Development of EEG Data Acquisition System based on FPGA Zedboard

被引:0
作者
Gani, Hendra Saputra [1 ]
Wijaya, Sastra Kusuma [1 ]
Prawito [1 ]
Toresano, La Ode Husein Z. [1 ]
机构
[1] Univ Indonesia, Fak Matemat & Ilmu Pengetahuan Alam, Dept Fis, Kampus Baru UI, Depok, Indonesia
来源
PROCEEDINGS OF 2017 5TH INTERNATIONAL CONFERENCE ON INSTRUMENTATION, COMMUNICATIONS, INFORMATION TECHNOLOGY, AND BIOMEDICAL ENGINEERING (ICICI-BME): SCIENCE AND TECHNOLOGY FOR A BETTER LIFE | 2017年
关键词
Electroencephalography; EEG; FPGA; ADS1299TI; Data Acquisition; Zedboard;
D O I
暂无
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
An Electroencephalography (EEG) data acquisition system based on FPGA Zedboard (Diligent, USA) has been implemented using ADC (Analog to Digital Converter) front-end ADS1299FE (Texas Instrument, USA). The EEG instrument is a device used to measure the electrical brain activities on the scalp, and the voltage potential of EEG signal around 0.5-100 mu V with frequency 0.5 - 40 Hz. The data acquisition system has been configured with 2-chips ADS1299 Daisy-Chain mode and processing based on Zedboard FPGA. The data acquisition system was used with FPGA, because that is very powerful speed to acquire the ADC devices. The result of data acquisition system has been sent to PC (Personal Computer) and also displayed into the graphic format of EEG data. The EEG instrument has been successfully validated using EEC Simulator (NETECH Mini-Sim EEG) with frequency 2Hz. 5Hz, and also voltage tested in 30 uV, 50 uV and 100 uV. The validation result at 2Hz was obtained measurement with a maximum deviation average of 0.15uV, therefore frequency of 5Hz with a maximum deviation average of 0.18uV.
引用
收藏
页码:246 / 250
页数:5
相关论文
共 8 条
[1]  
Brozino J. D., 2000, BIOMEDICAL ENG HDB
[2]  
Lovelace JA, 2013, MIDWEST SYMP CIRCUIT, P1172, DOI 10.1109/MWSCAS.2013.6674862
[3]  
Lovelace JA, 2013, IEEE ENG MED BIO, P6361, DOI 10.1109/EMBC.2013.6611009
[4]  
Teplan M., 2002, Measurement Science Review, V2, P1, DOI DOI 10.1021/PR070350L
[5]  
Texas Instruments, 2012, LOW NOIS 8 CHANN 24
[6]   Data Acquisition Instrument for EEG based on Embedded System [J].
Toresano, La Ode Husein Z. ;
Wijaya, Sastra Kusuma ;
Prawito ;
Sudarmaji, Arief ;
Syakura, Abdan ;
Badri, Cholid .
BIOMEDICAL ENGINEERING'S RECENT PROGRESS IN BIOMATERIALS, DRUGS DEVELOPMENT, AND MEDICAL DEVICES, 2017, 1817
[7]  
Williams R., 2006, REAL TIME SYSTEMS DE
[8]   A real-time comparison between direct control, sequential pattern recognition control and simultaneous pattern recognition control using a Fitts' law style assessment procedure [J].
Wurth, Sophie M. ;
Hargrove, Levi J. .
JOURNAL OF NEUROENGINEERING AND REHABILITATION, 2014, 11