Design and validation of a CT-guided robotic system for lung cancer brachytherapy

被引:18
|
作者
Dou, Huaisu [1 ]
Jiang, Shan [1 ,2 ]
Yang, Zhiyong [1 ]
Sun, Luqing [1 ]
Ma, Xiaodong [1 ]
Huo, Bin [3 ]
机构
[1] Tianjin Univ, Sch Mech Engn, Tianjin 300350, Peoples R China
[2] Tianjin Univ, Ctr Adv Mech & Robot, Tianjin 300350, Peoples R China
[3] Tianjin Med Univ, Dept Oncol, Hosp 2, Tianjin 300211, Peoples R China
基金
中国国家自然科学基金;
关键词
CT-guided; lung cancer brachytherapy; surgical robot; treatment planning; PUNCTURE ROBOT; ACCURACY; BIOPSIES; SAFETY;
D O I
10.1002/mp.12435
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: Currently, lung brachytherapy in clinical setting is a complex procedure. Operation accuracy depends on accurate positioning of the template; however, it is difficult to guarantee the positioning accuracy manually. Application of robotic-assisted systems can simplify the procedure and improve the manual positioning accuracy. Therefore, a novel CT-guided robotic system was developed to assist the lung cancer brachytherapy. Methods: A four degree-of-freedom (DOF) robot, controlled by a lung brachytherapy treatment planning system (TPS) software, was designed and manufactured to assist the template positioning. Target position of the template can be obtained from the treatment plan, thus the robot is driven to the target position automatically. The robotic system was validated in both the laboratory and the CT environment. In laboratory environment, a 3D laser tracker and an inertial measurement unit (IMU) were used to measure the mechanical accuracy in air, which includes positioning accuracy and position repeatability. Working reliability was also validated in this procedure by observing the response reliability and calculating the position repeatability. Imaging artifacts and accuracy of the robot registration were validated in the CT environment by using an artificial phantom with fiducial markers. CT images were obtained and used to test the image artifact and calculate the registration accuracy. Phantom experiments were conducted to test the accuracy of needle insertion by using a transparent hydrogel phantom with a high imitation artificial phantom. Also, the efficiency was validated in this procedure by comparing time costs in manual positioning with robotic positioning under the same experimental conditions. Results: The robotic system achieved the positioning accuracy of 0.28 +/- 0.25 mm and the position repeatability of 0.09 +/- 0.11 mm. Experimental results showed that the robot was CT-compatible and responded reliably to the control commands. The mean registration accuracy of the robotic system was 0.49 +/- 0.29 mm. Phantom experiments indicated that the accuracy of needle insertion was 1.5 +/- 1.7 mm at a depth ranging from 30 to 80 mm. The time used to adjust the template to the target position was 12 min on average by robotic system automatically. An average of 30 min was saved compared with the manual positioning procedure in phantom experiments. Conclusions: This paper describes the design and experimental validation of a novel CT-guided robotic system for lung cancer brachytherapy. The robotic system was validated in a number of aspects which prove that it was capable of locating the template with clinically acceptable accuracy in the CT environment. All experimental results indicated that the system is reliable and ready to be applied to further studies on animals. (C) 2017 American Association of Physicists in Medicine
引用
收藏
页码:4828 / 4837
页数:10
相关论文
共 50 条
  • [11] A robotic needle-positioning and guidance system for CT-guided puncture: Ex vivo results
    Kettenbach, Joachim
    Kara, Levent
    Toporek, Grzegorz
    Fuerst, Martin
    Kronreif, Gernot
    MINIMALLY INVASIVE THERAPY & ALLIED TECHNOLOGIES, 2014, 23 (05) : 271 - 278
  • [12] CT-guided percutaneous cryoablation of central lung tumors
    Colak, Errol
    Tatli, Servet
    Shyn, Paul B.
    Tuncali, Kemal
    Silverman, Stuart G.
    DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY, 2014, 20 (04): : 316 - 322
  • [13] Implant time and process efficiency for CT-guided high-dose-rate brachytherapy for cervical cancer
    Mayadev, Jyoti
    Qi, Lihong
    Lentz, Susan
    Benedict, Stanley
    Courquin, Jean
    Dieterich, Sonja
    Mathai, Mathew
    Stern, Robin
    Valicenti, Richard
    BRACHYTHERAPY, 2014, 13 (03) : 233 - 239
  • [14] A hybrid robot system for CT-guided surgery
    Liu, Da
    Wang, Tianmiao
    Tang, Can
    Zhang, Fan
    ROBOTICA, 2010, 28 : 253 - 258
  • [15] Evaluation of robotic-assisted navigation system for CT-guided thoracic and abdominal lesion puncture: A prospective clinical study
    Jing, Yifan
    Zhang, Jian
    Jin, Yong
    Bai, Xuming
    JOURNAL OF CANCER RESEARCH AND THERAPEUTICS, 2024, 20 (04) : 1350 - 1356
  • [16] Value of focused lung ultrasound in diagnosing and monitoring pneumothorax after CT-guided lung biopsy
    Metzler, Emma Katrine
    Everloff, Nina
    Juul, Amanda Dandanell
    Laursen, Christian Borbjerg
    Graumann, Ole
    Pietersen, Pia Iben
    JOURNAL OF ULTRASOUND, 2025, : 11 - 18
  • [17] CT-guided Biopsy for the Diagnosis of Pulmonary Epithelioid Hemangioendothelioma Mimicking Metastatic Lung Cancer
    Sasaki, Akiko
    Egashira, Hideto
    Sugimoto, Hideyasu
    Seki, Kenichi
    Tsukiyama, Toshitaka
    Ichita, Chikamasa
    Tokoro, Shinnosuke
    Takizawa, Satoshi
    Kitagawa, Lzumi
    Teshime, Shinichi
    Kako, Makoto
    INTERNAL MEDICINE, 2018, 57 (24) : 3631 - 3635
  • [18] Evaluation of an Interactive Breath-Hold Control System in CT-Guided Lung Biopsy
    Schoth, F.
    Plumhans, C.
    Kraemer, N.
    Mahnken, A.
    Friebe, M.
    Guenther, R. W.
    Krombach, G.
    ROFO-FORTSCHRITTE AUF DEM GEBIET DER RONTGENSTRAHLEN UND DER BILDGEBENDEN VERFAHREN, 2010, 182 (06): : 507 - 511
  • [19] Clinical value of an electromagnetic navigation system for CT-guided percutaneous lung biopsy of peripheral lung lesions
    Zhang, Yao
    Huang, Haozhe
    Cao, Shuhui
    Teng, Jiajun
    Wang, Yue
    Ling, Xuxinyi
    Zhou, Yan
    Xu, Jianlin
    Li, Wentao
    Zhong, Hua
    JOURNAL OF THORACIC DISEASE, 2021, 13 (08) : 4885 - 4893
  • [20] Management of pneumothorax after percutaneous CT-guided lung biopsy
    Yamagami, T
    Nakamura, T
    Iida, S
    Kato, T
    Nishimura, T
    CHEST, 2002, 121 (04) : 1159 - 1164