In aqua vivo EPID dosimetry

被引:59
作者
Wendling, Markus [1 ]
McDermott, Leah N. [1 ]
Mans, Anton [1 ]
Olaciregui-Ruiz, Igor [1 ]
Pecharroman-Gallego, Raul [1 ]
Sonke, Jan-Jakob [1 ]
Stroom, Joep [1 ]
van Herk, Marcel [1 ]
Mijnheer, Ben J. [1 ]
机构
[1] Antoni van Leeuwenhoek Hosp, Netherlands Canc Inst, Dept Radiat Oncol, NL-1066 CX Amsterdam, Netherlands
关键词
in vivo dosimetry; dose verification; EPID; quality assurance; lung cancer; MODULATED ARC THERAPY; CONE-BEAM CT; TREATMENT VERIFICATION; DOSE VERIFICATION; IMRT TREATMENTS; PROSTATE IMRT; RADIOTHERAPY; 3D; ALGORITHM;
D O I
10.1118/1.3665709
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: At the Netherlands Cancer Institute-Antoni van Leeuwenhoek Hospital in vivo dosimetry using an electronic portal imaging device (EPID) has been implemented for almost all high-energy photon treatments of cancer with curative intent. Lung cancer treatments were initially excluded, because the original back-projection dose-reconstruction algorithm uses water-based scatter-correction kernels and therefore does not account for tissue inhomogeneities accurately. The aim of this study was to test a new method, in aqua vivo EPID dosimetry, for fast dose verification of lung cancer irradiations during actual patient treatment. Methods: The key feature of our method is the dose reconstruction in the patient from EPID images, obtained during the actual treatment, whereby the images have been converted to a situation as if the patient consisted entirely of water; hence, the method is termed in aqua vivo. This is done by multiplying the measured in vivo EPID image with the ratio of two digitally reconstructed transmission images for the unit-density and inhomogeneous tissue situation. For dose verification, a comparison is made with the calculated dose distribution with the inhomogeneity correction switched off. IMRT treatment verification is performed for each beam in 2D using a 2D gamma evaluation, while for the verification of volumetric-modulated arc therapy (VMAT) treatments in 3D a 3D gamma evaluation is applied using the same parameters (3%, 3 mm). The method was tested using two inhomogeneous phantoms simulating a tumor in lung and measuring its sensitivity for patient positioning errors. Subsequently five IMRT and five VMAT clinical lung cancer treatments were investigated, using both the conventional back-projection algorithm and the in aqua vivo method. The verification results of the in aqua vivo method were statistically analyzed for 751 lung cancer patients treated with IMRT and 50 lung cancer patients treated with VMAT. Results: The improvements by applying the in aqua vivo approach are considerable. The percentage of gamma values <= 1 increased on average from 66.2% to 93.1% and from 43.6% to 97.5% for the IMRT and VMAT cases, respectively. The corresponding mean gamma value decreased from 0.99 to 0.43 for the IMRT cases and from 1.71 to 0.40 for the VMAT cases, which is similar to the accepted clinical values for the verification of IMRT treatments of prostate, rectum, and head-and-neck cancers. The deviation between the reconstructed and planned dose at the isocenter diminished on average from 5.3% to 0.5% for the VMAT patients and was almost the same, within 1%, for the IMRT cases. The in aqua vivo verification results for IMRT and VMAT treatments of a large group of patients had a mean gamma of approximately 0.5, a percentage of gamma values <= 1 larger than 89%, and a difference of the isocenter dose value less than 1%. Conclusions: With the in aqua vivo approach for the verification of lung cancer treatments (IMRT and VMAT), we can achieve results with the same accuracy as obtained during in vivo EPID dosimetry of sites without large inhomogeneities. (C) 2012 American Association of Physicists in Medicine. [DOI: 10.1118/1.3665709]
引用
收藏
页码:367 / 377
页数:11
相关论文
共 19 条
[1]  
[Anonymous], 2008, IAEA-TECDOC-1583
[2]  
BIR IPEM NPSA SCR RCR, 2008, SAF RAD
[3]   Lessons from recent accidents in radiation therapy in France [J].
Derreumaux, S. ;
Etard, C. ;
Huet, C. ;
Trompier, F. ;
Clairand, I. ;
Bottollier-Depois, J. -F. ;
Aubert, B. ;
Gourmelon, P. .
RADIATION PROTECTION DOSIMETRY, 2008, 131 (01) :130-135
[4]   Impact of simple tissue inhomogeneity correction algorithms on conformal radiotherapy of lung tumours [J].
Engelsman, M ;
Damen, EMF ;
Koken, PW ;
van 't Veld, AA ;
van Ingen, KM ;
Mijnheer, BJ .
RADIOTHERAPY AND ONCOLOGY, 2001, 60 (03) :299-309
[5]   Catching errors with in vivo EPID dosimetry [J].
Mans, A. ;
Wendling, M. ;
McDermott, L. N. ;
Sonke, J. -J. ;
Tielenburg, R. ;
Vijlbrief, R. ;
Mijnheer, B. ;
van Herk, M. ;
Stroom, J. C. .
MEDICAL PHYSICS, 2010, 37 (06) :2638-2644
[6]   3D Dosimetric verification of volumetric-modulated arc therapy by portal dosimetry [J].
Mans, Anton ;
Remeijer, Peter ;
Olaciregui-Ruiz, Igor ;
Wendling, Markus ;
Sonke, Jan-Jakob ;
Mijnheer, Ben ;
van Herk, Marcel ;
Stroom, Joep C. .
RADIOTHERAPY AND ONCOLOGY, 2010, 94 (02) :181-187
[7]   Quality assurance of volumetric modulated arc therapy: Evaluation and comparison of different dosimetric systems [J].
Masi, L. ;
Casamassima, F. ;
Doro, R. ;
Francescon, P. .
MEDICAL PHYSICS, 2011, 38 (02) :612-621
[8]   Clinical experience with EPID dosimetry for prostate IMRT pre-treatment dose verification [J].
McDermott, L. N. ;
Wendling, M. ;
van Asselen, B. ;
Stroom, J. ;
Sonke, J. -J. ;
van Herk, M. ;
Mijnheer, B. J. .
MEDICAL PHYSICS, 2006, 33 (10) :3921-3930
[9]   3D in vivo dose verification of entire hypo-fractionated IMRT treatments using an EPID and cone-beam CT [J].
McDermott, Leah N. ;
Wendling, Markus ;
Nijkamp, Jasper ;
Mans, Anton ;
Sonke, Jan-Jakob ;
Mijnheer, Ben J. ;
van Herk, Marcel .
RADIOTHERAPY AND ONCOLOGY, 2008, 86 (01) :35-42
[10]   Replacing pretreatment verification with in vivo EPID dosimetry for prostate IMRT [J].
McDermott, Leah N. ;
Wendling, Markus ;
Sonke, Jan-Jakob ;
van Herk, Marcel ;
Mijnheer, Ben J. .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2007, 67 (05) :1568-1577