Correlation of Pulse Wave Transit Time with Pulmonary Artery Pressure in a Porcine Model of Pulmonary Hypertension

被引:10
作者
Mueller-Graf, Fabian [1 ,2 ]
Merz, Jonas [1 ]
Bandorf, Tim [1 ]
Albus, Chiara Felicitas [1 ]
Henkel, Maike [1 ]
Krukewitt, Lisa [1 ,3 ]
Kuehn, Volker [3 ]
Reuter, Susanne [2 ]
Vollmar, Brigitte [2 ]
Pulletz, Sven [1 ]
Boehm, Stephan H. [1 ]
Reuter, Daniel A. [1 ]
Zitzmann, Amelie [1 ]
机构
[1] Univ Med Ctr Rostock, Dept Anaesthesiol & Intens Care Med, Schillingallee 35, D-18057 Rostock, Germany
[2] Univ Med Ctr Rostock, Rudolf Zenker Inst Expt Surg, D-18057 Rostock, Germany
[3] Univ Rostock, Inst Commun Engn, Richard Wagner Str 31, D-18119 Rostock, Germany
关键词
pulmonary artery pressure (PAP); pulmonary hypertension (PH); pulse wave transit time (PWTT); pulse arrival time (PAT); pulse wave velocity (PWV); BLOOD-PRESSURE; PREEJECTION PERIOD; VELOCITY; IMPEDANCE; HYPOXIA; LABOR; INDEX; FLOW; MRI; EIT;
D O I
10.3390/biomedicines9091212
中图分类号
Q5 [生物化学]; Q7 [分子生物学];
学科分类号
071010 ; 081704 ;
摘要
For the non-invasive assessment of pulmonary artery pressure (PAP), surrogates like pulse wave transit time (PWTT) have been proposed. The aim of this study was to invasively validate for which kind of PAP (systolic, mean, or diastolic) PWTT is the best surrogate parameter. To assess both PWTT and PAP in six healthy pigs, two pulmonary artery Mikro-Tip (TM) catheters were inserted into the pulmonary vasculature at a fixed distance: one in the pulmonary artery trunk, and a second one in a distal segment of the pulmonary artery. PAP was raised using the thromboxane A2 analogue U46619 (TXA) and by hypoxic vasoconstriction. There was a negative linear correlation between PWTT and systolic PAP (r = 0.742), mean PAP (r = 0.712) and diastolic PAP (r = 0.609) under TXA. During hypoxic vasoconstriction, the correlation coefficients for systolic, mean, and diastolic PAP were consistently higher than for TXA-induced pulmonary hypertension (r = 0.809, 0.778 and 0.734, respectively). Estimation of sPAP, mPAP, and dPAP using PWTT is feasible, nevertheless slightly better correlation coefficients were detected for sPAP compared to dPAP. In this study we establish the physiological basis for future methods to obtain PAP by non-invasively measured PWTT.
引用
收藏
页数:13
相关论文
共 38 条
[1]   Clinical and Hemodynamic Correlates of Pulmonary Arterial Stiffness in Incident, Untreated Patients With Idiopathic Pulmonary Arterial Hypertension [J].
Chemla, Denis ;
Weatherald, Jason ;
Lau, Edmund M. T. ;
Savale, Laurent ;
Boucly, Athenals ;
Attal, Pierre ;
Jais, Xavier ;
Parent, Florence ;
Girerd, Barbara ;
Simonneau, Gerald ;
Montani, David ;
Humbert, Marc ;
Sitbon, Olivier ;
Herve, Philippe .
CHEST, 2018, 154 (04) :882-892
[2]   DETERMINATION OF PULSE-WAVE VELOCITIES WITH COMPUTERIZED ALGORITHMS [J].
CHIU, YC ;
ARAND, PW ;
SHROFF, SG ;
FELDMAN, T ;
CARROLL, JD .
AMERICAN HEART JOURNAL, 1991, 121 (05) :1460-1470
[3]  
DOIG JR, 1982, BRIT J OBSTET GYNAEC, V89, P145, DOI 10.1111/j.1471-0528.1982.tb04682.x
[4]   The ARRIVE guidelines 2.0: Updated guidelines for reporting animal research [J].
du Sert, Nathalie Percie ;
Hurst, Viki ;
Ahluwalia, Amrita ;
Alam, Sabina ;
Avey, Marc T. ;
Baker, Monya ;
Browne, William J. ;
Clark, Alejandra ;
Cuthill, Innes C. ;
Dirnagl, Ulrich ;
Emerson, Michael ;
Garner, Paul ;
Holgate, Stephen T. ;
Howells, David W. ;
Karp, Natasha A. ;
Lazic, Stanley E. ;
Lidster, Katie ;
MacCallum, Catriona J. ;
Macleod, Malcolm ;
Pearl, Esther J. ;
Petersen, Ole H. ;
Rawle, Frances ;
Reynolds, Penny ;
Rooney, Kieron ;
Sena, Emily S. ;
Silberberg, Shai D. ;
Steckler, Thomas ;
Wuerbel, Hanno .
PLOS BIOLOGY, 2020, 18 (07)
[5]   Is the meta-analysis of correlation coefficients accurate when population correlations vary? [J].
Field, AP .
PSYCHOLOGICAL METHODS, 2005, 10 (04) :444-467
[6]   2015 ESC/ERS Guidelines for the diagnosis and treatment of pulmonary hypertension The Joint Task Force for the Diagnosis and Treatment of Pulmonary Hypertension of the European Society of Cardiology (ESC) and the European Respiratory Society (ERS) Endorsed by: Association for European Paediatric and Congenital Cardiology (AEPC), International Society for Heart and Lung Transplantation (ISHLT) [J].
Galie, Nazzareno ;
Humbert, Marc ;
Vachiery, Jean-Luc ;
Gibbs, Simon ;
Lang, Irene ;
Torbicki, Adam ;
Simonneau, Gerald ;
Peacock, Andrew ;
Noordegraaf, Anton Vonk ;
Beghetti, Maurice ;
Ghofrani, Ardeschir ;
Gomez Sanchez, Miguel Angel ;
Hansmann, Georg ;
Klepetko, Walter ;
Lancellotti, Patrizio ;
Matucci, Marco ;
McDonagh, Theresa ;
Pierard, Luc A. ;
Trindade, Pedro T. ;
Zompatori, Maurizio ;
Hoeper, Marius .
EUROPEAN HEART JOURNAL, 2016, 37 (01) :67-+
[7]   PULSE TRANSIT-TIME AS AN INDICATOR OF ARTERIAL BLOOD-PRESSURE [J].
GEDDES, LA ;
VOELZ, MH ;
BABBS, CF ;
BOURLAND, JD ;
TACKER, WA .
PSYCHOPHYSIOLOGY, 1981, 18 (01) :71-74
[8]   Continuous blood pressure measurement by using the pulse transit time: comparison to a cuff-based method [J].
Gesche, Heiko ;
Grosskurth, Detlef ;
Kuechler, Gert ;
Patzak, Andreas .
EUROPEAN JOURNAL OF APPLIED PHYSIOLOGY, 2012, 112 (01) :309-315
[9]   Modeling of Acute Pulmonary Arterial Hypertension in Pigs Using a Stable Thromboxane A2 Analogue (U46619): Dose Adjustment and Assessment of Hemodynamic Reactions [J].
Goncharova, N. S. ;
Andreeva, E. M. ;
Vakhrushev, A. D. ;
Leandro, H. I. Condori ;
Murashova, L. A. ;
Voronin, S. E. ;
Korobchenko, L. E. ;
Mitrofanova, L. B. ;
Skorik, Y. A. ;
Galagudza, M. M. ;
Moiseeva, O. M. ;
Mikhaylov, E. N. .
BULLETIN OF EXPERIMENTAL BIOLOGY AND MEDICINE, 2021, 170 (06) :729-733
[10]   Novel Noninvasive Assessment of Pulmonary Arterial Stiffness Using Velocity Transfer Function [J].
Gupta, Ankur ;
Sharifov, Oleg F. ;
Lloyd, Steven G. ;
Tallaj, Jose A. ;
Aban, Inmaculada ;
Dell'italia, Louis J. ;
Denney, Thomas S., Jr. ;
Gupta, Himanshu .
JOURNAL OF THE AMERICAN HEART ASSOCIATION, 2018, 7 (18)