Metal Neck and Liner Impingement in Ceramic Bearing Total Hip Arthroplasty

被引:21
作者
Lee, Young-Kyun [1 ]
Yoo, Jeong Loon [1 ]
Koo, Kyung-Hoi [1 ]
Yoon, Kang Sup [1 ]
Kim, Hee Jong [1 ]
机构
[1] Seoul Natl Univ, Coll Med, Dept Orthopaed Surg, Seoul, South Korea
关键词
total hip arthroplasty; impingement; ceramic liner; microcrack; failure; ACETABULAR COMPONENTS; EARLY OSTEOLYSIS; REPLACEMENT; HEAD; POLYETHYLENE; FRACTURE; WEAR; PROSTHESIS; STABILITY; SIMULATOR;
D O I
10.1002/jor.21246
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Although impingement between the neck of the metallic stem and the ceramic liner has been suspected to be the cause of ceramic liner failure in ceramic-on-ceramic total hip arthroplasty (THA), no report has directly demonstrated microscopic damage on ceramic liner. We performed 18 reoperations on 18 patients who had undergone third generation ceramic-on-ceramic THA. Considering impingement, 16 patients, who were reoperated more than 1 year after previous ceramic bearing THA, were evaluated. Retrieved alumina liners, showing evidence of impingement, were examined by means of visual inspection and scanning electron microscopy (SEM). Four of the 16 hips showed neck notching and black stained liners, evidence of metallic neck to ceramic impingement. Impinged alumina bearings had been implanted for an average of 62.5 months (range: 35-99 months) before reoperation. SEM of the black stained area demonstrated disruptive wear and loss of surface integrity. Furthermore, one liner had multiple microcracks, and its cross-sectional SEM analysis revealed one microcrack propagating into the deep portion of the ceramic liner. Our observations suggest that metal neck-to-ceramic impingement in ceramic-on-ceramic THA can cause microcrack formation in ceramic liner. (C) 2010 Orthopaedic Research Society. Published by Wiley Periodicals, Inc. J Orthop Res 29:218-222, 2011
引用
收藏
页码:218 / 222
页数:5
相关论文
共 34 条
  • [1] [Anonymous], 1920, The phenomena of Rupture and Flow in Solids
  • [2] The influence of head and neck geometry on stability of total hip replacement - A mechanical test study
    Bader, R
    Scholz, R
    Steinhauser, E
    Zimmermann, S
    Busch, R
    Mittelmeier, W
    [J]. ACTA ORTHOPAEDICA SCANDINAVICA, 2004, 75 (04): : 415 - 421
  • [3] The significance of metal staining on alumina femoral heads in total hip arthroplasty
    Bal, B. Sonny
    Rahaman, Mohamed N.
    Aleto, Thomas
    Miller, F. Scott
    Traina, Francesco
    Toni, Aldo
    [J]. JOURNAL OF ARTHROPLASTY, 2007, 22 (01) : 14 - 19
  • [4] Impingement and rim wear associated with early osteolysis after a total hip replacement - A case report
    Barrack, RL
    Schmalzried, TP
    [J]. JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2002, 84A (07) : 1218 - 1220
  • [5] Concerns about ceramics in THA
    Barrack, RL
    Barak, C
    Skinner, HB
    [J]. CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2004, (429) : 73 - 79
  • [6] Ceramic-on-ceramic bearings in total hip arthroplasty
    Bierbaum, BE
    Nairus, J
    Kuesis, D
    Morrison, JC
    Ward, D
    [J]. CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2002, (405) : 158 - 163
  • [7] Cracking and impingement in ultra-high-molecular-weight polyethylene acetabular liners
    Birman, MV
    Noble, PC
    Conditt, MA
    Li, S
    Mathis, KB
    [J]. JOURNAL OF ARTHROPLASTY, 2005, 20 (07) : 87 - 92
  • [8] Range of motion and stability in total hip arthroplasty with 28-, 32-, 38-, and 44-mm femoral head sizes - An in vitro study
    Burroughs, BR
    Hallstrom, B
    Golladay, GJ
    Hoeffel, D
    Harris, WH
    [J]. JOURNAL OF ARTHROPLASTY, 2005, 20 (01) : 11 - 19
  • [9] The effect of the orientation of the acetabular and femoral components on the range of motion of the hip at different head-neck ratios
    D'Lima, DD
    Urquhart, AG
    Buehler, KO
    Walker, RH
    Colwell, CW
    [J]. JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2000, 82A (03) : 315 - 321
  • [10] Optimizing acetabular component position to minimize impingement and reduce contact stress
    D'Lima, DD
    Chen, PC
    Colwell, CW
    [J]. JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2001, 83A : 87 - 91