Effect of Repetitive Loading on the Mechanical Properties of Synthetic Hernia Repair Materials

被引:26
作者
Eliason, Braden J. [1 ]
Frisella, Margaret M. [1 ]
Matthews, Brent D. [1 ]
Deeken, Corey R. [1 ]
机构
[1] Washington Univ, Sch Med, Dept Surg, Sect Minimally Invas Surg, St Louis, MO 63110 USA
关键词
MESH;
D O I
10.1016/j.jamcollsurg.2011.05.018
中图分类号
R61 [外科手术学];
学科分类号
摘要
BACKGROUND: Hernia repair materials undergo repeated loading while in the body, and the impact on mechanical properties is unknown. It was hypothesized that exposure to repetitive loading would lead to decreased tensile strength and increased strain, and that these differences would become more pronounced with greater loading and unloading sequences. STUDY DESIGN: Polypropylene, expanded polytetrafluoroethylene, composite barrier, and partially absorbable meshes were evaluated. Twenty specimens (7.5 X 7.5 cm) were prepared from each material. Five specimens were subjected to ball burst testing to determine baseline biomechanical properties. Cycles of 10, 100, and 1,000 loading sequences were also performed (n = 5 each). RESULTS: BardMesh (CR Bard/Davol), Dualmesh (WL Gore), and Prolene (Ethicon) exhibited significantly reduced tensile strength; BardMesh, Proceed (Ethicon), Prolene, Prolene (Atrium Medical), ProLite Ultra (Atrium Medical), and Ultrapro (Ethicon) exhibited significantly increased strain after exposure to 1,000 cycles compared with their baseline properties. BardMesh and Prolene demonstrated both reduced tensile strength and increased strain values after 1,000 cycles, suggesting that repetitive loading has the greatest effects on these materials. In addition, BardMesh and Prolene exhibited progressively worsening effects as the number of cycles was increased. CONCLUSIONS: Deterioration of the tensile strength of the mesh or an increase in the ability of the mesh material to stretch (ie, increased strain values) could potentially lead to hernia recurrence or a poor functional result. However, the results of this study should not be interpreted to mean that hernia repair materials will fail in the body after only 10, 100, or 1,000 cycles. The conditions used in this study were more extreme than most physiologic scenarios and were intended as a pilot investigation into how the mechanical properties of hernia repair materials are affected by in vitro cyclic testing. (J Am Coll Surg 2011;213:430-435. (C) 2011 by the American College of Surgeons)
引用
收藏
页码:430 / 435
页数:6
相关论文
共 11 条
[1]   The extracellular matrix as a biologic scaffold material [J].
Badylak, Stephen F. .
BIOMATERIALS, 2007, 28 (25) :3587-3593
[2]   Normal intraabdominal pressure in healthy adults [J].
Cobb, WS ;
Burns, JM ;
Kercher, KW ;
Matthews, BD ;
Norton, HJ ;
Heniford, BT .
JOURNAL OF SURGICAL RESEARCH, 2005, 129 (02) :231-235
[3]  
DAYTON MT, 1986, ARCH SURG-CHICAGO, V121, P954
[4]   Physicomechanical evaluation of absorbable and nonabsorbable barrier composite meshes for laparoscopic ventral hernia repair [J].
Deeken, Corey R. ;
Abdo, Michael S. ;
Frisella, Margaret M. ;
Matthews, Brent D. .
SURGICAL ENDOSCOPY AND OTHER INTERVENTIONAL TECHNIQUES, 2011, 25 (05) :1541-1552
[5]   Physicomechanical Evaluation of Polypropylene, Polyester, and Polytetrafluoroethylene Meshes for Inguinal Hernia Repair [J].
Deeken, Corey R. ;
Abdo, Michael S. ;
Frisella, Margaret M. ;
Matthews, Brent D. .
JOURNAL OF THE AMERICAN COLLEGE OF SURGEONS, 2011, 212 (01) :68-79
[6]  
Junge K, 2001, Hernia, V5, P113
[7]   Functional assessment and tissue response of short- and long-term absorbable surgical meshes [J].
Klinge, U ;
Schumpelick, V ;
Klosterhalfen, B .
BIOMATERIALS, 2001, 22 (11) :1415-1424
[8]   The lightweight and large porous mesh concept for hernia repair [J].
Klosterhalfen, B ;
Junge, K ;
Klinge, U .
EXPERT REVIEW OF MEDICAL DEVICES, 2005, 2 (01) :103-117
[9]  
LERDSIRISOPON S, 2011, SAGES ANN M SAN ANT
[10]  
RUIZJASBON F, 2010, NORRBY 6 MONTHS RESU