Measuring stable carbon and nitrogen isotopes by IRMS and 14C by AMS on samples with masses in the microgram range: Performances of the system installed at CEDAD-University of Salento

被引:13
作者
Maruccio, Lucio [1 ]
Quarta, Gianluca [1 ]
Braione, Eugenia [1 ]
Calcagnile, Lucio [1 ]
机构
[1] Univ Salento, Dept Math & Phys Ennio de Giorgi, CEDAD Ctr Dating & Diagnost, Via Monteroni, I-73100 Lecce, Italy
关键词
Carbon and nitrogen isotopes; Radiocarbon dating; Low mass samples; Accelerator mass spectrometry; ELEMENTAL ANALYZER; ION-SOURCE; RADIOCARBON;
D O I
10.1016/j.ijms.2017.05.014
中图分类号
O64 [物理化学(理论化学)、化学物理学]; O56 [分子物理学、原子物理学];
学科分类号
070203 ; 070304 ; 081704 ; 1406 ;
摘要
A new experimental set-up dedicated to the simultaneous measurements of stable carbon and nitrogen isotopic ratios by IRMS (Isotope Ratio Mass Spectrometry) and C-14 by AMS (Accelerator Mass Spectrometry) has been installed at CEDAD (Centre for Dating and Diagnostics)-University of Salento. The new set-up is formed by different functional blocks: an elemental analyzer, an IRMS spectrometer, a gas handling unit and a new hybrid sputtering ion source, connected to the AMS system. The general features of the new system are presented as well as the technical solutions adopted for its integration with the existing AMS system. Different tests were performed by analyzing samples of known isotopic composition and of different masses in order to define optimal operating conditions. In particular it is shown how radiocarbon measurements with precision of 0.6-0.7% and 1% are possible with the new system on samples of mass of the order of 20 pg and 10 pg, respectively. Background levels of the order of 10(-14) (C-14/C-12) have been also demonstrated. The new system has then significantly extended the experimental potential of the Centre in several research fields where the possibility to date very low mass samples is either a crucial advantage or mandatory such as in environmental sciences and cultural heritage dating. (C) 2017 Elsevier B.V. All rights reserved.
引用
收藏
页码:1 / 7
页数:7
相关论文
共 12 条
[1]   AixMICADAS, the accelerator mass spectrometer dedicated to 14C recently installed in Aix-en-Provence, France [J].
Bard, Edouard ;
Tuna, Thibaut ;
Fagault, Yoann ;
Bonvalot, Lise ;
Wacker, Lukas ;
Fahrni, Simon ;
Synal, Hans-Arno .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION B-BEAM INTERACTIONS WITH MATERIALS AND ATOMS, 2015, 361 :80-86
[2]   A new system for the simultaneous measurement of δ13C and δ15N by IRMS and radiocarbon by AMS on gaseous samples: Design features and performances of the gas handling interface [J].
Braione, Eugenia ;
Maruccio, Lucio ;
Quarta, Gianluca ;
D'Elia, Marisa ;
Calcagnile, Lucio .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION B-BEAM INTERACTIONS WITH MATERIALS AND ATOMS, 2015, 361 :387-391
[3]   The new AMS system at CEDAD for the analysis of 10Be, 26Al, 129I and actinides: Set-up and performances [J].
Calcagnile, Lucio ;
Quarta, Gianluca ;
Maruccio, Lucio ;
Synal, Hans-Arno ;
Mueller, Arnold Milenko .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION B-BEAM INTERACTIONS WITH MATERIALS AND ATOMS, 2015, 361 :100-104
[4]   High-resolution accelerator-based mass spectrometry: precision, accuracy and background [J].
Caleagnile, L ;
Quarta, G ;
D'Elia, M .
APPLIED RADIATION AND ISOTOPES, 2005, 62 (04) :623-629
[5]   Improving a gas ion source for 14C AMS [J].
Fahrni, S. M. ;
Wacker, L. ;
Synal, H-A ;
Szidat, S. .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION B-BEAM INTERACTIONS WITH MATERIALS AND ATOMS, 2013, 294 :320-327
[6]   Atmospheric Δ 14CO2 trend in Western European background air from 2000 to 2012 [J].
Levin, Ingeborg ;
Kromer, Bernd ;
Hammer, Samuel .
TELLUS SERIES B-CHEMICAL AND PHYSICAL METEOROLOGY, 2013, 65
[7]   Choice of dietary protein of vegetarians and omnivores is reflected in their hair protein 13C and 15N abundance [J].
Petzke, KJ ;
Boeing, H ;
Metges, CC .
RAPID COMMUNICATIONS IN MASS SPECTROMETRY, 2005, 19 (11) :1392-1400
[8]  
Reimer PJ, 2004, RADIOCARBON, V46, P1299
[9]   ON-LINE RADIOCARBON MEASUREMENTS OF SMALL SAMPLES USING ELEMENTAL ANALYZER AND MICADAS GAS ION SOURCE [J].
Ruff, M. ;
Fahrni, S. ;
Gaeggeler, H. W. ;
Hajdas, I. ;
Suter, M. ;
Synal, H-A ;
Szidat, S. ;
Wacker, L. .
RADIOCARBON, 2010, 52 (04) :1645-1656
[10]   Small sample Accelerator Mass Spectrometry for biomedical applications [J].
Salehpour, M. ;
Hakansson, K. ;
Possnert, G. .