Optimisation of a current generation ICP-QMS and benchmarking against MC-ICP-MS spectrometry for the determination of lead isotope ratios in environmental samples

被引:6
|
作者
Usman, Abida [1 ,2 ]
Ander, E. Louise [2 ]
Bailey, Elizabeth H. [1 ]
Nelms, Simon [3 ]
Pashley, Vanessa [2 ]
Young, Scott D. [1 ]
Chenery, Simon R. [2 ]
机构
[1] Univ Nottingham, Sch Biosci, Nottingham, England
[2] British Geol Survey, Ctr Environm Geochem, Inorgan Geochem, Nottingham, England
[3] Thermo Fisher Sci, Paisley, Renfrew, Scotland
关键词
PLASMA-MASS-SPECTROMETRY; DETECTOR DEAD-TIME; PRECISION; PB; ACCURATE; QUADRUPOLE; URANIUM; MODEL; BIAS;
D O I
10.1039/c8ja00290h
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Novel ANOVA methodology was used to benchmark ICP-QMS against MC-ICP-MS for Pb isotope ratios, demonstrating "fitness-for-purpose" in environmental source apportionment. The precision and accuracy of lead (Pb) isotope measurements obtained from quadrupole-based mass spectrometers (ICP-QMS) are considered to be limited by a number of factors originating in different components of the instruments. In this study, experimental and instrumental protocols were optimised for determining lead isotope ratios in urban soil digests. Experimental measures included individual dilution of all samples and isotopic standards (SRM-981, NIST) to a single Pb concentration intended to produce an intensity which was high enough to negate blanks and interferences but low enough to ensure the detector operated only in pulse counting mode. Instrumental protocols included batch dead time correction, optimisation of dwell time and the number of scans employed and correction of mass discrimination by sequential application of both internal (Tl-203/Tl-205 ratio) and external (SRM-981, NIST) standards. This optimised methodology was benchmarked against multi-collector mass spectrometer (MC-ICP-MS) measurements of Pb isotope ratios using replicate digest solutions of the same soil; but after these had been subjected to Pb separation using an ion-exchange procedure. On the assumption that MC-ICP-MS measurements are more accurate, small additive and multiplicative differences were observed in only the 4th decimal place. ANOVA was used to compare the precisions of the two techniques demonstrating equal precisions c. 0.08% for Pb-207/Pb-206, suggesting a sample heterogeneity limitation. By contrast, for Pb-207/Pb-204, the worst-case ratio, ICP-QMS had a 10-fold poorer precision, despite negligible interference from Hg-204, implying an instrumental limitation. The study concludes that ICP-QMS can provide valuable source apportionment information for most Pb isotope ratios but further efforts should focus on improving assay of the Pb-207/Pb-204 ratio.
引用
收藏
页码:2184 / 2194
页数:11
相关论文
共 50 条
  • [2] High precision determination of chromium isotope ratios in geological samples by MC-ICP-MS
    Halicz, Ludwik
    Yang, Lu
    Teplyakov, Nataliya
    Burg, Avihu
    Sturgeon, Ralph
    Kolodny, Yehoshua
    JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2008, 23 (12) : 1622 - 1627
  • [3] Determination of iodine 129 by ICP-QMS in environmental samples
    Bienvenu, P
    Brochard, E
    Excoffier, E
    Piccione, M
    CANADIAN JOURNAL OF ANALYTICAL SCIENCES AND SPECTROSCOPY, 2004, 49 (06): : 423 - 428
  • [4] Determination of europium isotope ratios in natural waters by MC-ICP-MS
    Arantes de Carvalho, Gabriel Gustinelli
    Oliveira, Pedro Vitoriano
    Yang, Lu
    JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2017, 32 (05) : 987 - 995
  • [5] Direct measurement of lead isotope ratios in low concentration environmental samples by MC-ICP-MS and multi-ion counting
    Cocherie, Alain
    Robert, Michele
    CHEMICAL GEOLOGY, 2007, 243 (1-2) : 90 - 104
  • [6] Measurement of isotope ratios on transient signals by MC-ICP-MS
    Günther-Leopold, I
    Wernli, B
    Kopajtic, Z
    Günther, D
    ANALYTICAL AND BIOANALYTICAL CHEMISTRY, 2004, 378 (02) : 241 - 249
  • [7] Application of ICP-QMS for the determination of plutonium in environmental samples for safeguards purposes
    Godoy, Maria Luiza D. P.
    Godoy, Jose Marcus
    Roldao, Luiz Alfredo
    JOURNAL OF ENVIRONMENTAL RADIOACTIVITY, 2007, 97 (2-3) : 124 - 136
  • [8] Precise and accurate determination of lead isotope ratios in mining wastes by ICP-QMS as a tool to identify their source
    Margui, E.
    Iglesias, M.
    Queralt, I.
    Hidalgo, M.
    TALANTA, 2007, 73 (04) : 700 - 709
  • [9] High precision determination of silver isotope ratios in commercial products by MC-ICP-MS
    Yang, Lu
    Dabek-Zlotorzynska, Ewa
    Celo, Valbona
    JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2009, 24 (11) : 1564 - 1569
  • [10] Precise determination of lead isotope ratios by MC-ICP-MS without matrix separation exemplified by unique samples of diverse origin and history
    Karasinski, Jakub
    Bulska, Ewa
    Halicz, Ludwik
    Tupys, Andrii
    Wagner, Barbara
    JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2023, 38 (11) : 2468 - 2476