Comparative repeatability of two handheld fractional exhaled nitric oxide monitors

被引:30
|
作者
Kapande, Kirsty M. [1 ]
McConaghy, Laura A. [2 ]
Douglas, Isabella [3 ]
McKenna, Sonia [3 ]
Hughes, Jenny L. [3 ]
McCance, David R. [3 ]
Ennis, Madeline [1 ]
Shields, Michael D. [1 ]
机构
[1] Queens Univ Belfast, Ctr Infect & Immun, Belfast BT9 7BL, Antrim, North Ireland
[2] Victoria Infirm, NHS Greater Glasgow & Clyde, Glasgow G42 9TY, Lanark, Scotland
[3] Belfast Hlth & Social Care Trust, Royal Hosp Grp, Belfast, Antrim, North Ireland
关键词
child; handheld; airways inflammation; repeatability; fractional exhaled nitric oxide; CHILDHOOD ASTHMA; STANDARD TECHNIQUE; CHILDREN; DEVICE; INFLAMMATION; ANALYZER; ADULTS;
D O I
10.1002/ppul.21591
中图分类号
R72 [儿科学];
学科分类号
100202 ;
摘要
Background The use of portable fractional exhaled nitric oxide (FENO) devices is increasingly common in the diagnosis and management of allergic airways inflammation. Methods We tested two handheld FENO devices, to determine (a) if there was adequate intradevice repeatability to allow the use of single breath testing, and (b) if the devices could be used interchangeably. In a mixed pediatric population, including normal, asthmatic, and children with peanut allergies, 858 paired values were collected from the NIOX-MINO (R) and/or the NObreath (R) devices. Results The NIOX-MINO (R) showed excellent repeatability (mean difference of 0.1 with 95% limits of agreement between -7.93 to 7.72?ppb), while the NObreath (R) showed good repeatability (mean difference of -1.61 with 95% limits of agreement between -14.1 and 10.8?ppb). Intradevice repeatability was good but not adequate and the NIOX-MINO (R) systematically produced higher results than the NObreath (R) [mean difference of 7.8?ppb with 95% limits of agreement from -11.55 to 27.52?ppb (-33% to 290%)]. Conclusions Our results support the manufacturer's advice that single breath testing is appropriate for the NIOX-MINO (R). NObreath (R) results indicate that the mean of more than one breath should be utilized. The devices cannot be used interchangeably. Pediatr Pulmonol. 2012; 47:546550. (c) 2011 Wiley Periodicals, Inc.
引用
收藏
页码:546 / 550
页数:5
相关论文
共 50 条
  • [1] Comparative Repeatability of Two Handheld Fractional Exhaled Nitric Oxide Monitors
    Kapande, K. M.
    Shields, M. D.
    McConaghy, L. A.
    IRISH JOURNAL OF MEDICAL SCIENCE, 2011, 180 : S460 - S460
  • [2] Fractional exhaled nitric oxide measurement with a handheld device
    Magori, Erhard
    Hiltawsky, Karsten
    Fleischer, Maximilian
    Simon, Elfriede
    Pohle, Roland
    von Sicard, Oliver
    Tawil, Angelika
    JOURNAL OF BREATH RESEARCH, 2011, 5 (02)
  • [3] Intra-device repeatability of the NIOX-MINO handheld fractional exhaled nitric oxide monitor
    Williams, Spencer
    Sopuluchukwu, Franklin
    Sivagnanasithiyar, Sindhu
    Robertson, Meredith
    Iles, Richard W.
    EUROPEAN RESPIRATORY JOURNAL, 2019, 54
  • [4] Comparison of Two Handheld Fractional Exhaled Nitric Oxide Measurements in the Assessment of Asthma Patients
    Kim, Sujeong
    Song, Woo-Jung
    Lee, Jong-Myung
    Kang, Hye-Ryun
    Park, Heung-Woo
    Cho, Sang Heon
    Min, Kyung-Up
    JOURNAL OF ALLERGY AND CLINICAL IMMUNOLOGY, 2015, 135 (02) : AB174 - AB174
  • [5] Within-day repeatability of fractional exhaled nitric oxide in severe asthma
    Wang, Ran
    Alahmadi, Fahad
    Niven, Robert
    Fowler, Stephen
    EUROPEAN RESPIRATORY JOURNAL, 2020, 56
  • [6] Same-day repeatability of fractional exhaled nitric oxide in severe asthma
    Wang, Ran
    Alahmadi, Fahad
    Niven, Robert
    Fowler, Stephen J.
    EUROPEAN RESPIRATORY JOURNAL, 2021, 57 (05)
  • [7] Between-occasion repeatability of fractional exhaled nitric oxide measurements in children
    Baranski, Kamil
    Zejda, Jan E.
    JORNAL BRASILEIRO DE PNEUMOLOGIA, 2018, 44 (01) : 52 - 54
  • [8] Validation study of fractional exhaled nitric oxide measurements using a handheld monitoring device
    Gill, Michelle
    Graff, Gavin R.
    Adler, Alan J.
    Dweik, Raed A.
    JOURNAL OF ASTHMA, 2006, 43 (10) : 731 - 734
  • [9] Short-term repeatability of fractional exhaled nitric oxide (FENO) measurement in children
    Baranski, Kamil
    Zejda, Jan Eugeniusz
    EUROPEAN RESPIRATORY JOURNAL, 2016, 48
  • [10] A prospective trial comparing two handheld devices to measure fractional exhaled nitric oxide to the "gold-standard-technique".
    Thanikkel, Leo
    Vogt, Andreas
    Micic, Srdjan
    Delliehausen, Nadja
    Moeller, Alexander
    EUROPEAN RESPIRATORY JOURNAL, 2024, 64