Adaptive non-local means on local principle neighborhood for noise/artifacts reduction in low-dose CT images

被引:11
作者
Zhang, Yuanke [1 ,2 ]
Lu, Hongbing [1 ]
Rong, Junyan [1 ]
Meng, Jing [2 ]
Shang, Junliang [2 ]
Ren, Pinghong [2 ]
Zhang, Junying [3 ]
机构
[1] Fourth Mil Med Univ, Sch Biomed Engn, Xian 710032, Shaanxi, Peoples R China
[2] Qufu Normal Univ, Sch Informat Sci & Engn, Rizhao 276826, Shandong, Peoples R China
[3] Xidian Univ, Sch Comp Sci & Technol, Xian 710071, Shaanxi, Peoples R China
基金
中国国家自然科学基金;
关键词
artifacts suppression; low-dose CT; nonlocal means (NLM); parameter estimation; principle component analysis (PCA); COMPUTED-TOMOGRAPHY; SINOGRAM RESTORATION; NOISE-REDUCTION; RECONSTRUCTION;
D O I
10.1002/mp.12388
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: Low-dose CT (LDCT) technique can reduce the x-ray radiation exposure to patients at the cost of degraded images with severe noise and artifacts. Non-local means (NLM) filtering has shown its potential in improving LDCT image quality. However, currently most NLM-based approaches employ a weighted average operation directly on all neighbor pixels with a fixed filtering parameter throughout the NLM filtering process, ignoring the non-stationary noise nature of LDCT images. In this paper, an adaptive NLM filtering scheme on local principle neighborhoods (PC-NLM) is proposed for structure-preserving noise/artifacts reduction in LDCT images. Methods: Instead of using neighboring patches directly, in the PC-NLM scheme, the principle component analysis (PCA) is first applied on local neighboring patches of the target patch to decompose the local patches into uncorrelated principle components (PCs), then a NLM filtering is used to regularize each PC of the target patch and finally the regularized components is transformed to get the target patch in image domain. Especially, in the NLM scheme, the filtering parameter is estimated adaptively from local noise level of the neighborhood as well as the signal-to-noise ratio (SNR) of the corresponding PC, which guarantees a "weaker" NLM filtering on PCs with higher SNR and a "stronger" filtering on PCs with lower SNR. The PC-NLM procedure is iteratively performed several times for better removal of the noise and artifacts, and an adaptive iteration strategy is developed to reduce the computational load by determining whether a patch should be processed or not in next round of the PC-NLM filtering. Results: The effectiveness of the presented PC-NLM algorithm is validated by experimental phantom studies and clinical studies. The results show that it can achieve promising gain over some state-of-the-art methods in terms of artifact suppression and structure preservation. Conclusions: With the use of PCA on local neighborhoods to extract principal structural components, as well as adaptive NLM filtering on PCs of the target patch using filtering parameter estimated based on the local noise level and corresponding SNR, the proposed PC-NLM method shows its efficacy in preserving fine anatomical structures and suppressing noise/artifacts in LDCT images. (C) 2017 American Association of Physicists in Medicine
引用
收藏
页码:E230 / E241
页数:12
相关论文
共 34 条
  • [1] Breast radiotherapy in the prone position primarily reduces the maximum out-of-field measured dose to the ipsilateral lung
    Becker, Stewart J.
    Elliston, Carl
    DeWyngaert, Keith
    Jozsef, Gabor
    Brenner, David
    Formenti, Silvia
    [J]. MEDICAL PHYSICS, 2012, 39 (05) : 2417 - 2423
  • [2] Wavelet Based Noise Reduction in CT-Images Using Correlation Analysis
    Borsdorf, Anja
    Raupach, Rainer
    Flohr, Thomas
    Hornegger, Joachim
    [J]. IEEE TRANSACTIONS ON MEDICAL IMAGING, 2008, 27 (12) : 1685 - 1703
  • [3] Current concepts - Computed tomography - An increasing source of radiation exposure
    Brenner, David J.
    Hall, Eric J.
    [J]. NEW ENGLAND JOURNAL OF MEDICINE, 2007, 357 (22) : 2277 - 2284
  • [4] Cancer Risks from CT Scans: Now We Have Data, What Next?
    Brenner, David J.
    Hall, Eric J.
    [J]. RADIOLOGY, 2012, 265 (02) : 330 - 331
  • [5] A review of image denoising algorithms, with a new one
    Buades, A
    Coll, B
    Morel, JM
    [J]. MULTISCALE MODELING & SIMULATION, 2005, 4 (02) : 490 - 530
  • [6] Thoracic low-dose CT image processing using an artifact suppressed large-scale nonlocal means
    Chen, Yang
    Yang, Zhou
    Hu, Yining
    Yang, Guanyu
    Zhu, Yongcheng
    Li, Yinsheng
    Luo, Limin
    Chen, Wufan
    Toumoulin, Christine
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2012, 57 (09) : 2667 - 2688
  • [7] Statistical image reconstruction for polyenergetic X-ray computed tomography
    Elbakri, IA
    Fessler, JA
    [J]. IEEE TRANSACTIONS ON MEDICAL IMAGING, 2002, 21 (02) : 89 - 99
  • [8] Inferring the eigenvalues of covariance matrices from limited, noisy data
    Everson, R
    Roberts, S
    [J]. IEEE TRANSACTIONS ON SIGNAL PROCESSING, 2000, 48 (07) : 2083 - 2091
  • [9] Adaptive streak artifact reduction in computed tomography resulting from excessive x-ray photon noise
    Hsieh, J
    [J]. MEDICAL PHYSICS, 1998, 25 (11) : 2139 - 2147
  • [10] Joliffe I.T., 1986, Principal Component Analysis