Adjustment of the lateral and longitudinal size of scanned proton beam spots using a pre-absorber to optimize penumbrae and delivery efficiency

被引:40
作者
Titt, Uwe [1 ]
Mirkovic, Dragan [1 ]
Sawakuchi, Gabriel O. [1 ]
Perles, Luis A. [1 ]
Newhauser, Wayne D. [1 ]
Taddei, Phillip J. [1 ]
Mohan, Radhe [1 ]
机构
[1] Univ Texas MD Anderson Canc Ctr, Dept Radiat Phys, Unit 94, Houston, TX 77030 USA
关键词
MONTE-CARLO SIMULATIONS; BRAGG PEAK; MULTIPLE COULOMB; MEDICAL-CENTER; RIDGE FILTER; PENCIL BEAMS; THERAPY; RADIOTHERAPY; DESIGN; SYSTEM;
D O I
10.1088/0031-9155/55/23/S10
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
In scanned-beam proton therapy, the beam spot properties, such as the lateral and longitudinal size and the minimum achievable range, are influenced by beam optics, scattering media and drift spaces in the treatment unit. Currently available spot scanning systems offer few options for adjusting these properties. We investigated a method for adjusting the lateral and longitudinal spot size that utilizes downstream plastic pre-absorbers located near a water phantom. The spot size adjustment was characterized using Monte Carlo simulations of a modified commercial scanned-beam treatment head. Our results revealed that the pre-absorbers can be used to reduce the lateral full width at half maximum (FWHM) of dose spots in water by up to 14 mm, and to increase the longitudinal extent from about 1 mm to 5 mm at residual ranges of 4 cm and less. A large factor in manipulating the lateral spot sizes is the drift space between the pre-absorber and the water phantom. Increasing the drift space from 0 cm to 15 cm leads to an increase in the lateral FWHM from 2.15 cm to 2.87 cm, at a water-equivalent depth of 1 cm. These findings suggest that this spot adjustment method may improve the quality of spot-scanned proton treatments.
引用
收藏
页码:7097 / 7106
页数:10
相关论文
共 27 条
[1]   Ridge filter design for proton therapy at Hyogo Ion Beam Medical Center [J].
Akagi, T ;
Higashi, A ;
Tsugami, H ;
Sakamoto, H ;
Masuda, Y ;
Hishikawa, Y .
PHYSICS IN MEDICINE AND BIOLOGY, 2003, 48 (22) :N301-N312
[2]  
[Anonymous], 2005, MCNPX VERSION 2 5 0
[3]   Therapeutic step and shoot proton beam spot-scanning with a multi-leaf collimator: A Monte Carlo study [J].
Bues, M ;
Newhauser, WD ;
Titt, U ;
Smith, AR .
RADIATION PROTECTION DOSIMETRY, 2005, 115 (1-4) :164-169
[4]  
CHU WT, 2006, COL OH ICRU IAEA M 1
[5]  
Coutrakon G, 2003, AIP CONF PROC, V680, P1116
[6]   MULTIPLE COULOMB SCATTERING OF 160 MEV PROTONS [J].
GOTTSCHALK, B ;
KOEHLER, AM ;
SCHNEIDER, RJ ;
SISTERSON, JM ;
WAGNER, MS .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION B-BEAM INTERACTIONS WITH MATERIALS AND ATOMS, 1993, 74 (04) :467-490
[7]   MAGNETIC SCANNING SYSTEM FOR HEAVY-ION THERAPY [J].
HABERER, T ;
BECHER, W ;
SCHARDT, D ;
KRAFT, G .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1993, 330 (1-2) :296-305
[8]   Spread-out Bragg peak and monitor units calculation with the Monte Carlo code MCNPX [J].
Herault, J. ;
Iborra, N. ;
Serrano, B. ;
Chauvel, P. .
MEDICAL PHYSICS, 2007, 34 (02) :680-688
[9]   SPOT SCANNING SYSTEM FOR PROTON RADIOTHERAPY [J].
KANAI, T ;
KAWACHI, K ;
KUMAMOTO, Y ;
OGAWA, H ;
YAMADA, T ;
MATSUZAWA, H .
MEDICAL PHYSICS, 1980, 7 (04) :365-369
[10]   Commissioning of a conformal irradiation system for heavy-ion radiotherapy using a layer-stacking method [J].
Kanai, Tatsuaki ;
Kanematsu, Nobuyuki ;
Minohara, Shinichi ;
Komori, Masataka ;
Torikoshi, Masami .
MEDICAL PHYSICS, 2006, 33 (08) :2989-2997