Left Ventricular Flow Dynamics with the HeartMate3 Left Ventricular Assist Device: Effect of Inflow Cannula Position and Speed Modulation

被引:12
作者
Ortiz, Sean [1 ]
Vu, Vi [1 ]
Montes, Ricardo [1 ]
May-Newman, Karen [1 ]
机构
[1] San Diego State Univ, Dept Mech Engn, Bioengn Program, San Diego, CA USA
关键词
flow; velocity; vortex; left ventricular assist device; pulsatility; thromboembolism; PULSATILE-FLOW; RISK-FACTORS; IMPLANTATION; THROMBUS; STASIS; MODEL;
D O I
10.1097/MAT.0000000000001523
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Improper left ventricular assist device (LVAD) inflow cannula (IC) positioning creates areas of stasis and low pulsatility that predispose thromboembolism, but may be mitigated with LVAD speed modulation. A mock loop study was performed to assess the sensitivity of left ventricle (LV) flow architecture to IC position and speed modulation during HeartMate3 support. System pressure, flow, and the time-resolved velocity field were measured within a transparent silicone LV for three IC angles and three IC insertion depths at matched levels of cardiac function and LVAD speed. Inflow cannula angulation towards the septum increased the resistance to LVAD flow as well as increasing the size and energy of the counter-clockwise (CCW) vortex. Apical velocity was reduced compared to IC angulation towards the mitral valve, but regional pulsatility was maintained across all angles and LVAD speeds. Increased IC protrusion decreased LVAD flow resistance, increasing velocity within the IC but reducing flow and pulsatility in the adjacent apical region. Increasing LVAD flow resistance improves aortic valve opening and strengthens the CCW vortex which directs inflow towards the septum, producing higher blood residence time and shear activation potential. Despite this impact on flow architecture, pulsatility reduction with increased LVAD speed was minimal with the HeartMate3 speed modulation feature.
引用
收藏
页码:1301 / 1311
页数:11
相关论文
共 37 条
[1]   INTERMACS Analysis of Stroke During Support With Continuous-Flow Left Ventricular Assist Devices Risk Factors and Outcomes [J].
Acharya, Deepak ;
Loyaga-Rendon, Renzo ;
Morgan, Charity J. ;
Sands, Kara A. ;
Pamboukian, Salpy V. ;
Rajapreyar, Indranee ;
Holman, William L. ;
Kirklin, James K. ;
Tallaj, Jose A. .
JACC-HEART FAILURE, 2017, 5 (10) :703-711
[2]   Radio logic assessment of HeartMate II position: Minimal pump migration after long-term support [J].
Adamson, Robert M. ;
Bower, Bruce L. ;
Sundareswaran, Kartik S. ;
Farrar, David J. ;
Dembitsky, Walter P. .
JOURNAL OF HEART AND LUNG TRANSPLANTATION, 2015, 34 (12) :1617-1623
[3]   Principles of HeartMate II Implantation to Avoid Pump Malposition and Migration [J].
Adamson, Robert M. ;
Mangi, Abeel A. ;
Kormos, Robert L. ;
Farrar, David J. ;
Dembitsky, Walter P. .
JOURNAL OF CARDIAC SURGERY, 2015, 30 (03) :296-299
[4]   Intraventricular vortex properties in nonischemic dilated cardiomyopathy [J].
Bermejo, Javier ;
Benito, Yolanda ;
Alhama, Marta ;
Yotti, Raquel ;
Martinez-Legazpi, Pablo ;
Perez del Villar, Candelas ;
Perez-David, Esther ;
Gonzalez-Mansilla, Ana ;
Santa-Marta, Cristina ;
Barrio, Alicia ;
Fernandez-Aviles, Francisco ;
del Alamo, Juan C. .
AMERICAN JOURNAL OF PHYSIOLOGY-HEART AND CIRCULATORY PHYSIOLOGY, 2014, 306 (05) :H718-H729
[5]   Research approaches for studying flow-induced thromboembolic complications in blood recirculating devices [J].
Bluestein, D .
EXPERT REVIEW OF MEDICAL DEVICES, 2004, 1 (01) :65-80
[6]   Left Ventricular Assist Device Inflow Cannula Insertion Depth Influences Thrombosis Risk [J].
Chivukula, Venkat Keshav ;
Beckman, Jennifer A. ;
Li, Song ;
Masri, S. Carolina ;
Levy, Wayne C. ;
Lin, Shin ;
Cheng, Richard K. ;
Farris, Stephen D. ;
Wood, Gregory ;
Dardas, Todd F. ;
Kirkpatrick, James N. ;
Koomalsingh, Kevin ;
Zimpfer, Daniel ;
Mackensen, G. Burkhard ;
Chassagne, Fanette ;
Mahr, Claudius ;
Aliseda, Alberto .
ASAIO JOURNAL, 2020, 66 (07) :766-773
[7]   Small Left Ventricular Size Is an Independent Risk Factor for Ventricular Assist Device Thrombosis [J].
Chivukula, Venkat Keshav ;
Beckman, Jennifer A. ;
Prisco, Anthony R. ;
Lin, Shin ;
Dardas, Todd F. ;
Cheng, Richard K. ;
Farris, Stephen D. ;
Smith, Jason W. ;
Mokadam, Nahush A. ;
Mahr, Claudius ;
Aliseda, Alberto .
ASAIO JOURNAL, 2019, 65 (02) :152-159
[8]   Left Ventricular Assist Device Inflow Cannula Angle and Thrombosis Risk [J].
Chivukula, Venkat Keshav ;
Beckman, Jennifer A. ;
Prisco, Anthony R. ;
Dardas, Todd ;
Lin, Shin ;
Smith, Jason W. ;
Mokadam, Nahush A. ;
Aliseda, Alberto ;
Mahr, Claudius .
CIRCULATION-HEART FAILURE, 2018, 11 (04)
[9]   Combined experimental and numerical analysis of the flow structure into the left ventricle [J].
Domenichini, F. ;
Querzoli, G. ;
Cenedese, A. ;
Pedrizzetti, G. .
JOURNAL OF BIOMECHANICS, 2007, 40 (09) :1988-1994
[10]   The effect of aortic valve incompetence on the hemodynamics of a continuous flow, ventricular assist device in a mock circulation [J].
Garcia, Mario A. Zamarripa ;
Enriquez, Luz A. ;
Dembitsky, Walter ;
May-Newman, Karen .
ASAIO JOURNAL, 2008, 54 (03) :237-244