Physical characterization of photon-counting tomosynthesis

被引:1
作者
Berggren, Karl [1 ,2 ]
Lundqvist, Mats [1 ]
Cederstrom, Bjorn [1 ]
Danielsson, Mats [2 ]
Fredenberg, Erik [1 ]
机构
[1] Philips Healthcare, S-17141 Solna, Sweden
[2] Alabanova Univ Ctr, Royal Inst Technol KTH, Dept Phys, S-10691 Stockholm, Sweden
来源
MEDICAL IMAGING 2015: PHYSICS OF MEDICAL IMAGING | 2015年 / 9412卷
关键词
Tomosynthesis; Mammography; Photon-Counting; Modelling; MTF; NPS; Shift invariance; Stationarity; DIGITAL BREAST TOMOSYNTHESIS; CASCADED SYSTEMS-ANALYSIS; MAMMOGRAPHY; RESOLUTION; DETECTORS; 2-VIEW;
D O I
10.1117/12.2075192
中图分类号
O43 [光学];
学科分类号
070207 ; 0803 ;
摘要
Tomosynthesis is emerging as a next generation technology in mammography. Combined with photon-counting detectors with the ability for energy discrimination, a novel modality is enabled - spectral tomosynthesis. Further advantages of photon-counting detectors in the context of tomosynthesis include elimination of electronic noise, efficient scatter rejection (in some geometries) and no lag. Fourier-based linear-systems analysis is a well-established method for optimizing image quality in two-dimensional x-ray systems. The method has been successfully adapted to three-dimensional imaging, including tomosynthesis, but several areas need further investigation. This study focuses on two such areas: 1) Adaption of the methodology to photon-counting detectors, and 2) violation of the shift-invariance and stationarity assumptions in non-cylindrical geometries. We have developed a Fourier-based framework to study the image quality in a photon-counting tomosynthesis system, assuming locally linear, stationary, and shift-invariant system response. The framework includes a cascaded-systems model to propagate the modulation-transfer function (MTF) and noise-power spectrum (NPS) through the system. The model was validated by measurements of the MTF and NPS. High degrees of non-shift invariance and non-stationarity were observed, in particular for the depth resolution as the angle of incidence relative the reconstruction plane varied throughout the imaging volume. The largest effects on image quality in a given point in space were caused by interpolation from the inherent coordinate system of the x-rays to the coordinate system that was used for reconstruction. This study is part of our efforts to fully characterize the spectral tomosynthesis system, we intend to extend the model further to include the detective-quantum efficiency, observer modelling, and spectral effects.
引用
收藏
页数:12
相关论文
共 20 条
[1]   Linear response theory for detectors consisting of discrete arrays [J].
Albert, M ;
Maidment, ADA .
MEDICAL PHYSICS, 2000, 27 (10) :2417-2434
[2]   Physical characterization of a scanning photon counting digital mammography system based on Si-strip detectors [J].
Aslund, Magnus ;
Cederstrom, Bjorn ;
Lundqvist, Mats ;
Danielsson, Mats .
MEDICAL PHYSICS, 2007, 34 (06) :1918-1925
[3]   Dual-energy contrast-enhanced digital breast tomosynthesis - a feasibility study [J].
Carton, A-K ;
Gavenonis, S. C. ;
Currivan, J. A. ;
Conant, E. F. ;
Schnall, M. D. ;
Maidment, A. D. A. .
BRITISH JOURNAL OF RADIOLOGY, 2010, 83 (988) :344-350
[4]  
Fredenberg E., 2009, PHOTON COUNTING DETE
[5]   Energy resolution of a photon-counting silicon strip detector [J].
Fredenberg, Erik ;
Lundqvist, Mats ;
Cederstrom, Bjorn ;
Aslund, Magnus ;
Danielsson, Mats .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2010, 613 (01) :156-162
[6]  
Gang G. J., 2013, P SOC PHOTO-OPT INS, V8668
[7]   Digital Breast Tomosynthesis in the Diagnostic Environment: A Subjective Side-by-Side Review [J].
Hakim, Christiane M. ;
Chough, Denise M. ;
Ganott, Marie A. ;
Sumkin, Jules H. ;
Zuley, Margarita L. ;
Gur, David .
AMERICAN JOURNAL OF ROENTGENOLOGY, 2010, 195 (02) :W172-W176
[8]   Overview of the evidence on digital breast tomosynthesis in breast cancer detection [J].
Houssami, Nehmat ;
Skaane, Per .
BREAST, 2013, 22 (02) :101-108
[9]   Optimization of dual-energy imaging systems using generalized NEQ and imaging task [J].
Richard, S. ;
Siewerdsen, J. H. .
MEDICAL PHYSICS, 2007, 34 (01) :127-139
[10]   Cascaded systems analysis of noise reduction algorithms in dual-energy imaging [J].
Richard, Samuel ;
Siewerdsen, Jeffrey H. .
MEDICAL PHYSICS, 2008, 35 (02) :586-601