Effective Dose to Patients from SPECT and CT During Myocardial Perfusion Imaging

被引:7
|
作者
Brindhaban, Ajit [1 ]
机构
[1] Kuwait Univ, Dept Radiol Sci, POB 31470, Sulibikhat 90805, Kuwait
关键词
myocardial perfusion imaging; SPECT/CT; administered activity; CT dose index; effective dose; ATTENUATION CORRECTION; NATIONAL-SURVEY;
D O I
10.2967/jnmt.119.233874
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Radiation dose to patients from imaging modalities is measured or calculated to assess the risk of the procedure and compare it with the benefits. Periodic review of image acquisition methods and the radiation dose used are an essential part of optimization in medical imaging. The aim of this study was to estimate patient radiation dose from SPECT myocardial perfusion imaging (MPI) using CT images for attenuation correction. Methods: SPECT and CT image acquisition parameters such as administered activity (AA), CT dose index (CTDIvol), and dose-length product for 415 patients who had undergone SPECT MPI using CT attenuation correction were reviewed. Effective dose (ED) for the SPECT part, the CT part and the total ED for the procedure were calculated. AA, CTDIvol, and ED values were compared between the 2 sexes and between body mass indexes (BMIs), imaging scanner models, and imaging centers. Statistical analyses were performed using t tests and 1-way ANOVA at P < 0.05 level of significance. Results: The range of AAs used for MPI was found to be 1,206-1,964 MBq per patient regardless of sex. The resulting mean ED of 8.8 mSv for men was significantly lower (P = 0.002) than the 10.4 mSv for women for SPECT. The range of CTDIvol was 1.12-3.97 mGy, resulting in an mean ED of 0.8 mSv for men, significantly lower (P, 0.001) than the 1.1 mSv for women for CT. The average combined EDs for male and female patients were 9.6 and 11.5 mSv, respectively. A positive correlation was found between AA and patient BMI (r = 0.48; P < 0.001), indicating patient size-related AAs. However, CTDIvol was found to depend only on the scanner model, regardless of BMI. Conclusion: The ED from SPECT/CT MPI studies was around 11 mSv, with 10 mSv being from the SPECT part of the study. The extra risk to the patients from CT imaging for attenuation correction is small compared with the benefit incurred from accurate diagnosis.
引用
收藏
页码:143 / 147
页数:5
相关论文
共 50 条
  • [41] Dose reduction in half-time myocardial perfusion SPECT-CT with multifocal collimation
    Lyon, Morgan C.
    Foster, Courtney
    Ding, Xinhong
    Dorbala, Sharmila
    Spence, Don
    Bhattacharya, Manojeet
    Vija, A. Hans
    DiCarli, Marcelo F.
    Moore, Stephen C.
    JOURNAL OF NUCLEAR CARDIOLOGY, 2016, 23 (04) : 657 - 667
  • [42] Dose reduction in half-time myocardial perfusion SPECT-CT with multifocal collimation
    Morgan C. Lyon
    Courtney Foster
    Xinhong Ding
    Sharmila Dorbala
    Don Spence
    Manojeet Bhattacharya
    A. Hans Vija
    Marcelo F. DiCarli
    Stephen C. Moore
    Journal of Nuclear Cardiology, 2016, 23 : 657 - 667
  • [43] Half-Time SPECT Myocardial Perfusion Imaging with Attenuation Correction
    Ali, Iftikhar
    Ruddy, Terrence D.
    Almgrahi, Abdulaziz
    Anstett, Frank G.
    Wells, R. Glenn
    JOURNAL OF NUCLEAR MEDICINE, 2009, 50 (04) : 554 - 562
  • [44] Dynamic CT myocardial perfusion imaging
    Nieman, Koen
    Balla, Sujana
    JOURNAL OF CARDIOVASCULAR COMPUTED TOMOGRAPHY, 2020, 14 (04) : 303 - 306
  • [45] Dynamic CT myocardial perfusion imaging
    Caruso, Damiano
    Eid, Marwen
    Schoepf, U. Joseph
    Jin, Kwang Nam
    Varga-Szemes, Akos
    Tesche, Christian
    Mangold, Stefanie
    Spandorfer, Adam
    Laghi, Andrea
    De Cecco, Carlo N.
    EUROPEAN JOURNAL OF RADIOLOGY, 2016, 85 (10) : 1893 - 1899
  • [46] Minimizing Patient-Specific Tracer Dose in Myocardial Perfusion Imaging Using CZT SPECT
    van Dijk, Joris D.
    Jager, Pieter L.
    Ottervanger, Jan Paul
    Slump, Cornelis H.
    de Boer, Jaep
    Oostdijk, Adrianus H. J.
    van Dalen, Jorn A.
    JOURNAL OF NUCLEAR MEDICINE TECHNOLOGY, 2015, 43 (01) : 36 - 40
  • [47] Advances in technical aspects of myocardial perfusion SPECT imaging
    Piotr J. Slomka
    James A. Patton
    Daniel S. Berman
    Guido Germano
    Journal of Nuclear Cardiology, 2009, 16 : 255 - 276
  • [48] Myocardial Perfusion Spect Imaging in Dextrocardia: A Case Report
    Ozdemir, Semra
    Gazi, Emine
    MOLECULAR IMAGING AND RADIONUCLIDE THERAPY, 2013, 22 (02) : 70 - 72
  • [49] Impact of SPECT myocardial perfusion imaging on cardiac care
    Lugomirski, Peter
    Chow, Benjamin J. W.
    Ruddy, Terrence D.
    EXPERT REVIEW OF CARDIOVASCULAR THERAPY, 2014, 12 (11) : 1247 - 1249
  • [50] Clinical Utility of CT-based Attenuation-correction in Myocardial Perfusion SPECT Imaging
    Hatipoglu, Filiz
    Cetin, Neslihan
    MOLECULAR IMAGING AND RADIONUCLIDE THERAPY, 2023, 32 (02) : 138 - 145