Rapid electrical immunoassay of the cardiac biomarker troponin I through dielectrophoretic concentration using imbedded electrodes

被引:45
|
作者
Sharma, Abhinav [1 ]
Han, Chang-Ho [1 ]
Jang, Jaesung [1 ,2 ,3 ]
机构
[1] Ulsan Natl Inst Sci & Technol, Sch Mech & Nucl Engn, Ulsan 44919, South Korea
[2] Ulsan Natl Inst Sci & Technol, Dept Biomed Engn, Ulsan 44919, South Korea
[3] Ulsan Natl Inst Sci & Technol, Sch Mat Sci & Engn, Ulsan 44919, South Korea
基金
新加坡国家研究基金会;
关键词
Cardiovascular disease; Carbon nanotubes; Cardiac troponin I; Dielectrophoresis; Immunoassay; MYOCARDIAL INJURY; PEPTIDE NANOTUBES; IMMUNOSENSOR; BIOSENSORS; MYOGLOBIN; SERUM; NANOPARTICLES; MANIPULATION; ADSORPTION; CHALLENGES;
D O I
10.1016/j.bios.2016.03.056
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
Rapidity and high sensitivity are critical factors for the diagnoses of heart attacks, and cardiac troponin I (cTnI) is at present a clinical standard for its diagnosis. Here we report a rapid, label-free, and highly sensitive single-walled carbon nanotube (SWCNT) electrical immunosensor, featuring two pairs of electrodes. Two concentration electrodes (gaps: 25 and 80 gm) and two detection electrodes (source and drain; gap: 20 gm; width: 50 gm) were used for dielectrophoretic concentration of cTnI on the SWCNT channels and resistance measurements of the dielectrophoresis (DEP)-concentrated cTnI, respectively. The two concentration electrodes were imbedded between upper and lower dielectric layers, facing each other, underneath the -COOH-functionalized SWCNT channels deposited between the detection electrodes. Therefore, the gap between these imbedded concentration electrodes can be reduced to maximize the electric field intensity for DEP-mediated concentration of cTnI, thereby greatly reducing the detection time (1 min) and enhancing the limit of detection (0.7-0.8 pg mL(-1)). Relative resistance changes of the SWCNTs were measured as cTnl concentration in Tris-Borate-EDTA (TBE; 0.0025 x) and human serum diluted 500-fold with 0.0025 x TBE decreased from 100 ng mL(-1) to 1 pg mL(-1), and they were shown to be linear with the logarithm of cTnI concentration (R-2=0.99 and 0.97, respectively). These immunosensors also showed high specificity over another cardiac biomarker, myoglobin, TBE medium (0.0025 x), and 500-fold diluted human serum. The DEP-capture of cTnI depended on the frequency of the applied electric field, demonstrating the qualitative nature of the real part of the Clausius-Mossotti factor for cTnl. (C) 2016 Elsevier B.V. All rights reserved.
引用
收藏
页码:78 / 84
页数:7
相关论文
共 22 条
  • [11] Zero Background Visualizing Phosphorescence Lateral Flow Immunoassay of Cardiac Troponin I for Rapid and Accurate Diagnosis of Myocardial Infarction
    Song, Qingwei
    Liao, Zixuan
    Zheng, Lin
    Fu, Pan
    Qian, Sihua
    Liang, Wei
    Zheng, Jianping
    Wang, Kaizhe
    Wang, Yuhui
    ANALYTICAL CHEMISTRY, 2025, 97 (06) : 3651 - 3660
  • [12] Development of lateral flow immunoassay system based on superparamagnetic nanobeads as labels for rapid quantitative detection of cardiac troponin I
    Xu, QuanFu
    Xu, Hong
    Gu, Hongchen
    Li, JingBo
    Wang, Yanyan
    Wei, Meng
    MATERIALS SCIENCE & ENGINEERING C-BIOMIMETIC AND SUPRAMOLECULAR SYSTEMS, 2009, 29 (03): : 702 - 707
  • [13] Highly sensitive amperometric detection of cardiac troponin I using sandwich aptamers and screen-printed carbon electrodes
    Jo, Hunho
    Her, Jin
    Lee, Heehyun
    Shim, Yoon-Bo
    Ban, Changill
    TALANTA, 2017, 165 : 442 - 448
  • [14] Sensitive immunoassay of cardiac troponin I using an optimized microelectrode array in a novel integrated microfluidic electrochemical device
    Yang Li
    Shaohua Zuo
    Laiqian Ding
    Pengpeng Xu
    Kun Wang
    Yuanchang Liu
    Jingmin Li
    Chong Liu
    Analytical and Bioanalytical Chemistry, 2020, 412 : 8325 - 8338
  • [15] Fabrication and Characterization of Aluminium Interdigitated Electrodes (IDE) Hybrid with Zinc Oxide (ZnO) Nanoparticles for Detection of Cardiac Troponin I (cTnI) Biomarker
    Tasakaren, Thavamani.
    Arshad, M. K. Md
    Nuzaihan, M. M. N.
    Letchumanan, Iswary
    Fathil, M. F. M.
    Ibau, C.
    2019 IEEE INTERNATIONAL CONFERENCE ON SENSORS AND NANOTECHNOLOGY (SN), 2019, : 49 - 52
  • [16] Interference of laboratory disinfection with trichloro-isocyanuric acid on cardiac troponin I measurement using the Vitros immunoassay system
    Zhao, Zhen
    Pan, Huifen
    Gu, Wenhong
    CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2017, 55 (04) : E80 - E83
  • [17] Highly sensitive detection of cardiac troponin I in human serum using gold nanoparticle-based enhanced sandwich immunoassay
    Kim, Wan-Joong
    Cho, Hyo Young
    Kim, Bong Kyu
    Huh, Chul
    Chung, Kwang Hyo
    Ahn, Chang-Geun
    Kim, Young Jun
    Kim, Ansoon
    SENSORS AND ACTUATORS B-CHEMICAL, 2015, 221 : 537 - 543
  • [18] A novel integrated lateral flow immunoassay platform for the detection of cardiac troponin I using hierarchical dendritic copper-nickel nanostructures
    Tan, Yuxin
    Zhang, Shirong
    Liu, Yilei
    Li, Jishun
    Zhang, Shenglan
    Pan, Hongcheng
    TALANTA, 2024, 277
  • [19] A Rapid, Direct, Quantitative, and Label-Free Detector of Cardiac Biomarker Troponin T Using Near-Infrared Fluorescent Single-Walled Carbon Nanotube Sensors
    Zhang, Jingqing
    Kruss, Sebastian
    Hilmer, Andrew J.
    Shimizu, Steven
    Schmois, Zeke
    De La Cruz, Flor
    Barone, Paul W.
    Reuel, Nigel F.
    Heller, Daniel A.
    Strano, Michael S.
    ADVANCED HEALTHCARE MATERIALS, 2014, 3 (03) : 412 - 423
  • [20] Photoelectrochemical Determination of Cardiac Troponin I as a Biomarker of Myocardial Infarction Using a Bi2S3 Film Electrodeposited on a BiVO4-Coated Fluorine-Doped Tin Oxide Electrode
    Monteiro, Thatyara Oliveira
    Neto, Antonio Gomes dos Santos
    de Menezes, Alan Silva
    Damos, Flavio Santos
    Luz, Rita de Cassia Silva
    Fatibello-Filho, Orlando
    BIOSENSORS-BASEL, 2023, 13 (03):