Design and validation of a CT-guided robotic system for lung cancer brachytherapy

被引:18
作者
Dou, Huaisu [1 ]
Jiang, Shan [1 ,2 ]
Yang, Zhiyong [1 ]
Sun, Luqing [1 ]
Ma, Xiaodong [1 ]
Huo, Bin [3 ]
机构
[1] Tianjin Univ, Sch Mech Engn, Tianjin 300350, Peoples R China
[2] Tianjin Univ, Ctr Adv Mech & Robot, Tianjin 300350, Peoples R China
[3] Tianjin Med Univ, Dept Oncol, Hosp 2, Tianjin 300211, Peoples R China
基金
中国国家自然科学基金;
关键词
CT-guided; lung cancer brachytherapy; surgical robot; treatment planning; PUNCTURE ROBOT; ACCURACY; BIOPSIES; SAFETY;
D O I
10.1002/mp.12435
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: Currently, lung brachytherapy in clinical setting is a complex procedure. Operation accuracy depends on accurate positioning of the template; however, it is difficult to guarantee the positioning accuracy manually. Application of robotic-assisted systems can simplify the procedure and improve the manual positioning accuracy. Therefore, a novel CT-guided robotic system was developed to assist the lung cancer brachytherapy. Methods: A four degree-of-freedom (DOF) robot, controlled by a lung brachytherapy treatment planning system (TPS) software, was designed and manufactured to assist the template positioning. Target position of the template can be obtained from the treatment plan, thus the robot is driven to the target position automatically. The robotic system was validated in both the laboratory and the CT environment. In laboratory environment, a 3D laser tracker and an inertial measurement unit (IMU) were used to measure the mechanical accuracy in air, which includes positioning accuracy and position repeatability. Working reliability was also validated in this procedure by observing the response reliability and calculating the position repeatability. Imaging artifacts and accuracy of the robot registration were validated in the CT environment by using an artificial phantom with fiducial markers. CT images were obtained and used to test the image artifact and calculate the registration accuracy. Phantom experiments were conducted to test the accuracy of needle insertion by using a transparent hydrogel phantom with a high imitation artificial phantom. Also, the efficiency was validated in this procedure by comparing time costs in manual positioning with robotic positioning under the same experimental conditions. Results: The robotic system achieved the positioning accuracy of 0.28 +/- 0.25 mm and the position repeatability of 0.09 +/- 0.11 mm. Experimental results showed that the robot was CT-compatible and responded reliably to the control commands. The mean registration accuracy of the robotic system was 0.49 +/- 0.29 mm. Phantom experiments indicated that the accuracy of needle insertion was 1.5 +/- 1.7 mm at a depth ranging from 30 to 80 mm. The time used to adjust the template to the target position was 12 min on average by robotic system automatically. An average of 30 min was saved compared with the manual positioning procedure in phantom experiments. Conclusions: This paper describes the design and experimental validation of a novel CT-guided robotic system for lung cancer brachytherapy. The robotic system was validated in a number of aspects which prove that it was capable of locating the template with clinically acceptable accuracy in the CT environment. All experimental results indicated that the system is reliable and ready to be applied to further studies on animals. (C) 2017 American Association of Physicists in Medicine
引用
收藏
页码:4828 / 4837
页数:10
相关论文
共 50 条
  • [31] Design and Validation of a CT- and MRI-Guided Robot for Percutaneous Needle Procedures
    Hungr, Nikolai
    Bricault, Ivan
    Cinquin, Philippe
    Fouard, Celine
    IEEE TRANSACTIONS ON ROBOTICS, 2016, 32 (04) : 973 - 987
  • [32] COMPLICATIONS OF CT-GUIDED PERCUTANEOUS LUNG BIOPSY AND ITS RISK FACTORS
    Ni, Y.
    Chen, K.
    Shi, G.
    Wan, H.
    RESPIROLOGY, 2011, 16 : 274 - 274
  • [33] ROLE OF CT AND CT-GUIDED NODAL BIOPSY IN STAGING OF PROSTATIC-CANCER
    AMIS, ES
    RADIOLOGY, 1994, 190 (02) : 309 - 310
  • [34] CT-guided percutaneous laser ablation of metastatic lung cancer: three cases report and literature review
    Zhao, Qiyu
    Tian, Guo
    Chen, Fen
    Zhong, Liyun
    Jiang, Tian'an
    ONCOTARGET, 2017, 8 (02) : 2187 - 2196
  • [35] Electromagnetic Navigation System for CT-Guided Biopsy of Small Lesions
    Appelbaum, Liat
    Sosna, Jacob
    Nissenbaum, Yizhak
    Benshtein, Alexander
    Goldberg, S. Nahum
    AMERICAN JOURNAL OF ROENTGENOLOGY, 2011, 196 (05) : 1194 - 1200
  • [36] Use of a surgical navigation system for CT-guided template production
    Widmann, Gerlig
    Widmann, Roland
    Widmann, Ekkehard
    Jaschke, Werner
    Bale, Reto
    INTERNATIONAL JOURNAL OF ORAL & MAXILLOFACIAL IMPLANTS, 2007, 22 (01) : 72 - 78
  • [37] CT-guided lung biopsy: Factors influencing diagnostic yield and complication rate
    Anderson, JM
    Murchison, J
    Patel, D
    CLINICAL RADIOLOGY, 2003, 58 (10) : 791 - 797
  • [38] Pleural and Peripheral Lung Lesions: Comparison of US- and CT-guided Biopsy
    Sconfienza, Luca Maria
    Mauri, Giovanni
    Grossi, Francesco
    Truini, Mauro
    Serafini, Giovanni
    Sardanelli, Francesco
    Murolo, Carmelina
    RADIOLOGY, 2013, 266 (03) : 930 - 935
  • [39] Comparison of a Robotic and Patient-Mounted Device for CT-Guided Needle Placement: A Phantom Study
    Scharll, Yannick
    Mitteregger, Alexander
    Laimer, Gregor
    Schwabl, Christoph
    Schullian, Peter
    Bale, Reto
    JOURNAL OF CLINICAL MEDICINE, 2022, 11 (13)
  • [40] Semiautomated robotic, CT-guided needle placement for postmortem CSF sampling - a Novel application of the Virtobot
    Franckenberg, Sabine
    Sieberth, Till
    Frauenfelder, Thomas
    Thali, Michael J.
    Ebert, Lars Christian
    ALL LIFE, 2021, 14 (01) : 75 - 79