Design and validation of a CT-guided robotic system for lung cancer brachytherapy

被引:18
|
作者
Dou, Huaisu [1 ]
Jiang, Shan [1 ,2 ]
Yang, Zhiyong [1 ]
Sun, Luqing [1 ]
Ma, Xiaodong [1 ]
Huo, Bin [3 ]
机构
[1] Tianjin Univ, Sch Mech Engn, Tianjin 300350, Peoples R China
[2] Tianjin Univ, Ctr Adv Mech & Robot, Tianjin 300350, Peoples R China
[3] Tianjin Med Univ, Dept Oncol, Hosp 2, Tianjin 300211, Peoples R China
基金
中国国家自然科学基金;
关键词
CT-guided; lung cancer brachytherapy; surgical robot; treatment planning; PUNCTURE ROBOT; ACCURACY; BIOPSIES; SAFETY;
D O I
10.1002/mp.12435
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: Currently, lung brachytherapy in clinical setting is a complex procedure. Operation accuracy depends on accurate positioning of the template; however, it is difficult to guarantee the positioning accuracy manually. Application of robotic-assisted systems can simplify the procedure and improve the manual positioning accuracy. Therefore, a novel CT-guided robotic system was developed to assist the lung cancer brachytherapy. Methods: A four degree-of-freedom (DOF) robot, controlled by a lung brachytherapy treatment planning system (TPS) software, was designed and manufactured to assist the template positioning. Target position of the template can be obtained from the treatment plan, thus the robot is driven to the target position automatically. The robotic system was validated in both the laboratory and the CT environment. In laboratory environment, a 3D laser tracker and an inertial measurement unit (IMU) were used to measure the mechanical accuracy in air, which includes positioning accuracy and position repeatability. Working reliability was also validated in this procedure by observing the response reliability and calculating the position repeatability. Imaging artifacts and accuracy of the robot registration were validated in the CT environment by using an artificial phantom with fiducial markers. CT images were obtained and used to test the image artifact and calculate the registration accuracy. Phantom experiments were conducted to test the accuracy of needle insertion by using a transparent hydrogel phantom with a high imitation artificial phantom. Also, the efficiency was validated in this procedure by comparing time costs in manual positioning with robotic positioning under the same experimental conditions. Results: The robotic system achieved the positioning accuracy of 0.28 +/- 0.25 mm and the position repeatability of 0.09 +/- 0.11 mm. Experimental results showed that the robot was CT-compatible and responded reliably to the control commands. The mean registration accuracy of the robotic system was 0.49 +/- 0.29 mm. Phantom experiments indicated that the accuracy of needle insertion was 1.5 +/- 1.7 mm at a depth ranging from 30 to 80 mm. The time used to adjust the template to the target position was 12 min on average by robotic system automatically. An average of 30 min was saved compared with the manual positioning procedure in phantom experiments. Conclusions: This paper describes the design and experimental validation of a novel CT-guided robotic system for lung cancer brachytherapy. The robotic system was validated in a number of aspects which prove that it was capable of locating the template with clinically acceptable accuracy in the CT environment. All experimental results indicated that the system is reliable and ready to be applied to further studies on animals. (C) 2017 American Association of Physicists in Medicine
引用
收藏
页码:4828 / 4837
页数:10
相关论文
共 50 条
  • [21] Study on the Clinical Application of CT-Guided Percutaneous Lung Biopsy
    Zhang, Peng
    Long, Feng
    Zhang, Youzhi
    Yang, Haihua
    JOURNAL OF THORACIC ONCOLOGY, 2015, 10 (09) : S620 - S620
  • [22] Percutaneous lung biopsy: comparison between an augmented reality CT navigation system and standard CT-guided technique
    Grasso, R. F.
    Faiella, E.
    Luppi, G.
    Schena, E.
    Giurazza, F.
    Del Vescovo, R.
    D'Agostino, F.
    Cazzato, R. L.
    Zobel, B. Beomonte
    INTERNATIONAL JOURNAL OF COMPUTER ASSISTED RADIOLOGY AND SURGERY, 2013, 8 (05) : 837 - 848
  • [23] A Miniature Optical Neuronavigation System for CT-Guided Stereotaxy
    von Jako, Christopher R.
    Perlaki, Gabor
    Orsi, Gergely
    Schwarcz, Attila
    Doczi, Tamas
    STEREOTACTIC AND FUNCTIONAL NEUROSURGERY, 2014, 92 (01) : 1 - 7
  • [24] LASER PROJECTION SYSTEM FOR RADIOTHERAPY AND CT-GUIDED BIOPSY
    NAGATA, Y
    NISHIDAI, T
    ABE, M
    HIRAOKA, M
    TAKAHASHI, M
    FUJIWARA, K
    OKAJIMA, K
    JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 1990, 14 (06) : 1046 - 1048
  • [25] LASER GUIDANCE-SYSTEM FOR CT-GUIDED PROCEDURES
    MIAUX, Y
    GUERMAZI, A
    GOSSOT, D
    BOURRIER, P
    ANGOULVANT, D
    KHAIROUNE, A
    TURKI, C
    BOUCHE, E
    RADIOLOGY, 1995, 194 (01) : 282 - 284
  • [26] Development and validation of a rapidly deployable CT-guided stereotactic system for external ventricular drainage: preclinical study
    Barath, Abhijeet S.
    Rusheen, Aaron E.
    Cabrera, Juan M. Rojas
    Shin, Hojin
    Blaha, Charles D.
    Bennet, Kevin E.
    Goerss, Stephan J.
    Lee, Kendall H.
    Oh, Yoonbae
    SCIENTIFIC REPORTS, 2021, 11 (01)
  • [27] Percutaneous CT-guided lung biopsy for the diagnosis of persistent pulmonary consolidation
    Brioulet, J.
    David, A.
    Sagan, C.
    Cellerin, L.
    Frampas, E.
    Morla, O.
    DIAGNOSTIC AND INTERVENTIONAL IMAGING, 2020, 101 (11) : 727 - 732
  • [28] Analysis of factors influencing accuracy and complications in CT-guided lung biopsy
    Nakatani, Miyuki
    Tanigawa, Noboru
    Kariya, Shuji
    Komemushi, Atsushi
    Yagi, Rie
    Sawada, Satoshi
    MINIMALLY INVASIVE THERAPY & ALLIED TECHNOLOGIES, 2012, 21 (06) : 415 - 422
  • [29] CT-guided percutaneous radiofrequency ablation for lung neoplasms adjacent to the pericardium
    Mu, Luwen
    Pan, Tao
    Lyu, Ning
    Sun, Lin
    Li, Shaolong
    Xie, Qiankun
    Deng, Haijing
    Wu, Peihong
    Liu, Hui
    Zhao, Ming
    LUNG CANCER, 2018, 122 : 25 - 31
  • [30] A prospective trial of CT-guided percutaneous microwave ablation for lung tumors
    Reisenauer, Janani S.
    Eiken, Patrick W.
    Callstrom, Matthew R.
    Johnson, Geoffrey B.
    Pierson, Karlyn
    Lechtenberg, Bettie
    Blackmon, Shanda H.
    JOURNAL OF THORACIC DISEASE, 2022, : 939 - 951