Preclinical Evaluation of Pulsed Field Ablation Electrophysiological and Histological Assessment of Thoracic Vein Isolation

被引:164
作者
Koruth, Jacob [1 ]
Kuroki, Kenji [1 ]
Iwasawa, Jin [1 ]
Enomoto, Yoshinari [1 ]
Viswanathan, Raju [2 ]
Brose, Richard [2 ]
Buck, Eric D. [2 ]
Speltz, Molly [3 ]
Dukkipati, Srinivas R. [1 ]
Reddy, Vivek Y. [1 ]
机构
[1] Icahn Sch Med Mt Sinai, Helmsley Electrophysiol Ctr, Dept Cardiol, New York, NY 10029 USA
[2] Farapulse Inc, Menlo Pk, CA USA
[3] R&M Speltz LLC, Stanchfield, MN USA
关键词
atrial fibrillation; catheter ablation; endocardium; phrenic nerve; pulmonary vein; TISSUE ABLATION; ELECTROPORATION;
D O I
10.1161/CIRCEP.119.007781
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background: Pulsed field ablation (PFA) is a uniquely tissue-selective, nonthermal cardiac ablation modality. Delivery parameters such as the electrical waveform composition and device design are critical to PFA's efficacy and safety, particularly tissue specificity. In a series of preclinical studies, we sought to examine the electrophysiological and histological effects of PFA and compare the safety and feasibility of durable pulmonary vein and superior vena cava (SVC) isolation between radiofrequency ablation and PFA waveforms. Methods: A femoral venous approach was used to gain right and left atrial access under general anesthesia in healthy swine. Baseline potentials in right superior pulmonary and inferior common vein and in SVC were assessed. Bipolar PFA was performed with monophasic (PFA(Mono)) and biphasic (PFA(Bi)) waveforms in 7 and 7 swine sequentially and irrigated radiofrequency ablation in 3 swine. Vein potentials were then assessed acutely, and at approximate to 10 weeks; histology was obtained. Results: All targeted veins (n=46) were successfully isolated on the first attempt in all cohorts. The PFA(Bi) waveform induced significantly less skeletal muscle engagement. Pulmonary vein isolation durability was assessed in 28 veins: including the SVC, durability was significantly higher in the PFA(Bi) group (18/18 PFA(Bi), 10/18 PFA(Mono), 3/6 radiofrequency, P=0.002). Transmurality rates were similar across groups with evidence of nerve damage only with radiofrequency. Pulmonary vein narrowing was noted only in the radiofrequency cohort. The phrenic nerve was spared in all cohorts but at the expense of incomplete SVC encirclement with radiofrequency. Conclusions: In this chronic porcine study, PFA-based pulmonary vein and SVC isolation were safe and efficacious with demonstrable sparing of nerves and venous tissue. This preclinical study provided the scientific basis for the first-in-human endocardial PFA studies.
引用
收藏
页数:9
相关论文
共 14 条
[11]   Ablation of Atrial Fibrillation With Pulsed Electric Fields An Ultra-Rapid, Tissue-Selective Modality for Cardiac Ablation [J].
Reddy, Vivek Y. ;
Koruth, Jacob ;
Jais, Pierre ;
Petru, Jan ;
Timko, Ferdinand ;
Skalsky, Ivo ;
Hebeler, Robert ;
Labrousse, Louis ;
Barandon, Laurent ;
Kralovec, Stepan ;
Funosako, Moritoshi ;
Mannuva, Boochi Babu ;
Sediva, Lucie ;
Neuzil, Petr .
JACC-CLINICAL ELECTROPHYSIOLOGY, 2018, 4 (08) :987-995
[12]   Intracardiac pulsed field ablation: Proof of feasibility in a chronic porcine model [J].
Stewart, Mark T. ;
Haines, David E. ;
Verma, Atul ;
Kirchhof, Nicole ;
Barka, Noah ;
Grassi, Erin ;
Howard, Brian .
HEART RHYTHM, 2019, 16 (05) :754-764
[13]   Low vulnerability of the right phrenic nerve to electroporation ablation [J].
van Driel, Vincent J. H. M. ;
Neven, Kars ;
van Wessel, Harri ;
Vink, Aryan ;
Doevendans, Pieter A. F. M. ;
Wittkampf, Fred H. M. .
HEART RHYTHM, 2015, 12 (08) :1838-1844
[14]   Pulmonary Vein Stenosis After Catheter Ablation Electroporation Versus Radiofrequency [J].
van Driel, Vincent J. H. M. ;
Neven, Kars G. E. J. ;
van Wessel, Harry ;
du Pre, Bastiaan C. ;
Vink, Aryan ;
Doevendans, Pieter A. F. M. ;
Wittkampf, Fred H. M. .
CIRCULATION-ARRHYTHMIA AND ELECTROPHYSIOLOGY, 2014, 7 (04) :734-738