SLO power calibration

被引:0
作者
Nygaard, RW
Schuchard, RA
机构
[1] Univ Missouri, Dept Ophthalmol, Kansas City, MO 64110 USA
[2] VA Rehab R&D Ctr, Decatur, GA 30322 USA
[3] Emory Univ, Sch Med, Decatur, GA 30322 USA
关键词
calibration; computer laser; ophthalmoscopy; SLO;
D O I
暂无
中图分类号
R49 [康复医学];
学科分类号
100215 ;
摘要
We present a method for calibrating the Scanning Laser Ophthalmoscope (SLO) that predicts radiant power at any of 256 grayscale values (gsv) and 12 polarized filter (polarizer) levels. Predicted power values, p(gsv), were determined by substitution into polynomials linearly transformed to old or new power at p(0) and p(255). This was compared with observed power values at 125 levels of attenuation/session. Prediction accuracy was the proportion of nonsignificant pairwise comparisons (t-test, p=0.0001). We found that power transformation between polarizers and within sessions has both linear and nonlinear characteristics. Within polarizer and between sessions, however, power transformation has linear characteristics. A 5th-degree polynomial was individually fit, at each polarizer, to session I power distributions of 9 gsv steps (0, 31, 63, 95, 127, 159, 191, 223, 255). When adjusted to p(255) and p(0) in new sessions, we obtained p(gsv) that predicted power at 25 gsv * 5 polarizers for 18 days with an accuracy of about 0.84. When only adjusted to p(255), predictive accuracy was 0.81.
引用
收藏
页码:129 / 133
页数:5
相关论文
共 19 条
[1]  
CULHAM LE, 1992, OPHTHAL PHYSL OPT, V12, P281, DOI 10.1111/j.1475-1313.1992.tb00398.x
[2]  
CULHAM LE, 1993, CLIN VISION SCI, V8, P551
[3]   Preferred retinal loci relationship to macular scotomas in a low-vision population [J].
Fletcher, DC ;
Schuchard, RA .
OPHTHALMOLOGY, 1997, 104 (04) :632-638
[4]  
Fletcher DC, 1994, OPHTHALMOLOGY CLIN N, V7, P257
[5]   IS THERE A SYSTEMATIC LOCATION FOR THE PSEUDO-FOVEA IN PATIENTS WITH CENTRAL SCOTOMA [J].
GUEZ, JE ;
LEGARGASSON, JF ;
RIGAUDIERE, F ;
OREGAN, JK .
VISION RESEARCH, 1993, 33 (09) :1271-1279
[6]   Functional assessment of macular hole surgery by scanning laser ophthalmoscopy [J].
Guez, JE ;
Le Gargasson, JF ;
Massin, P ;
Rigaudière, F ;
Grall, Y ;
Gaudric, A .
OPHTHALMOLOGY, 1998, 105 (04) :694-699
[7]  
Lei H, 1997, INVEST OPHTH VIS SCI, V38, P1812
[8]  
Remky A, 1996, INVEST OPHTH VIS SCI, V37, P2350
[9]   LOW-VISION READING WITH SEQUENTIAL WORD PRESENTATION [J].
RUBIN, GS ;
TURANO, K .
VISION RESEARCH, 1994, 34 (13) :1723-1733
[10]  
SACHS L, 1984, APPL STAT, P136