A Joint Delay-and-Sum and Fourier Beamforming Method for High Frame Rate Ultrasound Imaging

被引:2
|
作者
Zhao, Wei [1 ,2 ]
Feng, Shuai [1 ]
Wang, Yadan [1 ]
Wang, Yuanguo [1 ]
Han, Zhihui [1 ]
Peng, Hu [1 ]
机构
[1] Hefei Univ Technol, Dept Biomed Engn, Hefei 230009, Peoples R China
[2] Wannan Med Coll, Yijishan Hosp, Wuhu 241001, Peoples R China
来源
CMES-COMPUTER MODELING IN ENGINEERING & SCIENCES | 2020年 / 123卷 / 01期
基金
中国国家自然科学基金;
关键词
High frame rate imaging; delay-and-sum; Fourier transformation; angular spectrum; MINIMUM-VARIANCE BEAMFORMER; MEDICAL ULTRASOUND;
D O I
10.32604/cmes.2020.09387
中图分类号
T [工业技术];
学科分类号
08 ;
摘要
Frame rate is an important metric for ultrasound imaging systems, and high frame rates (HFR) benefit moving-target imaging. One common way to obtain HFR imaging is to transmit a plane wave. Delay-and-sum (DAS) beamformer is a conventional beamforming algorithm, which is simple and has been widely implemented in clinical application. Fourier beamforming is an alternative method for HFR imaging and has high levels of imaging efficiency, imaging speed, and good temporal dynamic characteristics. Nevertheless, the resolution and contrast performance of HFR imaging based on DAS or Fourier beamforming are insufficient due to the single plane wave transmission. To address this problem, a joint DAS and Fourier beamforming method is introduced in this study. The proposed method considers the different distributions of sidelobes in DAS imaging and Fourier imaging and combines the angular spectrum and DAS to reconstruct ultrasound images. The proposed method is evaluated on simulation and experimental phantom datasets to compare its performance with DAS and Fourier beamforming methods. Results demonstrate that the proposed method improves image effective dynamic range and resolution while also retaining a high frame rate of the ultrasound imaging systems. The proposed method improves the effective dynamic range along axial and lateral directions by 10 dB, compared to standard DAS and Fourier beamforming.
引用
收藏
页码:427 / 440
页数:14
相关论文
共 50 条
  • [31] High dynamic range ultrasound imaging with real-time Filtered-Delay Multiply and Sum beamforming
    Ramalli, Alessandro
    Dallai, Alessandro
    Bassi, Luca
    Scaringella, Monica
    Boni, Enrico
    Hine, Gabriel Emile
    Matrone, Giulia
    Savoia, Alessandro Stuart
    Tortoli, Piero
    2017 IEEE INTERNATIONAL ULTRASONICS SYMPOSIUM (IUS), 2017,
  • [32] iMAP Beamforming for High-Quality High Frame Rate Imaging
    Chernyakova, Tanya
    Cohen, Dan
    Shoham, Meged
    Eldar, Yonina C.
    IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 2019, 66 (12) : 1830 - 1844
  • [33] The Delay Multiply and Sum Beamforming Algorithm in Ultrasound B-Mode Medical Imaging
    Matrone, Giulia
    Savoia, Alessandro Stuart
    Caliano, Giosue
    Magenes, Giovanni
    IEEE TRANSACTIONS ON MEDICAL IMAGING, 2015, 34 (04) : 940 - 949
  • [34] Ultrasound Plane-Wave Imaging with Delay Multiply And Sum Beamforming and Coherent Compounding
    Matrone, Giulia
    Savoia, Alessandro S.
    Caliano, Giosue
    Magenes, Giovanni
    2016 38TH ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY (EMBC), 2016, : 3223 - 3226
  • [35] Regional-Lag Signed Delay Multiply and Sum Beamforming in Ultrafast Ultrasound Imaging
    Yan, Xin
    Qi, Yanxing
    Wang, Yinmeng
    Wang, Yuanyuan
    IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 2022, 69 (02) : 580 - 591
  • [36] Phase Aberration Correction in Transcranial Ultrasound Imaging using Averaged Sound Velocity Map in Delay-and-Sum Beamformer
    Mozaffarzadeh, Moein
    Minonzio, Claudio
    Verweij, Martin D.
    Hemm, Simone
    Daeichin, Verya
    2019 IEEE INTERNATIONAL ULTRASONICS SYMPOSIUM (IUS), 2019, : 1894 - 1897
  • [37] Elasticity Imaging using High Frame Rate Ultrasound Imaging
    Park, Suhyun
    Aglyamov, Salavat R.
    Scott, W. Guy
    Emelianov, Stanislav Y.
    2006 IEEE ULTRASONICS SYMPOSIUM, VOLS 1-5, PROCEEDINGS, 2006, : 602 - +
  • [38] Fast Marching based Tissue Adaptive Delay Estimation for Aberration Corrected Delay and Sum Beamforming in Ultrasound Imaging
    Asif, M. S.
    Malama, Gayathri
    Madhavanunni, A. N.
    Melapudi, Vikram
    Rahul, V
    Patil, Abhijit
    Langoju, Rajesh
    Panicker, Mahesh Raveendranatha
    2023 45TH ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE & BIOLOGY SOCIETY, EMBC, 2023, : 1 - 4
  • [39] High-resolution ultrasound imaging using unified pixel-based and filtered delay multiply and sum beamforming
    Guo, Hao
    Xie, Hui-Wen
    Zhou, Guang-Quan
    Nguyen, Nghia Q.
    Prager, Richard W.
    INTERNATIONAL ULTRASONICS SYMPOSIUM (IEEE IUS 2021), 2021,
  • [40] High Frame Rate Ultrasound Displacement Vector Imaging
    Kruizinga, Pieter
    Mastik, Frits
    Bosch, Johannes G.
    de Jong, Nico
    van Soest, Gijs
    van der Steen, Antonius F. W.
    2014 IEEE INTERNATIONAL ULTRASONICS SYMPOSIUM (IUS), 2014, : 707 - 709