.
引用
收藏
页码:663 / 674
页数:12
相关论文
共 14 条
  • [1] Exact consideration of data redundancies for spiral cone-beam CT
    Lauritsch, G
    Katsevich, A
    Hirsch, M
    MEDICAL IMAGING 2004: IMAGE PROCESSING, PTS 1-3, 2004, 5370 : 2034 - 2045
  • [2] Microlocal analysis of an FBP algorithm for truncated spiral cone beam data
    Katsevich, A
    JOURNAL OF FOURIER ANALYSIS AND APPLICATIONS, 2002, 8 (05) : 407 - 425
  • [3] Microlocal Analysis of an FBP Algorithm for Truncated Spiral Cone Beam Data
    Alexander Katsevich
    Journal of Fourier Analysis and Applications, 2002, 8 : 407 - 426
  • [4] Exact and efficient cone-beam reconstruction algorithm for a short-scan circle combined with various lines
    Dennerlein, F
    Katsevich, A
    Lauritsch, G
    Hornegger, J
    MEDICAL IMAGING 2005: IMAGE PROCESSING, PT 1-3, 2005, 5747 : 388 - 399
  • [5] An improved cone-beam reconstruction algorithm for the circular orbit
    Hu, H
    SCANNING, 1996, 18 (08) : 572 - 581
  • [6] Orthogonal-rotating tetrahedral scanning for cone-beam CT
    Ye, Ivan B.
    Wang, Ge
    DEVELOPMENTS IN X-RAY TOMOGRAPHY VIII, 2012, 8506
  • [7] Dual helical cone-beam CT for inspecting large object
    Zou, Xiaobing
    Zeng, Li
    Li, Zongjian
    JOURNAL OF X-RAY SCIENCE AND TECHNOLOGY, 2009, 17 (03) : 233 - 251
  • [8] Theoretically exact filtered backprojection-type inversion algorithm for spiral CT
    Katsevich, A
    SIAM JOURNAL ON APPLIED MATHEMATICS, 2002, 62 (06) : 2012 - 2026
  • [9] BPF-Type Reconstruction for Dual Helical Cone-Beam CT
    Zeng, Li
    Zou, Xiaobing
    CURRENT MEDICAL IMAGING, 2011, 7 (02) : 125 - 135
  • [10] Katsevich-type reconstruction for dual helical cone-beam CT
    Zeng, Li
    Zou, Xiaobing
    JOURNAL OF X-RAY SCIENCE AND TECHNOLOGY, 2010, 18 (04) : 353 - 367