From conventional proton-transfer-reaction mass spectrometry (PTR-MS) to universal trace gas analysis

被引:55
作者
Sulzer, Philipp [1 ]
Edtbauer, Achim [1 ]
Hartungen, Eugen [1 ]
Juerschik, Simone [1 ]
Jordan, Alfons [1 ]
Hanel, Gernot [1 ]
Feil, Stefan [1 ]
Jaksch, Stefan [1 ]
Maerk, Lukas [1 ]
Maerk, Tilmann D. [1 ,2 ]
机构
[1] IONICON Analyt Gesell mbH, A-6020 Innsbruck, Austria
[2] Univ Innsbruck, Inst Ionenphys & Angew Phys, A-6020 Innsbruck, Austria
关键词
PTR-MS; PTR-TOFMS; Chemical ionization; Trace gas analysis; Charge transfer; Direct injection mass spectrometry; REAL-TIME DETECTION; FLOW TUBE;
D O I
10.1016/j.ijms.2012.05.003
中图分类号
O64 [物理化学(理论化学)、化学物理学]; O56 [分子物理学、原子物理学];
学科分类号
070203 ; 070304 ; 081704 ; 1406 ;
摘要
We present here a slightly modified PTR-MS instrument that is not only capable to ionize trace compounds in air via proton-transfer-reactions (PTR) but is also able to ionize via charge-transfer-reactions (CTR) with help of reagent ions (Kr+ in particular) possessing higher ionization energies than common air constituents. This means that with minor adaptations a common PTR-MS instrument can be used for the analysis of nearly all available substance classes by using both PTR and/or CTR ionization. Especially in environmental research, the field of application where PTR-MS is used most widely, now not only trace volatile organic compounds (benzene, toluene, etc.) but additionally also very important (inorganic) substances, such as CO, CO2, CH4, NOx, and SO2, can be detected and quantified with the same instrument. As all ionizing agents are produced in a hollow cathode discharge ion source with good purity no additional mass filter is needed for reagent ion selection (as in other analytical methods employed) and remaining reagent ion impurities can be clearly distinguished from isobaric sample compounds due to the high mass resolution of the time-of-flight mass spectrometer used in the present PTR-MS instrument (PTR-TOF 8000). We present data obtained with various gas standards ranging from a "classical" PTR-MS aromatics mixture to samples containing molecules possessing ionization energies all the way up to 14 eV (CO). (c) 2012 Elsevier B.V. All rights reserved.
引用
收藏
页码:66 / 70
页数:5
相关论文
共 18 条
[1]   Detection of isocyanates and polychlorinated biphenyls using proton transfer reaction mass spectrometry [J].
Agarwal, B. ;
Juerschik, S. ;
Sulzer, P. ;
Petersson, F. ;
Jaksch, S. ;
Jordan, A. ;
Maerk, T. D. .
RAPID COMMUNICATIONS IN MASS SPECTROMETRY, 2012, 26 (08) :983-989
[2]   Use of proton transfer reaction time-of-flight mass spectrometry for the analytical detection of illicit and controlled prescription drugs at room temperature via direct headspace sampling [J].
Agarwal, B. ;
Petersson, F. ;
Juerschik, S. ;
Sulzer, P. ;
Jordan, A. ;
Maerk, T. D. ;
Watts, P. ;
Mayhew, C. A. .
ANALYTICAL AND BIOANALYTICAL CHEMISTRY, 2011, 400 (08) :2631-2639
[3]  
Anicich V.G., 2009, JPL PUBLICATION, V03-19
[4]   Flow tube studies of benzene charge transfer reactions from 250 to 1400 K [J].
Arnold, ST ;
Williams, S ;
Dotan, I ;
Midey, AJ ;
Morris, RA ;
Viggiano, AA .
JOURNAL OF PHYSICAL CHEMISTRY A, 1999, 103 (42) :8421-8432
[5]   Direct-injection mass spectrometry adds the time dimension to (B)VOC analysis [J].
Biasioli, Franco ;
Yeretzian, Chahan ;
Maerk, Tilmann D. ;
Dewulf, J. ;
Van Langenhove, H. .
TRAC-TRENDS IN ANALYTICAL CHEMISTRY, 2011, 30 (07) :1003-1017
[6]   PTR-MS monitoring of VOCs and BVOCs in food science and technology [J].
Biasioli, Franco ;
Yeretzian, Chahan ;
Gasperi, Flavia ;
Maerk, Tilmann D. .
TRAC-TRENDS IN ANALYTICAL CHEMISTRY, 2011, 30 (07) :968-977
[7]   Proton-Transfer Reaction Mass Spectrometry [J].
Blake, Robert S. ;
Monks, Paul S. ;
Ellis, Andrew M. .
CHEMICAL REVIEWS, 2009, 109 (03) :861-896
[8]   Detection of Plant Volatiles after Leaf Wounding and Darkening by Proton Transfer Reaction "Time-of-Flight" Mass Spectrometry (PTR-TOF) [J].
Brilli, Federico ;
Ruuskanen, Taina M. ;
Schnitzhofer, Ralf ;
Mueller, Markus ;
Breitenlechner, Martin ;
Bittner, Vinzenz ;
Wohlfahrt, Georg ;
Loreto, Francesco ;
Hansel, Armin .
PLOS ONE, 2011, 6 (05)
[9]   Measurements of volatile organic compounds in the earths atmosphere using proton-transfer-reaction mass spectrometry [J].
de Gouw, Joost ;
Warneke, Carsten .
MASS SPECTROMETRY REVIEWS, 2007, 26 (02) :223-257
[10]   On-line breath analysis with PTR-TOF [J].
Herbig, Jens ;
Mueller, Markus ;
Schallhart, Simon ;
Titzmann, Thorsten ;
Graus, Martin ;
Hansel, Armin .
JOURNAL OF BREATH RESEARCH, 2009, 3 (02)