Constant Temperature Drying Rate Tester: Real-Time Water Evaporation Measurement of Fabrics

被引:16
|
作者
Chau, Kam-Hong [1 ]
Tang, Ka-Po Maggie [1 ]
Kan, Chi-Wai [1 ]
机构
[1] Hong Kong Polytech Univ, Inst Text & Clothing, Hong Kong, Hong Kong, Peoples R China
关键词
Comfort; drying; evaporation; fabric; textile; water absorption; water transport; KNITTED FABRICS; SKIN TEMPERATURE; TRANSPORT TESTER; WOVEN FABRICS; WICKING; HEAT; EXERCISE; PERFORMANCE; BEHAVIOR; HOT;
D O I
10.1109/TIM.2018.2817400
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
A prototype of constant temperature drying rate (DRCT) tester (CTDRT) was built to evaluate the DRCT of fabrics. The sample platform is kept at constant temperature and was placed on the balance. The platform temperature is adjustable to simulate different skin temperatures. A heater was placed underneath the sample platform, but it did not have any contact with the sample platform. This novel noncontact heating system enables real-time measurement of water evaporation in the sample throughout the whole experiment. In addition to DRCT, comprehensive information such as water absorption, water spreading, and other fabric drying features can be obtained from CTDRT. In addition, negative pressure gradient system of CTDRT can help to maintain constant atmospheric conditions at negligible wind speed, thereby enhancing the stability and sensitivity of measurement. The DRCT, a key parameter of CTDRT, of 28 fabrics investigated is between 0.36 and 2.56 mL/h at the skin temperature of 37 degrees C. The validity and repeatability of CTDRT were also confirmed with statistical evidence. The testing duration for CTDRT ranges from 15 to 40 min, which is similar to those of other drying methods where heat is applied, but is much faster than the conventional nonheating methods.
引用
收藏
页码:2635 / 2648
页数:14
相关论文
共 50 条
  • [21] Real-time temperature measurement in stochastic rotation dynamics
    Fan, Rong
    Zachariah, Githin T.
    Padding, Johan T.
    Hartkamp, Remco
    PHYSICAL REVIEW E, 2021, 104 (03)
  • [22] Liquid crystal temperature measurement for real-time control
    Birrell, DC
    Eaton, JK
    APPLICATIONS OF DIGITAL IMAGE PROCESSING XXI, 1998, 3460 : 58 - 66
  • [23] Real-time temperature analogous measurement of gun charge
    Zhou, Yanhuang
    Chen, Guidong
    2003, Nanjing University of Science and Technology, Nanjing, China (15):
  • [24] Measurement of water evaporation rate for popcorning
    Chong, JIT
    Lam, DCC
    Tong, P
    INTERNATIONAL SYMPOSIUM ON ADVANCED PACKAGING MATERIALS: PROCESSES, PROPERTIES AND INTERFACES, PROCEEDINGS, 2000, : 131 - 134
  • [25] Real-Time Measurement of Substrate Water Retention and Loss
    Schulker, Brian
    Jackson, Brian
    Fonteno, William
    HORTSCIENCE, 2021, 56 (09) : S36 - S36
  • [26] A needle temperature microsensor for in vivo and real-time measurement of the temperature in acupoints
    Cui, RF
    Liu, JH
    Ma, WT
    Hu, HB
    Zhou, XD
    Li, HY
    Hu, JM
    SENSORS AND ACTUATORS A-PHYSICAL, 2005, 119 (01) : 128 - 132
  • [27] REAL-TIME MEASUREMENT OF LPE GROWTH-RATE IN GAP
    INATOMI, Y
    KURIBAYASHI, K
    JOURNAL OF CRYSTAL GROWTH, 1991, 114 (03) : 380 - 388
  • [28] Real-time measurement of diffusion exchange rate in biological tissue
    Williamson, Nathan H.
    Ravin, Rea
    Cai, Teddy X.
    Benjamini, Dan
    Falgairolle, Melanie
    O'Donovan, Michael J.
    Basser, Peter J.
    JOURNAL OF MAGNETIC RESONANCE, 2020, 317
  • [29] A real-time algorithm for assessing inhomogeneities in fabrics
    Weickert, J
    REAL-TIME IMAGING, 1999, 5 (01) : 15 - 22
  • [30] Study on the method of real-time measurement for ammunition charge temperature
    Liu, AM
    Li, XY
    Fan, YF
    Duan, JB
    Xue, CS
    ISTM/2001: 4TH INTERNATIONAL SYMPOSIUM ON TEST AND MEASUREMENT, VOLS 1 AND 2, CONFERENCE PROCEEDINGS, 2001, : 1553 - 1556